Back to Endress + Hauser Level measurement
Endress+Hauser logo

Datasheet summary

Micropilot FMR43

Engineering summary from PDF text extraction for Micropilot FMR43. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product Name: Micropilot FMR43 HART
  • Technology: Free-space radar
  • Document: TI01728F/00/EN/04.24-00

Application

  • Continuous, non-contact level measurement of liquids and solids in hygienic applications.
  • Process connections available for hygiene applications (e.g., Tri-Clamp or M24 adapter).
  • Suitable for CIP and SIP capabilities with protection class IP69.

Operating limits

  • Maximum measuring range: 15 m (49 ft)
  • Process temperature: –40 to +150 °C (–40 to +302 °F)
  • Pressure: –1 to +20 bar (–14.5 to +290 psi)

Performance characteristics

  • Accuracy: ±1 mm (±0.04 in)
  • Reference conditions:
    • As per IEC 62828-2
    • Ambient temperature TA = +21 to +33 °C (+70 to +91 °F)
    • Humidity = 5 to 80 % RH ± 5 %
    • Atmospheric pressure = 860 to 1060 mbar (12.47 to 15.37 psi)
    • Load with HART: 250 Ω
    • Supply voltage: 24 VDC ±3 VDC
    • Reflector: metal plate with diameter ≥ 1 m (40 in)
  • Maximum measurement error:
    • Liquids:
      • Up to 0.15 m: max. ±4 mm (±0.16 in)
      • 0.15 m: ±1 mm (±0.04 in)

    • Solids:
      • Up to 0.4 m: max. ±20 mm (±0.79 in)
      • 0.4 m: ±3 mm (±0.12 in)

  • Non-repeatability: ≤ 1 mm (0.04 in)
  • Warm-up time: ≤ 30 s (according to IEC 62828-4)

Feed water

  • Dielectric constant (εr):
    • For liquids: ≥ 1.2
    • For bulk solids: ≥ 1.6

Product water

Not specified in the extract.

Electrical

  • Output signal: 4 to 20 mA with superimposed digital communication protocol HART, 2-wire.
  • Power supply: 12 to 30 VDC on a direct current power unit.
  • Power consumption:
    • Non-hazardous area: Max. current limited to 500 mA.
    • Hazardous area: Max. current restricted to 100 mA by the transmitter power supply unit.
  • HART load: Min. 250 Ω.

Mechanical / hydraulic

  • Antenna materials: PTFE or PEEK for hygienic requirements.
  • Process connections: Various options including Tri-Clamp and M24 adapter concepts.
  • Housing weight: 0.2 kg (0.44 lb) (including electronics and onsite display).
  • Weight of process connection and antenna: Varies by type (e.g., Process connection MNPT/G ½, antenna 180 GHz; PTFE: 0.140 kg (0.31 lb)).
  • Materials in contact with process: Specific details provided for various process connections and seals (FKM, EPDM, PTFE cladding, PEEK cladding).
  • Surface roughness: Not specified in the extract.

Environment

  • Ambient temperature range:
    • For MNPT/G ½, M24 180 GHz, Tri-Clamp, Neumo Bio Control: –40 to +80 °C (–40 to +176 °F)
    • For MNPT/G ¾, MNPT/G 1½, G1, M24 80 GHz: –40 to +80 °C (–40 to +176 °F)
    • Permitted ambient temperature is reduced at higher process temperatures.
  • Storage temperature: –40 to +85 °C (–40 to +185 °F)
  • Operating height: Up to 5000 m (16404 ft) above sea level.
  • Degree of protection: IP66/68/69, NEMA type 4X/6P for mounted M12 connecting cable.
  • Vibration resistance: Guaranteed for 5 to 2000 Hz: 1.25 (m/s²)²/Hz, ~ 5 g.
  • Shock resistance: 30 g (18 ms) in all 3 axes.

Disclaimer