Datasheet summary
Ceragel CPS72
Engineering summary from PDF text extraction for Ceragel CPS72. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: Ceragel CPS72 and CPS72D
- Type: ORP electrodes, analog and digital with Memosens technology
- Intended Use: Process technology, hygienic and sterile applications with poison-resistant reference and ion trap.
- Approvals: ATEX, FM, and CSA for hazardous areas.
Application Areas
- Hygienic and sterile applications (sterilizable, autoclavable):
- Fermenters
- Biotechnology
- Pharmaceutical industry
- Food industry
- Process technology and monitoring:
- Rapidly changing ORP measured values
- Processes with a high proportion of electrode poisons such as H₂S
Key Features and Benefits
- Certified biocompatibility, no cytotoxicity
- Acrylamide-free bridging electrolyte
- Integrated temperature sensor (for Memosens versions)
- Poison-resistant reference with ion trap for extended service life
- Bridging electrolyte free of silver ions
- Suitable for CIP/SIP cleaning and autoclavable (depending on version, up to 140 °C / 284 °F)
- Short response times due to ceramic junction
- Long service life ensured by ion trap protection against poisoning
- Memosens technology (CPS72D) for digital, non-contact data transmission, ensuring process safety by eliminating moisture issues and galvanic decoupling.
- Automatic error message generation and immediate error detection with Memosens.
- Convenient calibration in the lab, quick replacement of precalibrated sensors, and predictive maintenance capabilities with Memosens.
Measuring Principle
- Measured Variable: Oxidation-Reduction Potential (ORP)
- Sensing Element: Gold or platinum electrode (instead of glass membrane).
- Reference System: Integrated Ag/AgCl reference system.
Operating Limits
- ORP Measuring Range: -1500 to 1500 mV
- Process Temperature Range:
- -15 to 140 °C (5 to 284 °F)
- -15 to 135 °C (5 to 275 °F) for sensors with Ex approval and analog sensors.
- Ambient Temperature Range:
- Operating: Not specified, but sensor must not be used below -15 °C (5 °F).
- Storage: 0 to 50 °C (32 to 120 °F).
- Process Pressure (absolute): 0.8 to 11 bar (12 to 160 psi)
Electrical Characteristics
- Memosens Technology (CPS72D):
- Inductive, non-contact data transmission.
- Digitizes measured values in the sensor.
- Galvanically decoupled from the medium.
- EMC safety guaranteed by screening.
- Intrinsically safe electronics for hazardous areas.
- Data Storage (CPS72D): Manufacturer data, calibration data, operating data can be saved in the sensor.
Mechanical Construction
- Materials:
- Electrode Shaft: Glass
- Reference System: Ag/AgCl
- Junction: Ceramic (sterilizable and autoclavable)
- ORP-sensitive Element: Platinum
- Bridging Electrolyte: Acrylamide-free, non-cytotoxic, gel.
- Process Connection: Pg 13.5 thread
- Plug-in Heads: GSA (non-Ex), ESA (Ex, for analog), Memosens (digital, Ex or non-Ex).
- Weight: 0.1 kg (0.22 lbs)
Certificates and Approvals
- Ex Approval (CPS72 ESA, CPS72D): ATEX/NEPSI, FM/CSA, IECEx.
- Biocompatibility: Certified according to ISO 10993-5:1993, USP <87>.
- TÜV Certificate (ESA and Memosens): Pressure resistance 16 bar (232 psi).
- Electromagnetic Compatibility (CPS72D): Interference emission and immunity as per EN 61326:2006.
Installation Instructions
- Orientation: Electrodes must not be installed upside down. Inclination angle must be at least 15° from the horizontal.
- Assembly: Finger-tighten electrode (3 Nm) when installing in Endress+Hauser assemblies.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.