Datasheet summary
CCS51
Engineering summary from PDF text extraction for CCS51. Verify every value with the OEM datasheet.
Summary
Product identification
- Product name: CCS51
- Product type: Sensor for determining free chlorine
- Document Title: CCS51
- Document Number: TI01424C/07/EN/01.18
- Revision Date: 2018-09-20
Application
The CCS51 is a free chlorine sensor designed for skid builders and end customers. It is suitable for measuring free chlorine in:
- Drinking water
- Food industry
- Utilities
- Pool water
Function and system design
Measuring principle
The sensor uses an amperometric measuring principle to determine free chlorine levels, based on hypochlorous acid (HOCl). HOCl diffuses through a sensor membrane and is reduced at a gold cathode, while silver is oxidized at a silver anode. The resulting current is proportional to the free chlorine concentration. A pH measurement is used to compensate for the dependency of HOCl concentration on pH. The transmitter calculates the concentration in mg/l (ppm) from the current signal.
The sensor comprises a membrane cap (measuring chamber with membrane) and a sensor shaft with an anode and a cathode embedded in plastic. The electrodes are separated from the medium by a membrane which prevents electrolyte leakage and contamination. Calibration is performed by colorimetric comparison measurement using the DPD method.
A complete measuring system includes the CCS51 sensor, a Flowfit CCA250 flow assembly, and a transmitter (e.g., Liquisys CCM223/253). Optional components include an immersion assembly (Flexdip CYA112) and a proximity switch.
Input
Measured variables
- Free chlorine (HOCl): [mg/l, μg/l, ppm, ppb]
- Temperature: [°C, °F]
Measuring ranges
- CCS51-**11AD*: 0 to 5 mg/l (ppm) HOCl
- CCS51-**11BF*: 0 to 20 mg/l (ppm) HOCl
Signal current
- CCS51-**11AD*: 33 to 63 nA per 1 mg/l (ppm) HOCl
- CCS51-**11BF*: 9 to 18 nA per 1 mg/l (ppm) HOCl
Electrical
- Sensor cable length: Maximum 3 m (9.8 ft)
Performance characteristics
Reference operating conditions
- Temperature: 20 °C (68 °F)
- pH value: pH 5.5 ±0.2
- Flow: 40 to 60 cm/s
- Medium: HOCl-free base medium, mains water
Response time
- T90: < 25 s (after completing polarization)
Measured value resolution of sensor
- CCS51-**11AD*: 0.03 µg/l (ppb) HOCl
- CCS51-**11BF*: 0.13 µg/l (ppb) HOCl
Measured error
- ±2 % or ±5 µg/l (ppb) of value measured (whichever is higher)
LOD (limit of detection)
- CCS51-**11AD*: 0.002 mg/l (ppm)
- CCS51-**11BF*: 0.002 mg/l (ppm)
LOQ (limit of quantification)
- CCS51-**11AD*: 0.005 mg/l (ppm)
- CCS51-**11BF*: 0.007 mg/l (ppm)
Repeatability
- CCS51-**11AD*: 0.0031 mg/l (ppm)
- CCS51-**11BF*: 0.0035 mg/l (ppm)
Nominal slope
- CCS51-**11AD*: 48 nA per 1 mg/l (ppm) ClO₂
- CCS51-**11BF*: 14 nA per 1 mg/l (ppm) ClO₂
Long-term drift
- < 1 % per month (mean value)
Polarization time
- Initial commissioning: 60 min
- Recommissioning: 30 min
Operating time of the electrolyte
- At maximum concentration and 55 °C: 60 days
- At 50 % of measuring range and 20 °C: 1 year
- At 10 % of measuring range and 20 °C: 2 years
Installation
Orientation
- Install the sensor at an angle of at least 15° to the horizontal. Do not install upside-down.
Immersion depth
- 77 mm (3.03 in)
Installation in Flowfit CCA151 flow assembly
- Minimum flow rate: 5 l/h (1.32 gal/h)
- Counterpressure must not exceed 1 bar (14.5 psi) and must remain constant.
- Avoid negative pressure.
- Heavily contaminated water should be filtered.
Installation in Flowfit CCA250 flow assembly
- Minimum flow rate: 30 l/h (7.92 gal/h)
- Counterpressure must not exceed 1 bar (14.5 psi) and must remain constant.
- Avoid negative pressure.
- Flow regulation is provided by a needle valve in the range of 30 to 120 l/h (7.9 to 30 gal/h).
Installation in other flow assemblies
- Flow velocity at the membrane must be at least 15 cm/s (0.49 ft/s).
- Flow direction should be upwards, and air bubbles must be removed.
- Flow must be directed to the membrane.
Installation in Flexdip CYA112 immersion assembly
- Ensure sufficient flow towards the sensor.
Environment
Ambient temperature range
- -20 to +60 °C (-4 to 140 °F)
Storage temperature
- Longterm storage: +0 to 35 °C (32 to 95 °F) (non-freezing)
- Storage up to max 48 h: 35 to 50 °C (95 to 122 °F) (non-freezing)
- Without electrolyte: –20 to 60 °C (–4 to 140 °F)
Degree of protection
- IP 68 (1.8 m (5.91 ft) water column over 7 days at 20 °C (68 °F))
Process
Process temperature
- +0 to 55 °C (32 to 130 °F), non-freezing
Process pressure
- Medium pressure directly upstream from the sensor membrane must not exceed 1 bar (14.5 psi) abs.
Flow velocity
| Velocity [cm/s] | Volume flow [l/h] (Flowfit CCA250) | Volume flow [l/h] (Flowfit CCA151) | Volume flow [l/h] (Flexdip CYA112) |
|---|---|---|---|
| Minimum | 30 | 5 | |
| Minimum | 15 | ||
| Maximum | 80 | 145 | 20 |
pH range
- Calibration: pH 4 to 8
- Measurement: pH 4 to 9
Mechanical construction
Dimensions
- See technical drawing (dimensions in mm (in)).
Weight
- Sensor with membrane cap and electrolyte (without protection cap and without adapter): Approx. 250 g (8.82 oz)
Materials
- Sensor shaft: PVC
- Cable sheathing: PVC
- Membrane: PVDF
- Membrane cap: PVDF
- Protection cap: PC Makrolon (polycarbonate), Kraiburg TPE TM5MED (seal), PC Makrolon (polycarbonate) (cover)
- Sealing ring: FKM
Certificates and approvals
- CE mark
- Declaration of Conformity: Product meets the requirements of harmonized European standards and EU directives.
Scope of delivery
- Disinfection sensor (membrane-covered, ⌀25 mm) with protection cap (ready for use)
- Bottle of electrolyte (50 ml (1.69 fl.oz))
- Replacement membrane cap in protection cap
- Operating Instructions
Accessories
- Maintenance kit CCV05: Contains membrane caps, electrolyte, and sealing sets.
- Flowfit CCA151: Flow assembly.
- Flowfit CCA250: Flow assembly for chlorine and pH/ORP sensors.
- Flexdip CYA112: Immersion assembly.
- Photometer PF-3: Compact hand-held photometer for determining free available chlorine.
- Adapter kit CCS5x(D) for CCA250: Adapter including O-rings and studs.
- Adapter kit CCS5x(D) for CYA112: Adapter including O-rings and studs.
- COY8: Zero-point gel for oxygen and chlorine sensors.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.