Datasheet summary
Prosonic S FDU91F
Engineering summary from PDF text extraction for Prosonic S FDU91F. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: Prosonic S FDU91F
- Type: Ultrasonic sensor for level and flow measurement
- Manufacturer: Endress+Hauser
- Document Revision: TI01471F/00/EN/01.20, dated 2020-09-14
Application
- Level Measurement: Continuous, non-contact measurement of liquids and bulk solids in silos, on conveyor belts, in material stockpiles, and in crushers.
- Flow Measurement: Measurement in open flumes and measuring weirs.
- Maximum Measuring Range: 10 m (33 ft) for liquids; 5 m (16 ft) for bulk solids.
Features
- Integrated temperature sensor for time-of-flight correction.
- Hermetically welded PVDF sensor for chemical resistance.
- Suitable for harsh ambient conditions with separate transmitter installation (up to 300 m / 984 ft).
- Self-cleaning effect to minimize deposit build-up.
- Weather resistant and flood-proof (IP68).
- International Dust-Ex and Gas-Ex certificates available.
Technical Specifications
Input
- Blocking Distance: 0.3 m (1.0 ft). Signals within this range cannot be measured.
Measuring Range
- Attenuation: Factors influencing the effective sensor range include surface conditions (calm, wavy, turbulent, frothy), dust formation, filling curtains, and temperature differences.
- Surface of liquid: 0 dB (calm) to 20 dB (very turbulent).
- Bulk solids surface: 40 dB (hard, rough) to 60 dB (soft).
- Dust: 0 dB (none) to 20 dB (major).
- Filling curtain: 0 dB (none) to 20 dB (large volumes).
- Temperature difference: 0 dB (up to 20 °C) to 20 dB (up to 80 °C).
- Range Chart: Provided to estimate the sensor range based on total calculated attenuation.
Operating Frequency
- 42 kHz
Power Supply
- Supply voltage is provided by the transmitter.
Process
- Process Temperature:
- Non-Ex: Max. 135 °C (275 °F) for 30 min (Tri-Clamp or flush mounting).
- Ex: –40 to +80 °C (–40 to +176 °F).
- Process Pressure: 0.7 to 4 bar (10.15 to 58 psi).
Mechanical Construction
- Dimensions: Various dimensions are shown for different connection types (FAU80 slip-on flange, G1 thread, NPT1 thread, Tri-Clamp).
- Weight: Approximately 1.6 kg (3.53 lb) including a 5 m (16 ft) cable.
- Materials:
- Sensor housing: 316L (1.4404/1.4435)
- Cable gland: PA
- Pipe adapter: CuZn nickel-plated
- O-ring: EPDM
- Connecting cable: PVC
Environment
- Degree of Protection: IP68 / NEMA6P (tested for 24 h at 1.83 m (6 ft) under water).
- Vibration Resistance: DIN EN 60068-2-64; 1 (m/s²)²/Hz for 3x100 min.
- Storage Temperature: Identical to process temperature.
- Thermal Shock Resistance: Based on DIN EN 60068-2-14.
- Electromagnetic Compatibility: Meets EN 61326 series and NAMUR Recommendation EMC (NE 21). Designed for use in an "industrial environment" and meets class A requirements for interference emission.
Certificates and Approvals
- CE mark
- RoHS (2011/65/EU)
- RCM-Tick marking
- Ex approval (details available in Product Configurator)
Installation
Level Measurement Conditions
- Recommended distance to vessel wall: 1/6 of vessel diameter.
- Avoid mounting in the center of the vessel.
- Do not measure through the filling curtain.
- No internal fixtures in the signal beam.
- For bulk solids, align sensor perpendicular to the product surface using the FAU40 alignment unit.
- The lower edge of the sensor should be inside the vessel.
- Maximum level must not enter the blocking distance.
Flow Measurement Conditions
- Mount sensor upstream of the maximum upstream level (Hmax) plus blocking distance.
- Position sensor in the center of the channel or weir.
- Align sensor perpendicular to the water surface.
- Observe mounting distance to flume constriction or weir edge.
- Protect sensor from sun and precipitation with a weather protection cover.
Installation Options
- Nozzle mounting (conditions for nozzle and maximum lengths specified).
- Mounting with U-rail or bracket.
- Mounting with FAU40 alignment unit.
- Mounting with a 1" sleeve welded to a grating.
- Flush mounting with slip-on flange (with or without adapter flange).
Accessories
- Sensor extension cable: Maximum total length 300 m (984 ft). Order number: 71027743.
- Mounting bracket for ceiling mounting: Material 316L (1.4404). Order No.: 71093130.
- FAU40 alignment unit: For aligning sensor with bulk solids surface, swivel range 15°.
- IP66 protective housing for RNB130 power supply unit: Order number: 51002468.
Supplementary Documentation
- Technical Information (TI) and Operating Instructions (BA) for FMU90 and FMU95 transmitters are available.
- Description of Device Parameters (GP) are available.
Ordering Information
- Detailed ordering information is available via the Endress+Hauser website and Product Configurator.
- 5-point linearity protocol: Available for the complete measuring system (sensor + transmitter) when specified during ordering. Span is defined by Empty (E) and Full (F) calibration values.
- E ≤ 10,000 mm (394 in).
- F = 100 to 9,700 mm (3.94 to 382 in).
- A ≥ 300 mm (11.8 in).
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.