Product identification

Product: Picomag electromagnetic flowmeter Product headline: Smart magmeter for utilities – intuitive · convenient · multivariable. Field of application: Economical magmeter for conductive liquids in secondary circuits across various industries.

Operating limits

  • Max. process pressure: 16 bar
  • Medium temperature range: -10 to 70°C (+14 to +158 °F)
  • Permissible short-term temperature (max. one hour): 85 °C (185 °F) (Repetition after 4 hours at the earliest)
  • Ambient temperature range: -10 to 60 °C

Feed water

  • Liquids: Conductive liquids

Product water

  • Measured variables: Volume flow, Temperature, Conductivity
  • Max. measurement error (Volume flow): +/-0.8% o.r. +/- 0.1% o.f.s
  • Measuring range (Volume flow): up to 750 l/min

Electrical

  • Power supply: 18 to 30V DC
  • Outputs: Current (4 to 20mA), Pulse, Switch, Voltage (2 to 10V)
  • Inputs: Status
  • Digital communication: IO-Link

Mechanical / hydraulic

  • Nominal diameter range: DN15 to 50 (1/2 to 2")
  • Wetted materials: PEEK, Stainless Steel, FKM, EPDM
  • Electrodes: 1.4404/316L
  • Sensor housing material: 1.4404/316L, Polycarbonate
  • Transmitter housing material: 1.4404/316L, Polycarbonate
  • Degree of protection: IP 65/67

Features

  • Simultaneous flow, conductivity, and temperature measurement.
  • Space-saving design with sensor and transmitter in one housing.
  • Intuitive operation and configuration via SmartBlue App using Bluetooth.
  • Quick offline configuration check with "knock" functionality.
  • High shock and vibration resistance.
  • Empty pipe detection.
  • Flexible integration with IO-Link and various I/Os.
  • 1.4" TFT color display with backlight.
  • cULus listing.
  • NSF drinking water approvals.