Back to Endress + Hauser Flow measurement
Endress+Hauser logo

Datasheet summary

Proline Cubemass C 500

PDF datasheet mirror for Proline Cubemass C 500 (Endress + Hauser Flow measurement). Verify with OEM before design.

Summary

Product identification

Product: Proline Cubemass C 500 Type: Coriolis flowmeter Headline: Compact sensor for smallest quantities, with a remote transmitter version with up to 4 I/Os. Application: Measurement of smallest flow rates in skids, test rigs, and industrial robotics.

Operating limits

Max. measurement error

  • Mass flow (liquid): ±0.10 %
  • Volume flow (liquid): ±0.10 %
  • Mass flow (gas): ±0.50 %
  • Density (liquid): ±0.0005 g/cm³

Measuring range

  • 0 to 1000 kg/h (0 to 37 lb/min)

Process conditions

  • Max. process pressure: PN 40, Class 300, 10K, 400 bar (5800 psi)
  • Medium temperature range: –50 to +205 °C (–58 to +401 °F)
  • Ambient temperature range (Standard): –40 to +60 °C (–40 to +140 °F)
  • Ambient temperature range (Option): –50 to +60 °C (–58 to +140 °F)

Feed water / Medium

  • Wetted materials (Measuring tube): 1.4539 (904L)
  • Wetted materials (Connection): 1.4539 (904L); 1.4404 (316/316L)

Product water / Measured variables

  • Mass flow
  • Density
  • Temperature
  • Volume flow
  • Corrected volume flow
  • Reference density
  • Concentration

Electrical

Outputs

  • 4 outputs:
    • 4-20 mA HART (active/passive)
    • 4-20 mA WirelessHART
    • 4-20 mA (active/passive)
    • Pulse/frequency/switch output (active/passive)
    • Double pulse output (active/passive)
    • Relay output

Inputs

  • Status input
  • 4-20 mA input

Digital communication

  • HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA

Power supply

  • DC 24 V
  • AC 100 to 230 V
  • AC 100 to 230 V / DC 24 V (non-hazardous area)

Mechanical / Hydraulic

  • Nominal diameter range: DN 1 to 6 (¹⁄₂₄ to ¹⁄₄")
  • Sensor housing material: 1.4301 (304), corrosion resistant
  • Sensor connection housing (standard): AlSi10Mg, coated
  • Sensor connection housing (option): 1.4301 (304); 1.4404 (316/316L)
  • Transmitter housing material: AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L; Polycarbonat
  • Degree of protection (Sensor remote version, standard): IP66/67, Type 4X enclosure
  • Degree of protection (Transmitter remote version): IP66/67, Type 4X enclosure

Other Features

  • Design: Compact single-tube design, lightweight sensor
  • Measurement: Multivariable (flow, density, temperature)
  • Transmitter: Remote version with up to 4 I/Os, backlit display with touch control and WLAN access.
  • Verification: Heartbeat Technology for measurement reliability and compliant verification.
  • Approvals: ATEX, IECEx, cCSAus, NEPSI, INMETRO, KC, CE, C-tick, EAC, CRN.
  • Safety: Functional safety according to IEC 61508 / IEC 61511.
  • Calibration: Performed on accredited calibration facilities (acc. to ISO/IEC 17025).

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.