Datasheet summary
CCS58E
Engineering summary from PDF text extraction for CCS58E. Verify every value with the OEM datasheet.
Summary
Product Identification
- Product Name: Memosens CCS58E
- Description: Digital sensor with Memosens technology for determining ozone.
- Applications: Wastewater (discharge guarantee), Drinking water (disinfection), Process water (hygienic packaging and bottling).
Measuring Principle
The CCS58E utilizes an amperometric measuring principle to determine ozone (O₃). Ozone diffuses through a membrane and is reduced to hydroxide ions (OH⁻) at the working electrode. A current flows, proportional to the ozone concentration, which is then used by the transmitter to calculate the concentration in mg/l (ppm). This process is largely independent of pH over a wide range.
Key Features and Benefits
- Robust Membrane Technology: High surfactant resistance, suitable for cleaning processes.
- High Specificity for Ozone: Ensures reliable measured values for disinfection.
- Heartbeat Technology: Offers advanced capabilities like an electrolyte counter for predictive maintenance.
- Fast Commissioning: Factory calibration and plug-and-play installation.
- Memosens Technology:
- Non-contact, digital signal transmission.
- Galvanic isolation, dust- and waterproof (IP 68).
- Sensor can be calibrated in a lab.
- Intrinsically safe electronics for use in hazardous areas.
- Stores sensor data (total hours, extreme condition hours, temperature hours, calibration history) for predictive maintenance and documentation.
- Easy calibration and quick sensor replacement, increasing measuring point availability.
- Cross-Sensitivity: No cross-sensitivities to free chlorine, free bromine, total chlorine, total bromine, hydrogen peroxide, peracetic acid. Minimum cross-sensitivity to chlorine dioxide. Surfactants do not affect performance.
Measuring System Components
A complete measuring system includes:
- Memosens CCS58E sensor
- Flowfit CYA27 flow assembly (or other appropriate mounting adapter like Flexdip CYA112)
- Measuring cable (CYK10, CYK20)
- Transmitter (e.g., Liquiline CM44x or CM44xR with appropriate firmware)
- Optional accessories: extension cable, proximity switch, pH sensor.
Input Parameters
- Measured Variables: Ozone, Temperature
- Ozone Units: mg/l, μg/l, ppm, ppb
- Temperature Units: °C, °F
- Measuring Range (Ozone): 0 to 2 mg/l (ppm). Not suitable for checking the absence of ozone.
Performance Characteristics
- Reference Operating Conditions:
- Temperature: 15 °C (59 °F) ±2 °C (±3.6 °F)
- pH Value: pH 7.2 ±0.2
- Flow: 140 cm/s (4.6 ft/s) ±5 cm/s (±0.16 ft/s)
- Sample Water: Drinking water
- Response Time (T90): < 8 min (440 s)
- Polarization Time:
- Initial commissioning: 60 min
- Recommissioning: 20 min
- Measured Value Resolution: At most 0.05% of the measured value above the limit of quantification (LOQ).
- Measurement Error: ±2% and ±5 µg/l (ppb) of measured value (whichever is higher).
- LOD (Limit of Detection): 0.018 mg/l (ppm)
- LOQ (Limit of Quantification): 0.061 mg/l (ppm)
- Repeatability: 0.055 mg/l (ppm) (for CCS58E-****31AC)
- Nominal Slope: 226 nA per 1 mg/l
- Long-Term Drift: 1 % per month
- Operating Time of Electrolyte: 3 to 6 months
- Operating Time of Membrane Cap: Replace once per year (with electrolyte).
- Intrinsic Consumption: Negligible.
Installation
- Orientation: Must be installed at an angle of at least 15° to the horizontal. Do not install upside-down.
- Immersion Depth: At least 55 mm (2.17 in).
- Assembly Installation (Flowfit CYA27):
- Minimum flow: 29 cm/s (1.0 ft/s) at the membrane; minimum volume flow of 5 l/h or 30 l/h depending on the assembly version.
- Maximum back pressure: 1 bar relative (14.5 psi relative).
- Avoid negative pressure.
- Filter heavily contaminated water to avoid buildup.
- General Flow Assembly Installation: Ensure minimum flow velocity of 29 cm/s (1.0 ft/s), upward flow direction, and direct exposure of the membrane to flow.
Environment and Process Limits
- Ambient Temperature: 0 to 55 °C (32 to 131 °F)
- Storage Temperature: 0 to 55 °C (32 to 131 °F) (without membrane cap and electrolyte)
- Degree of Protection: IP68
- Process Temperature: 0 to 45 °C (32 to 110 °F), non-freezing
- Pressure: 1 bar relative (14.5 psi relative), no pressure shocks or vibrations.
- pH Range:
- Calibration: pH 4 to 8
- Measurement: pH 4 to 9
- Material Resistance: pH 2 to 11
- Conductivity: 0.03 to 40 mS/cm. Can be used in low conductivity media like demineralized water.
- Flow: See "Installation" section.
Mechanical Construction
- Dimensions: Refer to diagram in document. Main dimensions approximately 160 mm length, 25 mm diameter.
- Weight:
- Membrane cap: 14.45 g (0.5 oz)
- Sensor, total: 93.45 g (3.3 oz)
- Materials:
- Membrane cap sleeve: PVC
- Sensor shaft: PVC
- Membrane: Plastic film
- Membrane holder: Stainless steel 1.4571
- Electrode body: PEEK
- Sealing ring: Silicone rubber
- Cable Specification: Max. 100 m (330 ft), including cable extension.
Certificates and Approvals
Current certificates and approvals are available via the Endress+Hauser Product Configurator on their website.
Ordering Information
- Detailed ordering information is available through sales organizations or the Product Configurator at www.endress.com.
- Scope of Delivery: Sensor, protection cap, electrolyte bottle, emery paper, Operating Instructions, manufacturer's certificate.
- Accessories: Maintenance kit CCV05, Memosens data cable CYK10, Memosens laboratory cable CYK20, Flowfit CYA27, Flexdip CYA112, photometer PF-3, adapter kits, quick fastener kit, zero-point gel COY8.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.