Datasheet summary
Proline Cubemass C 500
PDF datasheet mirror for Proline Cubemass C 500 (Endress + Hauser Flow measurement). Verify with OEM before design.
Summary
Product identification
Product: Proline Cubemass C 500 Type: Coriolis flowmeter Headline: Compact sensor for smallest quantities, with a remote transmitter version with up to 4 I/Os. Application: Measurement of smallest flow rates in skids, test rigs, and industrial robotics.
Operating limits
Max. measurement error
- Mass flow (liquid): ±0.10 %
- Volume flow (liquid): ±0.10 %
- Mass flow (gas): ±0.50 %
- Density (liquid): ±0.0005 g/cm³
Measuring range
- 0 to 1000 kg/h (0 to 37 lb/min)
Process conditions
- Max. process pressure: PN 40, Class 300, 10K, 400 bar (5800 psi)
- Medium temperature range: –50 to +205 °C (–58 to +401 °F)
- Ambient temperature range (Standard): –40 to +60 °C (–40 to +140 °F)
- Ambient temperature range (Option): –50 to +60 °C (–58 to +140 °F)
Feed water / Medium
- Wetted materials (Measuring tube): 1.4539 (904L)
- Wetted materials (Connection): 1.4539 (904L); 1.4404 (316/316L)
Product water / Measured variables
- Mass flow
- Density
- Temperature
- Volume flow
- Corrected volume flow
- Reference density
- Concentration
Electrical
Outputs
- 4 outputs:
- 4-20 mA HART (active/passive)
- 4-20 mA WirelessHART
- 4-20 mA (active/passive)
- Pulse/frequency/switch output (active/passive)
- Double pulse output (active/passive)
- Relay output
Inputs
- Status input
- 4-20 mA input
Digital communication
- HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
Power supply
- DC 24 V
- AC 100 to 230 V
- AC 100 to 230 V / DC 24 V (non-hazardous area)
Mechanical / Hydraulic
- Nominal diameter range: DN 1 to 6 (¹⁄₂₄ to ¹⁄₄")
- Sensor housing material: 1.4301 (304), corrosion resistant
- Sensor connection housing (standard): AlSi10Mg, coated
- Sensor connection housing (option): 1.4301 (304); 1.4404 (316/316L)
- Transmitter housing material: AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L; Polycarbonat
- Degree of protection (Sensor remote version, standard): IP66/67, Type 4X enclosure
- Degree of protection (Transmitter remote version): IP66/67, Type 4X enclosure
Other Features
- Design: Compact single-tube design, lightweight sensor
- Measurement: Multivariable (flow, density, temperature)
- Transmitter: Remote version with up to 4 I/Os, backlit display with touch control and WLAN access.
- Verification: Heartbeat Technology for measurement reliability and compliant verification.
- Approvals: ATEX, IECEx, cCSAus, NEPSI, INMETRO, KC, CE, C-tick, EAC, CRN.
- Safety: Functional safety according to IEC 61508 / IEC 61511.
- Calibration: Performed on accredited calibration facilities (acc. to ISO/IEC 17025).