Back to Endress + Hauser Flow measurement
Endress+Hauser logo

Datasheet summary

Proline-Promag H 200

PDF datasheet mirror for Proline-Promag H 200 (Endress + Hauser Flow measurement). Verify with OEM before design.

Summary

Product identification

  • Product name: Proline Promag H 200
  • Product line: Endress+Hauser Flow measurement
  • Product type: Electromagnetic flowmeter

Field of application

Sensor preferred for applications with highest requirements in the chemical and life sciences industries. Enables cost-effective and seamless integration into existing infrastructures with genuine loop-powered technology. Offers highest operational safety in hazardous areas thanks to its intrinsically safe design (Ex ia).

Features and specifications

  • Measuring principle: Electromagnetic
  • Product headline: Flowmeter for smallest flow rates with genuine loop-powered technology. Dedicated to the measurement of the smallest flow quantities.
  • Sensor features: Flexible installation concept with numerous hygienic process connections. Energy-saving flow measurement with no pressure loss due to cross-section constriction. Maintenance-free due to no moving parts. Liner made of PFA. Sensor housing made of stainless steel. Various electrode materials available.
  • Transmitter features: Convenient device wiring with separate connection compartment. Safe operation via display with touch control and background lighting, no need to open the device. Integrated verification with Heartbeat Technology. Loop-powered technology. Robust dual-compartment housing. Plant safety with worldwide approvals (SIL, Haz. area).

Operating limits

  • Max. measurement error (Volume flow): ±0.5 % o.r. ± 2 mm/s (0.08 in/s)
  • Measuring range: 0.06 dm³/min to 300 m³/h (0.015 to 80 gal/min)
  • Max. process pressure: PN 40, Class 150, 20K
  • Medium temperature range: –20 to +150 °C (–4 to +302 °F)
  • Ambient temperature range: –40 to +60 °C (–40 to +140 °F)

Wetted materials

  • Liner: PFA
  • Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum
  • Process Connections: stainless steel, 1.4404 (F316L); PVDF; PVC adhesive sleeve
  • Seals: O-ring seal (EPDM, FKM, Kalrez), aseptic molded seal (EPDM, FKM, silicone)
  • Grounding Rings: stainless steel, 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); tantalum

Mechanical / hydraulic

  • Nominal diameter range: DN 2 to 25 (¹⁄₁₂ to 1")
  • Sensor housing material: 1.4301 (304), corrosion resistant
  • Transmitter housing material: AlSi10Mg, coated
  • Degree of protection: IP66/67, type 4X enclosure

Electrical

  • Display/Operation: 4-line backlit display with touch control (operation from outside). Configuration via local display and operating tools possible. Remote display available.
  • Outputs: 4-20 mA HART (passive); Pulse/frequency/switch output (passive)
  • Inputs: None
  • Digital communication: HART, PROFIBUS PA, FOUNDATION Fieldbus
  • Power supply: DC 18 to 35 V (4-20 mA HART with/without pulse/frequency/switch output)

Approvals and certifications

  • Hazardous area approvals: ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC, JPN, UK Ex
  • Product safety: CE, C-Tick
  • Functional safety: Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
  • Metrological approvals and certificates: Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025). Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation).
  • Pressure approvals and certificates: CRN
  • Material certificates: 3.1 material

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.