Datasheet summary
Amberjet 1200 Na
Engineering summary from PDF text extraction for Amberjet 1200 Na. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: AMBERJET™ 1200 Na
- Product Type: Industrial Grade Strong Acid Cation Exchanger
- Matrix: Styrene divinylbenzene copolymer
- Functional group: Sulfonate
- Ionic form as shipped: Na⁺
Performance
Operating capacity and ionic leakage depend on water analysis, temperature, and regenerant level.
Typical Properties
- Physical form: Amber spherical beads
- Total exchange capacity: ≥ 2.00 eq/L (Na⁺ form) [1]
- Moisture holding capacity: 43 to 47% (Na⁺ form) [1]
- Shipping weight: 850 g/L
- Specific gravity: 1.26 to 1.30 (Na⁺ form)
- Particle size:
- Uniformity coefficient: ≤ 1.2 [1]
- Harmonic mean size: 620 ± 50 µm
- Fines content: < 0.300 mm : 0.1% max [1]
- Maximum reversible swelling: Na⁺ ⇌ H⁺ : 10%
[1] Contractual value. Test methods are available on request.
Suggested Operating Conditions
- Maximum operating temperature: 135°C
- Minimum bed depth: 800 mm
- Service flow rate: 5 to 50 BV*/h
- Maximum service velocity: 60 m/h
Regeneration
| Regenerant | Level (g/L) | Concentration (%) | Minimum contact time | Rinse (Slow/Fast) |
|---|---|---|---|---|
| NaCl | 50 to 240 | 10 | 20 minutes | 2 BV / 1 to 3 BV |
| HCl | 40 to 150 | 4 to 10 | 20 minutes | 2 BV at regen. flow / 1-3 BV at service flow |
| H₂SO₄ | 40 to 200 | 1 to 8 | 20 minutes | 2 BV at regen. flow / 1-3 BV at service flow |
- 1 BV (Bed Volume) = 1 m³ solution per m³ resin.
Limits of Use
AMBERJET 1200 Na resin is suitable for industrial uses. For pharmaceutical, food processing, or potable water applications, users should consult Rohm and Haas.
Hydraulic Characteristics
Figure 1 (in the original document) shows bed expansion as a function of backwash flow rate and water temperature. Figure 2 (in the original document) shows pressure drop data as a function of service flow rate and water temperature. Pressure drop data are valid at the start of the service run with clear water and a correctly classified bed.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.