Datasheet summary
Proline-Promag W 500
PDF datasheet mirror for Proline-Promag W 500 (Endress + Hauser Flow measurement). Verify with OEM before design.
Summary
Product identification
- Product Name: Proline Promag W 500
- Model: 5W5B
- Type: Electromagnetic flowmeter
- Headline: Specialist for demanding water and wastewater applications as remote version with up to 4 I/Os.
- Field of Application: Premium device for water and wastewater measurement, designed for reliable use in hazardous areas and under harsh conditions.
Operating limits
- Max. measurement error (Volume flow, standard): ±0.5 % o.r. ± 1 mm/s
- Max. measurement error (Volume flow, option - Flat Spec): ±0.2 % o.r. ± 2 mm/s
- Max. process pressure: PN 40, Class 300, 20K
- Ambient temperature range (Flange material carbon steel): –10 to +60 °C
- Ambient temperature range (Flange material stainless steel): –40 to +60 °C
Feed water
- Measuring principle: Electromagnetic
- Dedicated to: Measurement of industrial or municipal water and wastewater.
Product water
- International drinking water approvals: Yes
General specifications
- Nominal diameter range: DN 25 to 3000 (1 to 120")
- Measuring range (Volume flow): 0.5 m³/h to 263,000 m³/h
- Measured variables: Volume flow, conductivity, mass flow
Wetted materials
- Liner material hard rubber: 0 to +80 °C
- Liner material polyurethane: –20 to +50 °C
- Liner material PTFE: –20 to +90 °C
- Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum
Medium temperature range
- Liner material hard rubber: 0 to +80 °C
- Liner material polyurethane: –20 to +50 °C
- Liner material PTFE: –20 to +90 °C
Electrical
- Power supply:
- DC 24 V
- AC 100 to 230 V
- AC 100 to 230 V / DC 24 V (non-hazardous area)
- Outputs:
- 4-20 mA HART (active/passive)
- 4-20 mA (active/passive)
- Pulse/frequency/switch output (active/passive)
- Double pulse output (active/passive)
- Relay output
- Inputs:
- Status input
- 4-20 mA input
- Digital communication: HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
Mechanical / hydraulic
- Sensor housing material:
- DN 25 to 300 (1 to 12"): AlSi10Mg, coated
- DN 25 to 2400 (1 to 90"): Carbon steel with protective varnish
- Sensor connection housing (standard): AlSi10Mg, coated
- Sensor connection housing (option): Polycarbonate; 1.4409 (CF3M) similar to 316L
- Transmitter housing material: AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L; Polycarbonat
- Degree of protection:
- Compact version: IP66/67, type 4X enclosure
- Sensor remote version (standard): IP66/67, type 4X enclosure
- Sensor remote version (option): IP68, type 6P enclosure, with protective varnish according to EN ISO 12944 C5-M/Im1/Im2/Im3
- Display/Operation: 4-line backlit display with touch control (operation from outside). Configuration via local display and operating tools possible.
Certifications and Approvals
- Hazardous area approvals: ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC; JPN, UK Ex, KC
- Product safety: CE, C-tick, EAC marking
- Functional safety: IEC 61508, IEC 61511
- Metrological approvals: Calibration on accredited facilities (acc. to ISO/IEC 17025), NAMUR, Heartbeat Technology complies with ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
- Marine approvals: LR approval, DNV approval, ABS approval, BV approval
- Pressure approvals: CRN
- Material certificates: 3.1 material
- Hygienic approvals: ACS, KTW/W270, NSF 61, WRAS BS 6920