Product identification
- Product name: Proline Promass A 200
- Product type: Coriolis flowmeter
- Description: Genuine loop-powered flowmeter for accurate measurement of lowest flow rates. Ideal for continuous process control in demanding applications.
Field of application
The compact Promass A is ideal for continuous process control in demanding applications. Its unique self-drainable single-tube system enables accurate measurement of liquids and gases at lowest flow rates and high pressure. With its cost-effective genuine loop-powered technology and intrinsically safe design (Ex ia), Promass A 200 offers highest operational safety in hazardous areas. Heartbeat Technology enables safe processes.
Suitable for applications with smallest flow quantities in the chemical industry.
Sensor features
- Space-saving installation – compact, lightweight sensor.
- Highest product quality – self-drainable measuring tube design in all line sizes.
- Optimum process safety – resistant to corrosive ambient conditions and internal clogging.
Transmitter features
- Convenient device wiring – separate connection compartment.
- Safe operation – no need to open the device due to display with touch control, background lighting.
- Integrated verification – Heartbeat Technology.
- Loop-powered technology.
- Robust dual-compartment housing.
- Plant safety: worldwide approvals (SIL, Haz. area).
Operating limits
- Nominal diameter range: DN 1 to 4 (¹⁄₂₄ to ¹⁄₈")
- Max. process pressure: 430.9 bar (6250 psi)
- Medium temperature range: -50 to 205 °C (-58 to 401 °F)
- Ambient temperature range: -40 to 60 °C (-40 to +140 °F)
Feed water
(No specific feed water parameters are listed in the extract)
Product water
(No specific product water parameters are listed in the extract)
Performance
- Max. measurement error (Mass flow, liquid): ±0.1 %
- Max. measurement error (Volume flow, liquid): ±0.1 %
- Max. measurement error (Mass flow, gas): ±0.35 %
- Max. measurement error (Density, liquid): ±0.0005 g/cm³
- Measuring range: 0 to 450 kg/h (0 to 16.54 lb/min)
Electrical
- Power supply:
- DC 18 to 35 V (4‐20 mA HART with/without pulse/frequency/switch)
- DC 18 to 30 V (20 mA HART, 4‐20 mA)
- DC 9 to 32 V (PROFIBUS PA)
- Outputs:
- 4‐20 mA HART (passive)
- 4‐20 mA (passive)
- Pulse/frequency/switch output (passive)
- Inputs: None
- Digital communication: HART, PROFIBUS PA, FOUNDATION Fieldbus
Mechanical / hydraulic
- Wetted materials: Measuring tube: stainless steel, 1.4435 (316/316L); Alloy C22
- Sensor housing material: Stainless steel, 1.4404 (316L)
- Transmitter housing material: Stainless steel, CF-3M (316L, 1.4404); AlSi10MG, coated
- Degree of protection: IP66/67, Type 4X enclosure
- Display/Operation: 4-line backlit display with touch control (operation from outside)
Approvals and certificates
- Hazardous area approvals: ATEX, IECEx, cCSAus, INMETRO, NEPSI, UK Ex
- Product safety: CE, C-TICK
- Functional safety: Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
- Metrological: Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025); Heartbeat Technology complies with requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
- Pressure approvals: PED, CRN
- Material certificates: 3.1 material
- Hygienic approvals: 3-A, cGMP