Datasheet summary
Nivotester FTL325N
Engineering summary from PDF text extraction for Nivotester FTL325N. Verify every value with the OEM datasheet.
Summary
Product Identification
- Product Name: Nivotester FTL 325 N
- Type: Single channel and three channel isolating amplifier with NAMUR input
Application
- Level limit detection in liquid tanks and hazardous areas.
- Suitable for FM / CSA Division 1 (Zone 0 or Zone 20) measuring sensors.
- Liquid detection in pipes for dry running protection for pumps.
- Over-spill protection for tanks with combustible or noncombustible liquids dangerous to the environment.
- Two-point control and level limit detection with one switching unit.
Features and Benefits
- Intrinsically safe signal circuits (FM IS) for use in explosion hazardous areas.
- High functional safety through line monitoring and corrosion monitoring on tuning forks of Liquiphant M and Liquiphant S (high temperature) measuring sensors.
- Compact housing for series installation on standard rails.
- Simple wiring with plug-in terminal blocks.
- NAMUR interface to EN 50227 (DIN 19234; NAMUR) or IEC 60947-5-6.
Functionality
- The Nivotester supplies a DC voltage to NAMUR-specified sensors and transfers a control current that ranges from < 1.2 mA to > 2.1 mA depending on the switching status.
- The unit measures and evaluates this control current.
- A level alarm relay signals if the measuring sensor is covered or free.
- Faults, such as line interruption or short circuit, are also indicated.
- Fail-safe circuit options allow the relay to operate in a de-energized state for maximum or minimum fail-safe conditions upon fault or power failure.
- Function monitoring system with LED indication and relay de-energization upon fault.
Measuring System
A measuring device consists of one to three measuring sensors, a Nivotester unit, and control or signal devices. Compatible sensors include:
- Liquiphant M with FEL 56 or FEL 58 electronic inserts.
- Liquiphant S (high-temperature) with FEL 56 or FEL 58 electronic inserts.
- Any sensors specified to NAMUR or contact switches with an appropriate resistance circuit.
Input Parameters
- Input for FTL 325N: Galvanically isolated from power supply and output.
- Protection Type: FM Intrinsically Safe, Class I, II, III; Division 1, Groups A-G. CSA Intrinsically Safe, Class I, II, III; Division 1, Groups A-G.
- Connectable Measuring Sensors: Liquiphant M (FTL 50/51, FTL 50 H/51 H, FTL 51 C), Liquiphant S (HT) (FTL 70/71), sensors specified to EN 50227 or IEC 60947-5-6, and contact switches with appropriate resistance circuit.
- Measuring Sensor Power Supply: Supplied by the FTL 325 N Nivotester.
- Connecting Line: Two-wire, shielding not required.
- Line Resistance: Maximum 25 Ω per wire.
- Signal Transmission: Current signal on power supply line.
- Control Current Range: < 1.2 mA / > 2.1 mA.
- Line Interruption Monitoring: < 200 µA.
- Short-Circuit Monitoring: > 6.1 mA (can be switched off).
Output Parameters
- Relay Output per Channel: One potential-free relay contact for the level alarm.
- De-energized Current Safety Circuit: Functionality depends on electronic insert settings and Nivotester error current signal selection.
- Switch Delay: Approximately 0.5 seconds.
- Switching Power:
- AC version: Maximum 253 V~, 2 A~, 500 VA at cos ϕ ≥ 0.7.
- DC version: Maximum 40 V=, 2 A=, 80 W=.
- Life: At least 10⁵ switching operations at maximum contact load.
- Function Displays: LEDs for operation, level alarm, and fault.
- Fault Signal: See official PDF for details.
Electrical
- Power Supply:
- AC version: 85 to 253 VAC, 50/60 Hz.
- DC version: 20 to 60 VDC.
- Power Consumption:
- AC version: Maximum 1.75 W (one channel), 2.25 W (three channel).
- DC version: Maximum 1.2 W (one channel at 20 V), 2.25 W (three channel at 20 V).
- Setting Time/Length: Approximately 10 to 20 seconds after power supply is switched on.
- Galvanic Isolation: All input and output channels and relay contacts are galvanically isolated from each other.
Mechanical / Hydraulic
- Housing Material: Polycarbonate, light grey (RAL 7035).
- Front Cover Material: Polyamid PA6, blue.
- Installation: On top-hat DIN rail, 35 x 7.5 mm or 35 x 15 mm to EN 50022.
- Weight: Approximately 148 g (single channel), 250 g (three channel).
- Connection Terminals: Screw terminals for sensor power supply, limit value relay, fault signal relay, and power supply.
- Connection Cable Cross Section: Maximum 2.5 mm² (single core) or 1.5 mm² (two core).
- Dimensions: See official PDF for detailed dimensions.
Operating Conditions and Installation
- Operating Location: Unit must be housed in a cabinet outside explosion hazardous areas. Optional NEMA 4 (IP 65) protective housing available.
- Installation Position: Mount vertically on a DIN top-hat rail.
- Ambient Temperature:
- Single installation: -20°C to +60°C (-4°F to +140°F).
- Rail mounting, without gaps: -20°C to +50°C (-4°F to +122°F).
- Storage temperature: -20°C to +60°C (-4°F to +140°F).
- Installation in protective housing: -20°C to +40°C (-4°F to +104°F).
- Climatic Class: 3K3 and 3M2, according to EN 60721-3-3.
- Ingress Protection: NEMA 1 (IP 20).
- Electromagnetic Compatibility (EMC): Interference emission to EN 61326, Class B apparatus; interference emission to EN 61326, Appendix A (industry) and NAMUR Recommendation NE 21.
Certificates and Approvals
- CE mark.
- FM approved intrinsically safe, Class I, II, III; Division 1, Groups A-G.
- CSA intrinsically safe, Class I, II, III; Division 1, Groups A-G.
- EN 50227 (DIN 19234; NAMUR) or IEC 60947-5-6.
- EN 60529, EN 61010, EN 61326.
Accessories
- Protective housing, NEMA 4 rated (IP 66) with integrated top-hat rail. Dimensions: 180 x 182 x 165 mm (7.09” W x 7.17” H x 6.50” D). Part Number: 52010132.
Disclaimer: This summary is based on extracted text and may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.