Datasheet summary
Ceraliquid CPS41
Engineering summary from PDF text extraction for Ceraliquid CPS41. Verify every value with the OEM datasheet.
Summary
Product identification
Ceraliquid CPS41D and CPS41 pH electrodes, digital with Memosens technology, and analog.
Application
Designed for process engineering with ceramic diaphragm and KCl liquid electrolyte. Optionally available with integrated temperature sensor. Suitable for media with very low conductivity or a high proportion of organic solvents or alcohol in industries including chemical, organic chemicals, power stations, food, and laboratories. Features ATEX- and FM-approval for use in hazardous areas.
Benefits include:
- KCl liquid electrolyte, enabling use with very low conductivity values (≥ 0.1 μS/cm).
- Ceramic diaphragm with specified KCl flow rate.
- Can be used up to 11 bar (160 psi) abs. pressure if counterpressure is applied.
- Poison-resistant due to a separate reference capillary.
- Suitable for CIP / SIP.
- pH membrane glass for applications involving steam sterilization.
- Available in four lengths: 120, 225, 360 and 425 mm.
- Optional integrated Pt100, Pt1000 temperature sensor for analog sensors or NTC30K for digital sensors.
Memosens technology advantages:
- Maximum process safety via non-contact, inductive signal transmission.
- Data security through digital data transmission.
- Easy to use with sensor data saved in the sensor.
- Enables predictive maintenance by recording sensor load data in the sensor (Memobase Plus CYZ71D).
Operating limits
- Process temperature:
- Version AB, AC: -15 to 80 °C (5 to 176 °F)
- Version BB, BC: 0 to 135 °C (32 to 275 °F)
- Process pressure (absolute): 1 to 11 bar (15 to 160 psi) with counterpressure applied via a separate KCl vessel.
- Ambient temperature range: Not to be used below -15 °C (5 °F).
- Storage temperature: 0 to 50 °C (32 to 122 °F).
Feed water
- Minimum conductivity:
- AB, BB version (1 diaphragm): Min. 5 μS/cm
- AC, BC version (3 diaphragms): Min. 0.1 μS/cm
Product water
- pH range:
- Version AB, AC: 1 to 12 pH
- Version BB, BC: 0 to 14 pH
Electrical
- Digital Sensors (CPS41D) with Memosens Technology:
- Non-contact, inductive signal transmission.
- Intrinsically safe electronics for hazardous areas.
- Analog Sensors (CPS41):
- TOP68 plug-in head.
- Temperature sensor:
- CPS41D: NTC30K
- CPS41: Pt100, Pt1000
Mechanical / hydraulic
- Electrode shaft: Glass.
- pH membrane glasses: Type A, B.
- Diaphragm: Ceramic, suitable for sterilization.
- Process connection: Pg 13.5.
- Lengths: 120, 225, 360, and 425 mm.
- Weight: 0.1 kg (0.2 lbs).
- Plug-in heads:
- Memosens (CPS41D): Pressure resistance 16 bar (232 psi) rel.
- TOP68 (CPS41): Pressure resistance 16 bar (232 psi) rel.
- Reference system: Ag/AgCl reference lead, bridge electrolyte: liquid KCl, 3M, AgCl-free.
- Installation angle: At least 15° from the horizontal.
Certificates and approvals
- Ex approval for CPS41D: ATEX, IECEx, FM (IS Class I Div. 2 with CM42/CPM153), CSA (IS Class I Div. 2 with CM42/CPM153).
- Ex approval CPS41 (ESA, ESS): ATEX, CSA (IS Class I Div. 2 with CPM153), FM (Class I Div. 2 with CPM153).
- TÜV certificate for Memosens plug-in head: Pressure resistance 16 bar rel. (232 psi).
- TÜV certificate for TOP68 plug-in head: Pressure resistance 16 bar rel. (232 psi).
- Electromagnetic compatibility CPS41D: Interference emission and interference immunity as per EN 61326: 2012.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.