Back to Endress + Hauser Level measurement
Endress+Hauser logo

Datasheet summary

Liquiphant M FTL50

Engineering summary from PDF text extraction for Liquiphant M FTL50. Verify every value with the OEM datasheet.

Summary

Product Identification

  • Product Line: Liquiphant M
  • Models: FTL50, FTL51, FTL50H, FTL51H
  • Type: Vibronic point level switch for all kinds of liquids.
  • Certifications: FTL50H and FTL51H are certified for food and pharmaceutical industries.

Application

  • Point level detection (maximum or minimum) in tanks or pipes.
  • Suitable for use in hazardous areas, food, and pharmaceutical applications.
  • Ideal substitute for float switches, unaffected by flow, turbulence, bubbles, foam, vibration, solids content, or buildup.

Operating Limits

  • Process Temperatures: –50 °C to 150 °C (–58 °F to 302 °F)
  • Pressures: Up to 100 bar (1450 psi)
  • Viscosities: Up to 10,000 mm²/s (cSt)
  • Liquid Densities: ≥ 0.5 g/cm³ (SGU)

Model Variants

  • FTL50: Compact design.
  • FTL51: With extension pipe, standard up to 3 m (9.8 ft), and up to 6 m (20 ft) on request.
  • FTL50H/FTL51H: Hygienic versions with polished tuning fork and hygienic process connections.

Features and Benefits

  • Recommended for safety systems requiring functional safety to SIL2/SIL3 per IEC 61508/IEC 61511-1.
  • Design in accordance with ASME B31.3.
  • Recommended for sterile applications in the life science industry (design in accordance with ASME BPE).
  • No adjustment required for quick, low-cost startup.
  • No mechanically moving parts: no maintenance, no wear, long operating life.
  • Monitoring of tuning fork for damage.
  • Compact stainless steel housing available with IP69 protection rating.

Electronic Inserts (Summary)

The Liquiphant M system is modular with various electronic inserts offering different functionalities:

  • FEL51: 2-wire AC version, switches load directly into power supply circuit.
  • FEL52: 3-wire DC PNP version, switches load via transistor for PLC integration.
  • FEL54: Universal AC/DC version with relay output (2 floating change-over contacts).
  • FEL55: 8/16 mA signal transmission for PLC AI modules.
  • FEL56: NAMUR L-H edge signal (0.6 to 1.0 / 2.2 to 2.8 mA) for separate switching units.
  • FEL58: NAMUR H-L edge signal (2.2 to 3.5 / 0.6 to 1.0 mA) for separate switching units.
  • FEL57: PFM signal transmission for Endress+Hauser switching units (Nivotester) with proof test capability.
  • FEL50A: PROFIBUS PA communication.
  • FEL50D: For connecting to Density Computer FML621 for density measurement.

Performance Characteristics

  • Maximum Measured Error: +/- 1 mm (0.04 in) under reference conditions.
  • Repeatability: 0.1 mm (0.004 in).
  • Hysteresis: Approx. 2 mm (0.08 in).
  • Influence of Medium Temperature: Max. +1.8 to -2.8 mm.
  • Influence of Medium Density: Max. +4.8 to -3.5 mm (for 0.5 to 1.5 g/cm³).
  • Influence of Medium Pressure: Max. 0 to -2.5 mm (for 1 to 64 bar).

Operating Conditions

  • Ambient Temperature Range: -50 °C to +150 °C (depending on housing and electronic insert).
  • Storage Temperature: -50 °C to +80 °C.
  • Degree of Protection: Up to IP69 (compact stainless steel housing).
  • Vibration Resistance: See official PDF for details.
  • Electromagnetic Compatibility: See official PDF for details.

Mechanical Construction

  • Dimensions: Refer to the official PDF for detailed dimensions.
  • Materials: Stainless steel housing options available.
  • Process Connections: Various options available, including hygienic connections for H models.

Disclaimer: This summary is based on extracted text and may not include all figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.