Datasheet summary
CleanFit CPA472D
Engineering summary from PDF text extraction for CleanFit CPA472D. Verify every value with the OEM datasheet.
Summary
Product identification
Product: Cleanfit CPA472D Type: Retractable process assembly for pH and ORP sensors Version: Heavy duty
Application
- Chemical industry
- Biodiesel plants
- Power plants
- Plant design
- Tanks, process vats or pipes
Functions and Features
- Modular, robust, heavy-duty retractable assembly designed for harsh environments.
- Can move to measurement position with process pressures up to 10 bar (145 psi) (pneumatic actuation).
- Prevents assembly movement if a sensor is not installed (safety stop mechanism).
- Accommodates gel electrodes or pressure-overlay KCl electrodes.
- Features electrode rinsing in service position.
- Compatible with fully automated cleaning and calibration using Topcal CPC310.
- Service-friendly design.
- Locking mechanism for safety during sensor replacement.
- Pressure-tight stainless steel housing.
- Easy sensor replacement and Memosens laboratory calibration.
- Flexible material choices and modifications.
- Can be manually or pneumatically operated without interrupting the process for:
- Separating from the process and moving to rinse chamber.
- Rinsing with water or cleaning solution.
- Keeping wet during operation pauses.
- Dismounting.
- Sterilizing.
- Calibrating.
- Available in PEEK, PVDF, conductive PVDF, Alloy C22, and Stainless steel 1.4571 (AISI 316 Ti).
- Modular design with three modules in contact with the medium: Rinse chamber, electrode holder, and raised face.
Operating Limits
- Process temperature: 0 to 140 °C (32 to 284 °F)
- Process pressure:
- Manual actuation: 0 to 4 bar (0 to 58 psi) overpressure.
- Pneumatic actuation: 0 to 10 bar (0 to 145 psi) overpressure.
- Ambient temperature: Not below 0 °C (32 °F).
- Limit position switches (NAMUR type) maximum temperature: 90 °C (194 °F).
Feed Water / Process Medium
- Designed for aggressive chemicals.
Product Water / Output
- Not applicable to this product.
Mechanical / Hydraulic
- Materials in contact with medium:
- Electrode holder: PEEK, PVDF, conductive PVDF, Alloy C22, Stainless steel 1.4571 (AISI 316 Ti).
- Rinse chamber and raised face: PEEK, PVDF, conductive PVDF, Alloy C22, Stainless steel 1.4571 (AISI 316 Ti).
- Seals: EPDM/FPM (Viton®)/FFKM (Kalrez®).
- Materials not in contact with medium:
- Housing: Stainless steel 1.4404 (AISI 316 L).
- Seals: FPM.
- Limit position switches (NAMUR-type) front surface: PBT, cable PVC.
- Handle/protection cap: PVC.
- Immersion Depth Options:
- Standard version: Up to 146 mm (5.75"). Applicable with 225 mm gel electrodes, 360 mm gel electrodes with adapter, or 360 mm KCl electrodes.
- Long version: Up to 280 mm (7.87"). Applicable with 360 mm gel electrodes.
- A special version with an immersion depth of 378 mm (14.9") is available for 425 mm gel sensors.
- Process Connections:
- Flanges: DN 50 / DN 80 / ANSI 2" / JIS (IS10K50A) for tanks.
- Flow assembly pipes with sight glass: DN 50 / DN 80.
- Thread: G 1¼ with union nut (for metal versions).
- Weight: 7.5 to 12.0 kg (16.54 to 26.46 lbs) depending on material.
- Rinse Connections: 2 x G¼ (internal) or 2 x NPT ¼" (internal) or 2 x pipe 8 x 60 Swagelok as nozzle.
Electrical
- Limit Position Switches:
- Pneumatic: 3/2 way valve; thread M 12 x 1; connection for hoses with OD = 6 mm (0.24").
- Electric: Inductive (NAMUR type); cable length: 10 m (32.8 ft); housing material: stainless steel; thread M 12 x 1; nominal voltage: 8 V; 0II 1G EEx ia IIC T6; switching distance: 2 mm, flush.
- Transmitter Compatibility: Liquiline CM42, Mycom S CPM153, Liquiline CM44x/CM44xR, Liquisys CPM223/253, Mycom CPM153.
- Measuring Cables: CPK9, CPK1, CPK12.
- Data Cable: CYK10 Memosens.
Pneumatic Actuation
- Required Air Pressure: 5 to 6 bar (72.5 to 87 psi).
- Air Quality: Filtered (40 μm), free of water and oil.
- Air Line Minimum Nominal Diameter: 4 mm (0.16").
- Note: A pressure-reducing valve is recommended if air pressure can exceed 6 bar (87 psi). A pneumatic throttle is recommended for smoother operation.
Rinse Water Connection
- Pressure: 2 to max. 6 bar (30 to max. 87 psi).
- Requirements: Check valve and filter (100 μm) at the inlet side when using water. Chemically resistant ON/OFF valve and outlet valve recommended when using cleaning solutions with pneumatic actuation.
- Note: A pressure-reducing valve is recommended if water pressure can exceed 6 bar (87 psi).
Disclaimer
This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.