Back to Veolia E-Cell-EDI / EDR
Veolia logo

Datasheet summary

EDI Pro E-Cell EU Series

Engineering summary from PDF text extraction for EDI Pro E-Cell EU Series. Verify every value with the OEM datasheet.

Summary

Product identification

The PRO E-Cell EU Series consists of integrated Reverse Osmosis (RO) and Electrodeionization (EDI) machines, with capacities ranging from 10 to 23 m3/h at 50 Hz.

Key features include:

  • RO and EDI systems mounted on a common skid.
  • Two-pass RO with interpass caustic injection.
  • RO Pump VFDs and EDI Power control/DC Drive mounted on the skid.
  • RO permeate flush sequence on shutdown and divert to drain on startup.
  • Siemens control system with Simatic PLC and HMI interface.
  • CE certified.

Instrumentation

  • Flow Meters: RO Permeates, RO Concentrates, EDI Product, EDI Concentrate, EDI Electrode
  • Pressure Gauge: Pre & Post Cartridge Filter, Primary, Final, Interstage, RO Permeate, RO Concentrate, RO Pump Discharge, EDI Feed, EDI Product, EDI concentrate, EDI Electrode, EDI Dilute Feed
  • Pressure Switch: RO Feeds, RO2 Permeate, RO Concentrates
  • Pressure Transmitter: RO Primary, RO Final, RO Permeate, EDI Product, EDI Concentrate, EDI Electrode, EDI Dilute Feed
  • Conductivity/Temp: RO Feed, RO1 Permeate, EDI Product (resistivity)
  • pH Meter: RO Feed (both passes)
  • ORP: RO Feed pass1

Options

  • HART protocol instruments

Accessories

  • PRO E-CELL EU Clean-in-Place unit
  • PRO Multi-Media filters
  • PRO Activated Carbon and Softeners
  • PRO Chemical Feed Systems
  • Transfer Pumps and Storage Tanks

Operating limits

  • Design Recovery RO: 75% Pass 1; 85% Pass 2
  • Design Recovery EDI: 90-95%
  • Design Temperature: 15ºC
  • Operating Temperature Range: 5-29ºC
  • System Inlet Pressure: 2-4 Bar
  • Recovery Rate can vary +/- 5%.

Feed water

  • RO Inlet Connection:
    • PRO E-Cell EU-10DP: DN50 Flange PN10
    • PRO E-Cell EU-20DP: DN80 Flange PN10
  • Inlet Water Pressure: 2-4 bar

Product water

  • EDI Product Outlet Connection:
    • PRO E-Cell EU-10DP: DN40 Flange PN10
    • PRO E-Cell EU-20DP: DN50 Flange PN10
  • EDI Stack Performance:
    • 16 MOhm-cm
    • Silica Removal: <10 ppb
  • EDI product flow rate varies with conductivity, hardness/silica, and temperature.

Electrical

  • Installed Power (High Pressure Pumps):
    • PRO E-Cell EU-10DP: 26 KW (15kW TEFC Pass 1, 11kW TEFC Pass 2)
    • PRO E-Cell EU-20DP: 60 KW (30 kW TEFC Pass 1, 30 kW TEFC Pass 2)
  • EDI Stacks Power (DC drive, 400 VDC):
    • PRO E-Cell EU-10DP: 16 A max
    • PRO E-Cell EU-20DP: 32 A max
  • Utility Power Requirement: 400VAC, 50hz, 3-Phase

Mechanical / hydraulic

Materials of Construction

  • High-pressure piping: 1.4404 Stainless Steel
  • Low-pressure RO & EDI piping: DIN PVC-U
  • EDI dilute piping: DIN Standard Polypropylene
  • Frame: Epoxy-coated carbon steel
  • Enclosure: IP54
  • Fasteners/Victaulic: Zinc-plated

Membrane Elements and Housings

  • Membrane elements: Veolia AK-400
  • Average membrane flux: 22lmh pass 1, 33lmh pass 2
  • Membrane rejection: 99.0 to 99.5%
  • Housing material: Fiberglass
  • Housing Pressure Rating: 31 bar

EDI Stacks

  • Model: Veolia E-Cell 3X
  • Stack product flow rate, nominal: 5 m3/hr

Cartridge Filtration

  • Housing model: Veolia HX-0740-80-V-316-A-EMEA
  • Housing material: 1.4404 Stainless Steel
  • Cartridge filter: 1-µ nom, ROSave.Zs*, RO.Zs01-40XK

PRO E-Cell Models Specifics

ParameterPRO E-Cell EU-10DPPRO E-Cell EU-20DP
RO Feed Rate19 m3/hr34 m3/hr
RO Concentrate Rate4.8 m3/hr9.5 m3/hr
EDI Product Rate9-12.4 m3/hr18-24.8 m3/hr
EDI Concentrate Rate0.6 m3/hr1.2 m3/hr
EDI Electrode Rate0.2 m3/hr0.4 m3/hr
High Pressure Pumps
ModelCRN 20-14 Pass 1, CRN 15-14 Pass 2CRN 45-7 Pass 1, CRN 45-7 Pass 2
ManufacturerGrundfosGrundfos
Quantity22
Motor Power Pass 115kW TEFC30 kW TEFC
Motor Power Pass 211kW TEFC30 kW TEFC
Membrane Elements and Housings
Membrane Quantity20 Pass 1; 12 Pass 236 Pass 1; 18 Pass 2
Memb. Housing Style4 element long, 4 port6 element long, 4 port
Banking Arrangement3→2 Pass 1; 2→1 Pass 24→2 Pass 1; 2→1 Pass 2
EDI Stacks
Stack ModelVeolia E-Cell 3XVeolia E-Cell 3X
Stack Quantity24
Cartridge Filtration
Housing Quantity12
Cartridge Filter ModelROSave.Zs, RO.Zs01-40XKROSave.Zs, RO.Zs01-40XK
Cartridge Filter Length40-inch (120 cm)40-inch (102 cm)
Cartridge Filter Qty714

Installation and Utility Requirements

ConnectionPRO E-Cell EU-10DPPRO E-Cell EU-20DP
RO Inlet ConnectionDN50 Flange PN10DN80 Flange PN10
EDI Product OutletDN40 Flange PN10DN50 Flange PN10
RO1 Reject to DrainDN25 Flange PN10DN40 Flange PN10
RO2 Product DumpDN40 Flange PN10DN50 Flange PN10
EDI Product DumpDN40 Flange PN10DN50 Flange PN10
EDI Concentrate OutletDN15 Flange PN10DN15 Flange PN10
EDI Electrode OutletDN15 Flange PN10DN15 Flange PN10
Air Pressure5.5 barg, oil-free, 90 slpm5.5 barg, oil-free, 90 slpm
Drain to be Sized for22 m3/hr35 m3/hr

Skid Dimensions and Weights (approx.)

ParameterPRO E-Cell EU-10DPPRO E-Cell EU-20DP
Height2.4m2.4m
Width1.4m1.6m
Length5.8m7.2m
Shipping Weight3,500kg6,100kg
Operating Weight4,500kg7,800kg