Datasheet summary
Prosonic S FMU95
Engineering summary from PDF text extraction for Prosonic S FMU95. Verify every value with the OEM datasheet.
Summary
Product identification
The Prosonic S FMU95 is an ultrasonic transmitter for continuous, non-contact level measurement of fluids, pastes, sludge, and bulk materials. It supports up to 10 ultrasonic sensors (FDU90/91/91F/92/93/95).
Operating limits
- Measuring range: Up to 45 m (148 ft), depending on the sensor and material.
- Blocking distance: Level echoes within the blocking distance range cannot be evaluated. Blocking distances for individual sensors are detailed in TI00396F.
Feed water (Material Measured)
- Suitable for fluids, pastes, sludge, and powdery to coarse bulk materials.
Product water (Transmitter Output)
- Transmittable values via PROFIBUS DP: Main value (level 1 to 10), Distances, Temperatures, Averages/sums.
- Function blocks: 20 Analog Input Blocks (AI).
Electrical
- Supply voltage:
- AC version: 90 to 253 VAC (50/60 Hz)
- DC version: 10.5 to 32 VDC
- Power consumption:
- AC version: max. 23 VA
- DC version: max. 14 W (typically 8 W)
- Current consumption:
- AC version: max. 100 mA at 230 VAC
- DC version: max. 580 mA at 24 VDC
- Galvanic isolation: Between auxiliary energy, sensor inputs, and bus connection (PROFIBUS DP).
- Fuse:
- 400 mA T /AC
- 2 A T /DC
Mechanical / hydraulic
- Housing versions:
- Field housing (optional integrated display)
- DIN rail mounting housing (optional integrated display)
- DIN rail mounting housing with separate display and operating module
- Ingress protection:
- Field housing: IP66 / NEMA 4x
- DIN rail housing: IP20
- Separate display: IP65 / NEMA 4 (front panel), IP20 (rear panel)
- Materials:
- Field housing: PC-FR
- DIN rail housing: PBT-GF
- Separate display and operating module: PC
Performance characteristics
- Reference operating conditions: Temperature 24±5 °C, Pressure 960±100 mbar, Relative humidity 60±15 %.
- Maximum measuring error: ±0.2 % of the maximum span.
- Typical measuring error: Better than ±2 mm (0.08 in) + 0.17 % of the measured distance (includes linearity, repeatability, and hysteresis).
- Measured value resolution: 1 mm (0.04 in) with FDU90/FDU91.
- Measuring frequency: 0.2 Hz (5 sensors), 0.1 Hz (10 sensors).
Operating conditions: Environment
- Ambient temperature: −40 to +60 °C (−40 to +140 °F).
- Storage temperature: −40 to +60 °C (−40 to +140 °F).
- Climate class:
- Field housing: DIN EN 60721-3 4K2/4K5/4K6/4Z2/4Z5/4C3/4S4/4M2
- DIN rail housing: DIN EN 60721-3 3K3/3Z2/3Z5/3B1/3C2/3S3/3M1
- Vibration resistance:
- DIN rail housing: DIN EN 60068-2-64; 20 to 2000 Hz; 0.5 (m/s²)²/Hz
- Field housing: DIN EN 60068-2-64; 20 to 2000 Hz; 1.0 (m/s²)²/Hz
- Electromagnetic compatibility (EMC): According to EN 61326- series and NAMUR recommendation EMC (NE21). Meets class A requirements for industrial environments.
Human interface
- Display and operating module: 6-line plain text display, 15 languages selectable.
- Operation: Menu-guided, softkey operation.
- Diagnosis: Envelope curves on display, "FieldCare" operating program.
- Linearization: Up to 32 points, freely configurable.
- Locking: Via switch, key combination, or software code.
Certificates and Approvals
- CE mark: Confirms compliance with EC directives.
- Ex approval: Available certificates are listed in ordering information; requires strict adherence to associated safety instructions (XA) and control drawings (ZD).
Documentation
- Technical Information: TI00396F (for sensors FDU90/91/91F/92/93/95).
- Operating Instructions: BA00344F, BA00345F (Description of Instrument Functions), BA00346F (Slot-Index tables).
- Safety Instructions: XA00326F (for ATEX II 3D).
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.