Summary

Product identification

  • Product Name: Micropilot FMR62B
  • Protocol: HART
  • Technology: Free-space radar
  • Document: TI01684F/00/EN/05.24-00

Application

  • Continuous, non-contact level measurement of liquids, pastes, and sludges.
  • Process connections: Flanges.

Operating limits

  • Maximum measuring range: 80 m (262 ft)
  • Temperature: –196 to +450 °C (–321 to +842 °F)
  • Pressure: –1 to +160 bar (–14.5 to +2 321 psi)

Performance characteristics

  • Accuracy: ±1 mm (±0.04 in)
  • Measured value resolution: 1 mm (digital)
  • Maximum measured error:
    • Up to 0.8 m (2.62 ft) measuring distance: max. ±4 mm (±0.16 in)
    • Greater than 0.8 m (2.62 ft) measuring distance: ±1 mm (±0.04 in)
  • Non-repeatability: ≤ 1 mm (0.04 in)
  • Response time: < 1 s (damping switched off)
  • Influence of ambient temperature:
    • Digital output (HART): Average TC = 2 mm/10 K
    • Analog (current output 1): Zero point TC = 0.02 %/10 K, Span TC = 0.05 %/10 K
    • Analog (current output 2): Zero point TC = 0.08 %/10 K, Span TC = 0.08 %/10 K

Feed water

  • Media Groups:
    • A0 (εr 1.2 to 1.4)
    • A (εr 1.4 to 1.9)
    • B (εr 1.9 to 4)
    • C (εr 4 to 10)
    • D (εr >10)
  • Measurement in vessels with agitator: Measuring ranges vary by media group and antenna type, from 2 m (7 ft) to 60 m (197 ft).
  • Measurement in stilling well: Measuring ranges are 20 m (66 ft) for all media groups with an 80 mm (3 in) PTFE cladded flush mount antenna.
  • Measurement in bypass: Measuring ranges are 20 m (66 ft) for all media groups with an 80 mm (3 in) PTFE cladded flush mount antenna.
  • Influence of gas phase: Systematic measured error can occur due to gas phase pressure and temperature.

Product water

Not applicable for this product type.

Electrical

  • Operating frequency: Approx. 80 GHz
  • Transmission power: Peak power < 1.5 mW, Average output power < 70 µW
  • Output signal: HART (FSK ±0.5 mA over current signal, 1200 Bit/s)
  • Current output 1 (4 to 20 mA passive):
    • Operating modes: 4.0 to 20.5 mA, NAMUR NE 43 (3.8 to 20.5 mA), US mode (3.9 to 20.8 mA).
    • Assigned to: Level linearized.
  • Current output 2 (optional, 4 to 20 mA passive):
    • Operating modes: 4.0 to 20.5 mA, NAMUR NE 43 (3.8 to 20.5 mA), US mode (3.9 to 20.8 mA).
    • Can be assigned to: Level linearized, Distance, Terminal voltage, Electronics temperature, Sensor temperature, etc.
  • Switching output (optional):
    • Type: Open collector.
    • Electrical connection data: U = 16 to 35 VDC, I = 0 to 40 mA.
    • Internal resistor: < 880 Ω.
    • Insulation voltages: 1350 VDC (power supply), 500 VAC (ground).
  • Signal on alarm:
    • Current output: Minimum alarm 3.6 mA, Maximum alarm 22 mA, User-configurable 3.59 to 22.5 mA.
  • Load:
    • 4 to 20 mA passive, HART: Max load resistance depends on supply voltage (e.g., 30 VDC, 23 mA).
    • 4 to 20 mA passive, current output 2: Max load resistance depends on supply voltage.
  • Supply voltage:
    • Non-hazardous, Ex d, Ex e: 10.5 to 35 VDC
    • Ex i: 10.5 to 30 VDC
  • Nominal current: 4 to 20 mA
  • Power consumption: 0.9 W max.
  • Wireless HART data: Minimum start-up voltage 10.5 V, Start-up current < 3.6 mA.

Mechanical / hydraulic

  • Antenna types and maximum measuring ranges:
    • Horn, 316L, 65 mm (2.6 in): 80 m (262 ft)
    • Drip-off, PTFE, 50 mm (2 in): 50 m (164 ft)
    • Cladded flush mount, PTFE, 50 mm (2 in): 50 m (164 ft)
    • Cladded flush mount, PTFE, 80 mm (3 in): 80 m (262 ft)
  • Materials: Specific materials depend on the version (e.g., PTFE, 316L, aluminum).
  • Dimensions: Refer to official documentation for detailed dimensional drawings.
  • Weight: Refer to official documentation.
  • Cable entries: Depend on device version; route cables downwards to prevent moisture ingress.
  • Terminals: Clamping range for supply voltage is 0.5 to 2.5 mm² (20 to 14 AWG); for external ground terminal is 0.5 to 4 mm² (20 to 12 AWG).

Certificates and approvals

  • CE mark
  • RoHS
  • RCM marking
  • Ex approvals
  • Functional safety (SIL2, SIL3)
  • Pressure equipment (≤ 200 bar)
  • Radio approval (EN 302729, EN 302372, FCC, Industry Canada)

Disclaimer: Extraction may miss figures, footnotes, or revisions; contractual data must match the OEM PDF revision used on the project.