Datasheet summary
SP Series 0.5 x 1
Engineering summary from PDF text extraction for SP Series 0.5 x 1. Verify every value with the OEM datasheet.
Summary
Product Identification
- 3M™ Liqui-Cel™ SP-0.5x1 Series Membrane Contactor
Mechanical / Hydraulic
- Cartridge Configuration: No Baffle Membrane Array Design
- Membrane Type: UP
- Membrane/Potting Material: Polyolefin / Epoxy
- Priming Volume (approximate):
- Shellside: 1.7 ml
- Lumenside: 1.0 ml
- Housing Options and Characteristics:
- Material: Black Polyethylene
- Flange Connections:
- Shellside (Liquid Inlet/Outlet): Female Luer Lock
- Lumenside (vacuum): Female Luer Lock
- Weight (approximate):
- Dry: 8 grams
- Water-filled: 10 grams
- Nominal Dimensions:
- 47 mm (1.8 in)
- 49 mm (1.9 in)
- 28 mm (1.1 in)
- 13 mm (0.5 in)
- Note: All dimensions are nominal. Refer to liqui-cel.com for detailed housing drawings.
Operating Limits
- Liquid Flow Guidelines: 5 – 30 ml/min (Liquid must flow on the shellside)
- Pressure Guidelines*:
- Maximum Shellside LIQUID Working Temperature/Pressure:
- 5-20°C, 3.1 barg (41-68°F, 45 psig)
- 45°C, 1.7 barg (113°F, 25 psig)
- Note: Liquid pressure should always exceed gas pressure.
- Maximum Shellside LIQUID Working Temperature/Pressure:
- Gas side: Vacuum or Sweep (Lumenside)
- O₂ Removal Test Condition: Vacuum mode with water at 20°C. Vacuum: 50 torr.
Regulatory
- Complies with the limits as set by RoHS Directive 2011/65/EU Annex II; recasting 2002/95/EC.
- Constructed of FDA CFR title 21 compliant materials for wetted parts only at ambient temperatures.
Disclaimer: This summary is based on extracted text and may not include all figures, footnotes, or the latest revisions. For contractual data, always refer to the specific OEM PDF revision used on your project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.