Datasheet summary
Prosonic FMU42
Engineering summary from PDF text extraction for Prosonic FMU42. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: Prosonic M FMU42
- Technology: Ultrasonic measuring technology
- Type: Compact transmitter for non-contact level and flow measurement
Application
- Two-wire version for non-contact level and flow measurement
- Process connection: thread
- International explosion protection certificates available.
Operating limits
- Temperature: –40 to 80 °C (–40 to 176 °F)
- Pressure: 0.7 to 2.5 bar (10.15 to 36.25 psi)
Feed water (Process conditions)
- Process temperature range: –40 to +80 °C (–40 to +176 °F)
- Process pressure range: 0.7 to 2.5 bar abs. (10.15 to 36.25 psi)
Product water (Measurement capabilities)
- Measured variable: Distance D between sensor membrane and product surface. Calculated values include Level (L), Volume (V), and Flow (Q).
- Maximum measuring range:
- Liquids: 10 m (33 ft)
- Bulk solids: 5 m (16 ft)
- Blocking distance: 0.4 m (1.3 ft) for liquids and bulk solids.
- Measured value resolution: 2 mm (0.08 in)
- Maximum measured error: ±0.2 % of the maximum sensor span under reference operating conditions.
- Measuring error: Better than ±4 mm (±0.16 in) or 0.2 % of the measuring distance, whichever is larger.
- Response time (Two-wire version): ≥ 2 s (minimum values)
- Pulse frequency (Two-wire version): ≤ 0.5 Hz
- Accuracy influenced by gas phase: Very good accuracy for media with vapor pressure < 50 mbar at 20 °C. Accuracy may be impacted by high vapor pressures or outgassing media (e.g., ethanol, acetone, ammonia).
Electrical
- Output signal: 4 to 20 mA with HART protocol
- Load: Minimum load for HART communication: 250 Ω
- Output damping: User-defined, 0 to 255 s
- Supply voltage (Two-wire version):
- Standard: 14 to 36 V
- Ex ia: 8 to 30 V
- Ex d: 11 to 30 V
- Power consumption (Two-wire version): 51 to 800 mW
- Current consumption (Two-wire version): 3.6 to 22 mA
- HART ripple: 47 to 125 Hz: Uss = 200 mV (at 500 Ω)
- HART noise: 500 Hz to 10 kHz: Ueff = 2.2 mV (at 500 Ω)
Mechanical / hydraulic
- Process connections: Universal flange DN 80 PN16 / ANSI 3" 150 lbs / JIS 10K 80 and DN 100 PN16 / ANSI 4" 150 lbs / JIS 16K 100. Mounting bracket also available.
- Weight: Approximately 3 kg (6.62 lb)
- Materials in contact with process: Sensor: PVDF; Seal: EPDM or FKM; Flange: PP, PVDF or stainless steel 316L.
- Materials not in contact with process: Housing and cover: AlSi10Mg powder-coated.
- Cable entries: Thread G ½ or NPT ½; Cable gland: M20×1.5.
- Dimensions: See diagrams for F12 housing, T12 housing, sensor, mounting bracket, and flange.
Environment
- Ambient temperature range: –40 to +80 °C (–40 to +176 °F)
- Storage temperature: –40 to +80 °C (–40 to +176 °F)
- Climate class: DIN EN 60068-2-38 (Test Z/AD); DIN/IEC 68 T2-30Db
- Degree of protection: IP68, NEMA 6P (housing closed); IP20, NEMA 1 (housing open, display)
- Vibration resistance: DIN EN 60068-2-64/IEC 68-2-64: 20 to 2 000 Hz, 1 (m/s²)²/Hz; 3 x 100 min
Operability
- Languages: 7 languages supported (de, en, es, fr, ja, it, nl)
- Local operation: With display and operating module (VU331) or via PC with FieldCare.
- Remote operation: With Field Xpert, PC with Commubox FXA195 and FieldCare, or Commubox FXA291 with ToF Adapter FXA291 and FieldCare.
Certificates and approvals
- International explosion protection certificates available. Specific certificates can be selected via the Product Configurator.
Disclaimer
This summary is based on extracted text from a datasheet and may not include all details, figures, footnotes, or revisions. Contractual data must always be verified against the official OEM PDF revision used for the project.