Datasheet summary
Micropilot M FMR56
Engineering summary from PDF text extraction for Micropilot M FMR56. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Line: Level measurement
- Product Name: Micropilot M FMR56, FMR57
- Type: Free-space radar
Application
- Continuous, non-contact level measurement in powdery to granular bulk solids.
- PP-cladded horn antenna (FMR56); horn or parabolic antenna (FMR57).
- International explosion protection certificates.
- Linearity protocol (3-point, 5-point).
Operating Limits
- Process Temperature: –40 to +400 °C (–40 to 752 °F)
- Process Pressure: –1 to +16 bar (–14.5 to +232 psi)
Input
- Measured Variable: Distance from the reference point (process connection) to the product surface. Level is calculated based on the entered empty distance "E".
- Measuring Range:
- FMR56: Maximum 30 m (98 ft)
- FMR57: Maximum 70 m (230 ft)
- Operating Frequency: K-band (~ 26 GHz)
- Transmission Power:
- 1 m (3.3 ft): < 64 nW/cm²
- 5 m (16 ft): < 2.5 nW/cm²
Output
-
Output Signal:
- HART: FSK ±0.5 mA over current signal, 1200 Bit/s, Galvanic isolation: Yes.
- Bluetooth® wireless technology: Via SmartBlue app. Range > 10 m (33 ft). Encrypted communication.
- PROFIBUS PA: Manchester Bus Powered (MBP), 31.25 kBit/s, Voltage mode, Galvanic isolation: Yes.
- FOUNDATION Fieldbus: Manchester Bus Powered (MBP), 31.25 kBit/s, Voltage mode, Galvanic isolation: Yes.
- Switch output (optional for HART): Open collector, binary switching, failure mode: Non-conductive. Electrical connection data: U = 16 to 35 VDC, I = 0 to 40 mA. Internal resistor Rᵢ < 880 Ω. Insulation voltage: 1350 VDC (relative to power supply), 500 VAC (to ground).
-
Signal on Alarm:
- Current output: Configurable failure mode (NAMUR NE 43: Minimum alarm 3.6 mA, Maximum alarm 22 mA, or user-configurable 3.59 to 22.5 mA).
- Local display: Status signal (NAMUR NE 107), plain text display.
- Operating tool: Status signal (NAMUR NE 107), plain text display.
-
Linearization: Up to 32 points, manual or semi-automatic entry, for converting measured values to length or volume units. Preprogrammed tables for cylindrical and horizontal cylindrical vessels.
Power Supply
- 2-wire, 4-20mA HART, passive: Terminal voltage 10.4 to 35 VDC (depending on approval and ambient temperature). Maximum load R depends on supply voltage U₀.
- 4-wire, 4-20mA HART, active:
- 90 to 253 VAC (50 to 60 Hz), Category II. Max load 500 Ω.
- 10.4 to 48 VDC. Max load 500 Ω.
- PROFIBUS PA, FOUNDATION Fieldbus:
- Non-Ex: 9 to 32 VDC.
- Ex ia / IS: 9 to 30 VDC.
- Power Consumption: Varies by model (e.g., < 0.9 W for 2-wire HART, 6 VA for 4-wire 90-253VAC).
- Current Consumption:
- HART: 3.6 to 22 mA nominal.
- PROFIBUS PA: 14 mA nominal.
- FOUNDATION Fieldbus: 15 mA nominal.
Mechanical Construction
- Dimensions: See official PDF.
- Weight: See official PDF.
- Materials: Housing options include stainless steel (GT18), plastic (GT19), and die-cast aluminum (GT20). Antenna and process connection materials vary.
Performance Characteristics
- Reference Operating Conditions: Temperature +24 °C (±5 °C), Pressure 960 mbar abs. (±100 mbar), Humidity 60% (±15%).
- Maximum Measured Error:
- Digital output: ± 3 mm (± 0.12 in) sum of non-linearity, non-repeatability, hysteresis; ± 4 mm (± 0.2 in) offset/zero.
- Analog output: ± 0.02 % span (non-linearity, etc.); ± 0.03 % span (offset/zero).
- Measured Value Resolution:
- Digital: 1 mm
- Analog: 1 μA
- Response Time: Varies with tank height and sampling rate, typically < 0.8 s for < 10 m tank height.
- Influence of Ambient Temperature:
- Digital: Average TC = 3 mm/10 K.
- Analog: Zero point average TC = 0.02 %/10 K; Span average TC = 0.05 %/10 K.
Certificates and Approvals
- CE mark, RoHS, RCM-Tick, Ex approval, SIL2/SIL3, WHG, Marine approval, Radio standards (EN 302729, FCC, Industry Canada, Japanese), CRN approval.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.