Datasheet summary
Ceramax CPS341D
Engineering summary from PDF text extraction for Ceramax CPS341D. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: Ceramax CPS341D
- Type: pH sensor for use in the pharmaceutical industry and food production
- Technology: Digital with Memosens technology
- Document: TI00468C/07/EN/08.24-00
Application
- Food production (including highly viscous media)
- Beverage production and filling
- Quality control
- Pharmaceutical industry:
- Water treatment
- Active ingredient production
- Active ingredient preparation
- Fermentation
- Biotechnology
Measuring principle
- Measured Variable: pH value, Temperature
- pH Measurement: Based on the electrochemical potential generated by H⁺ ions on the pH-sensitive enamel, measured against an integrated Ag/AgCl reference system. The voltage is converted to pH using the Nernst equation.
- Measuring System: Comprises pH sensor CPS341D, electrolyte vessel (e.g., CPS341Z-D1), transmitter (e.g., Liquiline CM42, CM44x), and Memosens data cable (e.g., CYK10, CYK20). An optional ultrasonic sensor (CPS341Z-D2) can monitor electrolyte level.
Features and Benefits
- Continuous online measurement: Possible while the process is running.
- Direct installation: In vessel nozzles or piping.
- Self-cleaning: Achieved by the flowing medium.
- Long-term stability: Over many years.
- High mechanical loading capacity: Due to steel substrate.
- Corrosion resistance: High resistance to acids.
- Hygienic design: Suitable for in-line CIP and SIP cleaning.
- Memosens Technology Advantages:
- Maximum process safety.
- Data security via digital transmission.
- Easy to use with sensor data saved in the sensor.
- Predictive maintenance enabled by recording sensor load data.
Input Parameters
- pH Measuring Range:
- 0 to 10 pH (linear range)
- -2 to 14 pH (application range)
- Temperature Measuring Range: 0 to 140 °C (32 to 284 °F)
- Process Pressure Range: 0.8 to 7 bar (11.6 to 101.5 psi) (absolute)
- Minimum Conductivity: 50 μS/cm
Operating Limits
- Process Temperature Range: 0 to 140 °C (32 to 284 °F)
- Process Pressure Range: 0.8 to 7 bar (11.6 to 101.5 psi) (absolute)
- Ambient Temperature Range: Not specified, but risk of damage below 0 °C (32 °F) due to frost.
- Storage Temperature: 0 to 50 °C (32 to 122 °F)
- Degree of Protection: IP 68 (10 m (33 ft) head of water at 25 ˚C (77 ˚F) over 45 days, 1 mol/l KCl)
- Electromagnetic Compatibility (EMC): Interference emission and immunity as per EN 61326: 2012.
Mechanical and Hydraulic
- Dimensions: Varies with process connection type (e.g., Tri-Clamp DN50, Varivent DN50, milk pipe DN25/DN50, socket DN25/DN30). Total length approximately 295 mm (11.6 in).
- Weight: 0.6 kg (1.3 lbs)
- Materials:
- Sensor body: Glass-lined steel
- Adapter and terminal head: Stainless steel 1.4404 (AISI 316 L), PVDF, PTFE
- Electrolyte vessel: Stainless steel 1.4301 (AISI 304)
- Process connections: Stainless steel 1.4404 (AISI 316 L)
- Electrolyte Volume: 1.6 ml (0.05 fl oz) in the sensor.
- Plug-in head: Memosens, pressure resistance 16 bar (232 psi) (relative).
- Process Connections: M20, Nipple DN25, Nipple DN30, Varivent DN50/40, Dairy fitting DN50, Dairy fitting DN25, Tri-Clamp DN50.
- Mounting Orientation: Any angle (0 to 360 °).
Electrical
- Electrical Connection: Via measuring cable CYK10.
- Power Supply (for optional ultrasonic sensor CPS341Z-D2): 12 ... 30 V DC, max. 200 mA (PNP output).
Data Storage (Memosens Technology)
The sensor can store: Manufacturer data, serial number, order code, date of manufacture, calibration data (date, slope, zero point), application data (temperature/pH ranges, operating hours at extreme conditions), number of sterilizations, and glass membrane resistance. This data is viewable with compatible transmitters.
Accessories
- Electrolyte Vessel: CPS341Z-D1
- Ultrasonic Level Monitoring Sensor: CPS341Z-D2
- Cable with LED Indicator: CPS341Z-D3
- Relay: CPS341Z-D4 (type KCD2-R, P+F)
- KCl Electrolyte: Sterile, 1 l (0.26 gal) plastic bottle (CPS341Z-D5)
- Buffer Solutions: CPY20
- Memosens Data Cable: CYK10
Disclaimer: Extraction may miss figures, footnotes, or revisions; contractual data must match the OEM PDF revision used on the project.