Proline Promass H 100

PDF datasheet mirror for Proline Promass H 100 (Endress + Hauser Flow measurement). Verify with OEM before design.

Summary

Product identification

  • Product: Proline Promass H 100
  • Model: 8H1B
  • Type: Coriolis flowmeter
  • Headline: Chemically resistant single-tube flowmeter with an ultra-compact transmitter.

Field of application

Designed for applications requiring highest corrosion resistance with chemically aggressive fluids. Suitable for system integrators, skid builders, and equipment manufacturers where space is a premium.

Benefits

  • Maximum safety for chemically aggressive fluids – corrosion-resistant wetted parts.
  • Fewer process measuring points – multivariable measurement (flow, density, temperature).
  • Space-saving installation – no in/outlet run needs.
  • Space-saving transmitter – full functionality on the smallest footprint.
  • Time-saving local operation without additional software and hardware – integrated web server.
  • Integrated verification – Heartbeat Technology.

Operating limits

  • Max. process pressure: PN 40, Class 300, 20K
  • Medium temperature range:
    • Tantalum: –50 to +150 °C (–58 to +302 °F)
    • Zirconium: –50 to +205 °C (–58 to +401 °F)
  • Ambient temperature range:
    • Standard: –40 to +60 °C (–40 to +140 °F)
    • Option: –50 to +60 °C (–58 to +140 °F)

Feed water

  • Measuring principle: Coriolis
  • Measured variables: Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration.

Product water

  • Max. measurement error:
    • Mass flow (liquid): ±0.1 %
    • Volume flow (liquid): ±0.1 %
    • Mass flow (gas, Tantalum only): ±0.5 %
    • Density (liquid): ±0.0005 g/cm³
  • Measuring range: 0 to 70 000 kg/h (0 to 2570 lb/min)

Mechanical / hydraulic

  • Nominal diameter range: DN 8 to 50 (³⁄₈ to 2")
  • Wetted materials:
    • Measuring tube: Tantalum 2.5W; 702 (UNS R60702)
    • Connection: Tantalum; 702 (UNS R60702)
  • Sensor housing material: 1.4301 (304), corrosion resistant
  • Transmitter housing material:
    • Compact: AlSi10Mg, coated
    • Compact/ultra-compact: 1.4301 (304)
  • Degree of protection:
    • Standard: IP66/67, type 4X enclosure
    • Option: IP69

Electrical

  • Outputs:
    • 4–20 mA HART (active)
    • Pulse/frequency/switch output (passive)
  • Inputs: None
  • Digital communication: HART, Modbus RS485, EtherNet/IP, PROFIBUS DP, PROFINET
  • Power supply: DC 20 to 30 V

Approvals and Certifications

  • Hazardous area approvals: ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC
  • Product safety: CE, C-Tick, EAC marking
  • Metrological approvals and certificates: Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR. Heartbeat Technology complies with the requirements for traceable verification according to ISO 9001:2008 – Section 7.6a (TÜV SÜD attestation).
  • Pressure approvals and certificates: PED, CRN
  • Material certificates: 3.1 material

Display/Operation

  • 4-line backlit display available (no local operation)
  • Configuration via web browser and operating tools possible