Back to Endress + Hauser Level measurement
Endress+Hauser logo

Datasheet summary

Micropilot M FMR63B

Engineering summary from PDF text extraction for Micropilot M FMR63B. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product Name: Micropilot FMR63B HART
  • Document: TI01685F/00/EN/07.24-00
  • Product Type: Free-space radar level measurement instrument for hygienic applications.

Application

  • Continuous, non-contact level measurement of liquids in hygienic applications.
  • Process connections available: Tri-Clamp or M24 adapter concept for hygiene applications.

Operating limits

  • Maximum measuring range: 80 m (262 ft)
  • Temperature range: –40 to +200 °C (–40 to +392 °F)
  • Pressure range: –1 to +25 bar (–14.5 to +363 psi)

Feed water / Process medium

  • Media Groups:
    • A0 (εr 1.2 to 1.4)
    • A (εr 1.4 to 1.9)
    • B (εr 1.9 to 4)
    • C (εr 4 to 10)
    • D (εr >10)
  • Measurement in media with absorbing gas phase may require specialized solutions; contact Endress+Hauser.
  • Dielectric Constant (εr): Critical for measuring range, with different ranges specified for each media group and antenna type.

Product water / Measured variable

  • Measured variable: Distance between reference point and product surface. Level is calculated based on an entered "empty distance".
  • Accuracy: ±1 mm (±0.04 in)
  • Maximum measured error:
    • Up to 0.8 m (2.62 ft): max. ±4 mm (±0.16 in)
    • Greater than 0.8 m (2.62 ft): ±1 mm (±0.04 in)
  • Non-repeatability: ≤ 1 mm (0.04 in)
  • Measured value resolution:
    • Digital: 1 mm
    • Analog: 1 µA

Electrical

  • Operating frequency: Approx. 80 GHz
  • Transmission power: Peak power <1.5 mW, Average output power <70 µW
  • Output Signal: HART (FSK ±0.5 mA over current signal, 1200 Bit/s)
  • Current Output 1 (4 to 20 mA passive):
    • HART, 2-wire.
    • Operating modes: 4.0 to 20.5 mA, NAMUR NE 43 (3.8 to 20.5 mA factory setting), US mode (3.9 to 20.8 mA).
    • Assigned to the level measured value.
  • Current Output 2 (4 to 20 mA passive, optional):
    • Operating modes: 4.0 to 20.5 mA, NAMUR NE 43, US mode.
    • Can be assigned to various device variables including Level, Distance, Temperature, Amplitude, and optional Heartbeat Technology parameters.
  • Switching Output (optional):
    • Type: Open collector.
    • Switching behavior: Binary.
    • Electrical connection data: U = 16 to 35 VDC, I = 0 to 40 mA.
    • Insulation voltages: 1350 VDC (in relation to power supply), 500 VAC (in relation to ground).
  • Signal on Alarm (Current Output):
    • Minimum alarm (factory setting): 3.6 mA
    • Maximum alarm: 22 mA
    • User-configurable: 3.59 to 22.5 mA
  • Supply Voltage:
    • Non-hazardous, Ex d, Ex e: 10.5 to 35 VDC
    • Ex i: 10.5 to 30 VDC
  • Nominal current: 4 to 20 mA
  • Power consumption: 0.9 W max.
  • Wireless HART: Minimum start-up voltage 10.5 V, Start-up current < 3.6 mA.

Mechanical / Hydraulic

  • Antenna Materials: PTFE or PEEK for hygienic requirements.
  • Dimensions: See official PDF for detailed dimensional drawings.
  • Weight: See official PDF for exact values.
  • Materials: See official PDF for material specifications.
  • Process connections: For hygiene applications (e.g., Tri-Clamp or M24 adapter concept).
  • Nozzle Length (Hmax) Limits: Specified for different antenna types and nozzle diameters to ensure optimal performance.
  • Cable entries: Dependent on device version; various glands available (plastic, brass, stainless steel).

Performance Characteristics

  • Reference operating conditions: Temperature +24 °C (±5 °C), Pressure 960 mbar abs. (±100 mbar), Humidity 60 % (±15 %).
  • Response time: < 1 s (when damping is switched off, pulse frequency ≥ 5/s).
  • Influence of ambient temperature:
    • Digital output (HART): Average TC = 2 mm/10 K
    • Analog (current output): Zero point TC = 0.02 %/10 K, Span TC = 0.05 %/10 K (Output 1); Zero point TC = 0.08 %/10 K, Span TC = 0.08 %/10 K (Output 2).
  • Influence of gas phase: High pressure can lead to systematic measured errors; values provided for various gases and pressures.

Certificates and Approvals

  • CE mark
  • ASME BPE
  • RCM marking
  • Ex approvals
  • Functional safety: SIL2 as per IEC 61508, SIL3 for homogeneous redundancy.
  • Radio approval (EN 302372, FCC, Industry Canada).

Disclaimer

Extraction may miss figures, footnotes, or revisions; contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.