Product Identification

These are improved High Temperature Stable (HTS) Electropure™ EDI products, suitable for pharmaceutical OEMs. The XL-HTS™ and EXL-HTS™ models are available in various capacities.

Key Benefits:

  • Hot Water Sanitizable at 85°C / 185°F for 15 min up to 2 hrs.
  • Rapid Heat/Cool Cycle — up to 160 cycles.
  • New High Temperature FDA Urethane Materials.
  • Pharmaceutical Stainless Steel (EXL option).
  • Light, Compact, Easy to Ship and Handle.

Model Flow Rates:

  • EXL-810-HTS: 6.5-8.0 m³/h (28-35 gpm)
  • EXL-710-HTS: 5.0-7.0 m³/h (22-30 gpm)
  • EXL-610-HTS: 3.5-5.5 m³/h (15-24 gpm)
  • EXL-510-HTS: 2.5-4.0 m³/h (10-18 gpm)
  • XL-500-HTS: 1.3-2.5 m³/h (6-11 gpm)
  • XL-400-HTS: 0.7-1.5 m³/h (2.5-7.0 gpm)
  • XL-300-HTS: 300-1000 lph (1.5-4.5 gpm)
  • XL-200-HTS: 100-300 lph (0.5-1.5 gpm)
  • XL-100-HTS: 50-150 lph (0.2-0.7 gpm)

Electropure™ EDI Technology features:

  • No concentrate recirculation or brine injection.
  • Thin-Cell efficient technology.
  • Thin-Concentrate non-scaling technology.
  • Unique non-scaling electrode system.
  • Patented Excellion™ membranes.

Operating Limits

EDI Module Parameters - Recommended for Optimum Performance / Maximum:

  • Feedwater Conductivity Equivalent (Including CO₂): <9 μS/cm / <33 μS/cm
  • pH: 7.0 to 7.5 / 5 to 9.5
  • Temperature:
    • Operation: 5 to 45°C (40-113°F)
    • Sanitize: 72-85°C (162-185°F)
  • Concentrate Flow Direction: Upward
  • Feed Flow Direction: Upward

Feed water

Feedwater Impurity Limits - Recommended for Optimum Performance / Maximum:

  • Total Chlorine (as Cl₂) and Ozone (O₃): ND / ND
  • Iron (Fe): ND / <.01 ppm
  • Manganese (Mn): ND / <.01 ppm
  • Sulfide (S²⁻): ND / <.01 ppm
  • Total Hardness (as CaCO₃): <0.5 ppm / <1 ppm
  • Total Silica (as SiO₂): <0.2 ppm / <0.5 ppm
  • Total Organics (TOC as C): ND / <0.5 ppm

Mechanical / Hydraulic

XL Dimensions (Width x Height x Depth): 22x56x D cm (8.5x22x D in)

Specific XL Model Depths and Weights:

  • XL-100-HTS: Depth: 17 cm (6.5 in), Weight: 24kg/54lb
  • XL-200-HTS: Depth: 19 cm (7.5 in), Weight: 28kg/61lb
  • XL-300-HTS: Depth: 26 cm (10 in), Weight: 33kg/72lb
  • XL-400-HTS: Depth: 29 cm (11.5 in), Weight: 36kg/80lb
  • XL-500-HTS: Depth: 37 cm (14.5 in), Weight: 44kg/95lb