Datasheet summary
Liquiphant FTL51B
Engineering summary from PDF text extraction for Liquiphant FTL51B. Verify every value with the OEM datasheet.
Summary
Product identification
- Product name: Liquiphant FTL51B HART
- Type: Vibronic Point level switch for liquids
- Order code: (See nameplate or delivery note)
- Serial number: (See nameplate)
- Document version: Valid as of version 01.00.zz (Device firmware)
- Nameplate information: Includes manufacturer identification, order number, extended order code, serial number, technical data, degree of protection, firmware/hardware version, approval-related information, and DataMatrix code.
- Manufacturer: Endress+Hauser SE+Co. KG, Hauptstraße 1, 79689 Maulburg, Germany.
Storage and transport
- Storage temperature: –40 to +80 °C (–40 to +176 °F). Optional –50 °C (–58 °F) or –60 °C (–76 °F). With M12 plug, elbowed: –25 to +80 °C (–13 to +176 °F).
- Transport: Use original packaging. Hold by the housing, temperature spacer, flange, or extension pipe. Do not bend, shorten, or extend the tuning fork.
Mounting
- Orientation: Any orientation for short pipes (up to approx. 500 mm / 19.7 in). Vertical from above for long pipes.
- Clearance: Minimum distance between fork tip and tank/pipe wall: 10 mm (0.39 in).
- Viscosity:
- Low viscosity (< 2,000 mPa⋅s): Tuning fork can be within the installation socket.
- High viscosity (≤ 10,000 mPa⋅s): Tuning fork must be outside the installation socket.
- Buildup: Use short installation sockets and maintain distance from expected buildup.
- Support: Required for severe dynamic loads. Maximum lateral loading capacity of pipe extensions and sensors: 75 Nm (55 lbf ft). For marine approval, support is needed at least every 1,600 mm for extensions/sensors longer than 1,600 mm.
- Weld-in adapter: Leakage hole must point downwards.
Installation
- Tools: Open-ended wrench, Allen key.
- Alignment: Use markings on the hexagonal nut or weld-in adapter, or the II symbol on the flange/Tri-Clamp.
- Pipes: Suitable for flow velocity up to 5 m/s with 1 mPa⋅s viscosity and 1 g/cm³ density.
- Screwing in: Turn only by the hex bolt; 15 to 30 Nm (11 to 22 lbf ft). Do not turn the housing.
- Cable entry alignment: Housing can be rotated up to 380° by loosening the securing screw (max 1.5 turns).
- Housing cover: Do not lubricate the thread; it is coated with lubricant varnish.
- Display module rotation: Switch off supply voltage before rotating. Can be rotated up to 4 × 90° in each direction.
Electrical connection
- Tools: Screwdriver, Allen key.
- Cover screw: For hazardous area devices, ensure correct positioning of the securing screw.
- Potential equalization: Connect to the outer ground terminal if necessary. Keep the line short; minimum cross-section 2.5 mm² (14 AWG).
- Supply voltage:
- 10.5 to 35 VDC (non-Ex, Ex d, Ex e)
- 10.5 to 30 VDC (Ex i)
- Terminals:
- Supply voltage and internal ground: 0.5 to 2.5 mm² (20 to 14 AWG)
- External ground: 0.5 to 4 mm² (20 to 12 AWG)
- Cable outer diameter:
- Plastic entry: ⌀5 to 10 mm (0.2 to 0.38 in)
- Nickel-plated brass entry: ⌀7 to 10.5 mm (0.28 to 0.41 in)
- Stainless steel entry: ⌀7 to 12 mm (0.28 to 0.47 in)
- 4-20 mA HART: A 250 Ω HART communication resistor is required for low-impedance power supply. Maximum 6 V voltage drop across the resistor.
- Overvoltage protection:
- Without optional protection: Meets IEC/DIN EN 61326-1, test level 1000 V line to earth. Overvoltage category II, pollution degree 2.
- With optional protection: Spark-over voltage min. 400 VDC, nominal discharge current 10 kA.
- Wiring: Switch off supply voltage before connecting. Use specified cable glands for hazardous areas. Ensure cables are adequately insulated and temperature-stable.
- Terminal assignment:
- Single compartment housing: Terminals 1 (+), 2 (-), 3 (internal ground).
- Dual-compartment housing (L-shape): Terminals 1 (+), 2 (-), 3 (internal ground).
- Cable entries: Type depends on device version.
- M12 plug: Pin assignment: 1 (Signal +), 2 (Not assigned), 3 (Signal –), 4 (Ground).
- Degree of protection:
- Housing: IP66/IP68 NEMA 4X/6P.
- M12 plug (connected): IP66/67 NEMA Type 4X.
- M12 plug (unconnected/open housing): IP20, NEMA Type 1.
Post-connection check
- Verify device and cables for damage.
- Check cable requirements, strain relief, and tightness of glands.
- Confirm supply voltage and terminal assignment.
- Ensure all housing covers are installed and tightened.
Operation options
- Methods: Operating keys/DIP switches on electronic insert, optical keys on display (optional), Bluetooth® wireless technology (optional) with Smartblue app or FieldXpert/DeviceCare, operating tool (FieldCare/DeviceCare, handheld terminal, AMS, PDM).
- Electronic insert:
- Reset password key.
- Device reset keys (as-delivered state).
- Proof test wizard key (> 3 s).
- DIP switch for safety function (MIN/MAX).
- DIP switch for locking/unlocking.
- Safety function (DIP switch):
- MAX: Output switches to demand mode when covered (e.g., overfill protection).
- MIN: Output switches to demand mode when uncovered (e.g., dry run prevention).
- DIP switch settings have priority.
- Function test (operating key): Press key for > 3 s. Output changes to demand state; returns to normal if internal test passes. Can also be initiated via digital interfaces.
- Operating menu structure: Local display for simple configuration; operating tools for advanced applications. Wizards guide users through configuration.
- User roles: Operator and Maintenance with different write access levels, protected by an access code.
- Local display (optional): Operates through cover. Displays measured values, faults, and notices. Backlighting changes color (green to red) on error. Available with Bluetooth® option.
- Bluetooth® wireless technology (optional): Requires device display with Bluetooth, smartphone/tablet with Smartblue app or PC with DeviceCare/FieldXpert. Range up to 25 m (82 ft). Operating keys on display locked when connected. Initial password is the device serial number.
- Operating tool access: Via HART (e.g., Commubox FXA195) or CDI connection (Commubox FXA291).
- HistoROM data management: Stores data and device serial number. Transfers data when replacing electronic insert.
System integration
- Manufacturer ID: 17 (0x11)
- Device type code: 0x11C4
- Device revision: 1
- HART specification: 7
- DD revision: 1
- HART load: Minimum 250 Ω.
- Measured variables via HART:
- Primary variable (PV): Level limit detection (initial state depends on fork state and safety function).
- Secondary variable (SV): Sensor frequency (oscillation frequency).
- Tertiary variable (TV): Fork state (covered/uncovered).
- Quaternary variable (QV): Sensor temperature.
- Device variables codes: Level limit detection (0), Sensor frequency (1), Fork state (2), Sensor temperature (3), Terminal current (5), Terminal voltage (6).
- System units: Oscillation frequency in Hz. Temperature in °C, °F, or K.
Commissioning
- Preliminaries: Ensure correct current output settings to prevent overflow or dry running. Reconfigure range settings (LRV/URV) after changing current output settings.
- As-delivered state (default settings):
- Assign PV: Level limit detection (8/16 mA mode).
- Safety mode: MAX.
- Alarm condition: Min. 3.6 mA.
- DIP switch for locking: OFF.
- Bluetooth: ON.
- Density range: > 0.7 g/cm³.
- Switching times: 0.5 s when fork is covered.
Technical data
Input
- Measured variable: Level limit detection.
- Measuring range: See official PDF for exact tabulated values.
Output
- Output signal: 4 to 20 mA HART.
- Signal on alarm: See official PDF for exact tabulated values.
- Load: See official PDF for exact tabulated values.
- Damping: See official PDF for exact tabulated values.
- Switch output: See official PDF for exact tabulated values.
- Protocol-specific data: See official PDF for exact tabulated values.
- Wireless HART data: See official PDF for exact tabulated values.
- Heartbeat Technology: See official PDF for exact tabulated values.
Environment
- Ambient temperature range: See official PDF for exact tabulated values.
- Storage temperature: –40 to +80 °C (–40 to +176 °F).
- Humidity: See official PDF for exact tabulated values.
- Operating height: See official PDF for exact tabulated values.
- Climate class: See official PDF for exact tabulated values.
- Degree of protection: IP66/IP68 (NEMA 4X/6P).
- Vibration resistance: See official PDF for exact tabulated values.
- Shock resistance: See official PDF for exact tabulated values.
- Mechanical load: See official PDF for exact tabulated values.
- Pollution degree: 2.
- Electromagnetic compatibility (EMC): See official PDF for exact tabulated values.
Process
- Process temperature range: See official PDF for exact tabulated values.
- Thermal shock: See official PDF for exact tabulated values.
- Process pressure range: See official PDF for exact tabulated values.
- Overpressure limit: See official PDF for exact tabulated values.
- State of aggregation: Liquids.
- Medium density: See official PDF for exact tabulated values.
- Viscosity:
- Low viscosity: < 2,000 mPa⋅s
- High viscosity: ≤ 10,000 mPa⋅s
- Pressure tightness: See official PDF for exact tabulated values.
- Solids contents: See official PDF for exact tabulated values.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.