Back to Endress + Hauser Flow measurement
Endress+Hauser logo

Datasheet summary

Prosonic Flow Clamp On 91W

PDF datasheet mirror for Prosonic Flow Clamp On 91W (Endress + Hauser Flow measurement). Verify with OEM before design.

Summary

Product identification

  • Product line: Proline Prosonic Flow 91W
  • Measuring principle: Ultrasonic flow
  • Product headline: Device with automatic frequency scan for maximum measuring performance and cost-effective transmitter.
  • Description: Clamp-on flow measurement of process water, salt water, demineralized water, drinking & wastewater.

Operating limits

  • Medium temperature range:
    • -20 to +80°C (-4 to +176 °F)
    • 0 to +130°C (32 to +265 °F) option
  • Ambient temperature range: -20 to +60 °C (-4 to +140 °F)
  • Max. process pressure: N/A

Feed water

  • Wetted materials (Clamp on system):
    • Sensor holder: 1.4308/CF-8
    • Sensor housing: 1.4301/304
    • Strapping bands: 1.4301/304

Product water

  • Nominal diameter range: Single channel, 1 or 2 paths: DN15 to 2000 (1/2 to 80")

Performance

  • Max. measurement error (Volume flow):
    • ±3% o.r. for DN15
    • ±2% o.r. for DN25 to 200
    • ±2% o.r. above DN200
  • Measuring range: 0.3 to 10 m/s (1 to 33 ft/s)
  • Measured variables: Volume flow, sound velocity, flow velocity, signal strength, totalizer

Electrical

  • Outputs:
    • 1x 4-20 mA HART (active)
    • 1x Pulse/frequency/switch output (passive)
  • Digital communication: HART
  • Power supply:
    • AC 85 to 250 V
    • DC 20 to 28 V
    • DC 11 to 40 V

Mechanical / hydraulic

  • Degree of protection:
    • IP67, type 4X for transmitter
    • IP67 type 4X for sensors
    • IP68 type 6P for sensors (option). IP67 NEMA 4X
  • Transmitter housing material: AlSi10Mg, coated
  • Display/Operation: 2 lines backlit display with 3 push buttons

Approvals and certificates

  • Hazardous area approvals: Non hazardous area, FM, CSA
  • Other approvals and certificates: Flowmeter verification for DN15, 25, 40, 50 & 100 only, EAC marking
  • Product safety: CE, C-Tick, EAC marking
  • Metrological approvals and certificates: Flowmeter verification for DN15, 25, 40, 50 & 100 only