Datasheet summary
CleanFit CPA875
Engineering summary from PDF text extraction for CleanFit CPA875. Verify every value with the OEM datasheet.
Summary
Product Identification
The Endress+Hauser Cleanfit CPA875 is a modular retractable process assembly designed for in-line measurement with standard 12 mm sensors. It is suitable for sterile and hygienic applications, including parameters such as pH, ORP, oxygen, and NIR.
Application Industries
- Food and beverages
- Biotechnology
- Life sciences
- Special chemicals
Key Benefits
- High availability with minimal maintenance.
- Reliable measurements and accurate results.
- Enhanced product quality.
- Modular design for investment security.
- EHEDG-certified assembly (process connection and service chamber).
- Certified to FDA and USP Class VI.
- Available with 3-A certificate.
Functionality
The Cleanfit CPA875 allows for sensor removal, cleaning, sterilization, or calibration/adjustment without interrupting the process. It can be installed in both vessels and pipes.
Design and Options
- Drive Type: Manual or pneumatic drive.
- Manual drive supports process pressures up to 8 bar (116 psi).
- Pneumatic drive supports process pressures up to 16 bar (232 psi).
- Chamber Systems:
- Single-chamber system (service chamber).
- Double-chamber system (inner and front service chambers).
- Electrode Guide Strokes:
- 36 mm (e.g., for flow housing).
- 78 mm (e.g., for installation in vessels).
- Process Connections: Various common process connections are available, including Clamp, Aseptic DIN 11864, BioControl, BioConnect, Dairy fitting, ISO228 Thread, and Varivent.
Safety Features
- Locking Mechanism: Prevents movement from service to measuring position if the sensor is not installed.
- Limit Position Locking:
- Pneumatic drive: Remains in the selected position if compressed air fails.
- Manual drive: Features an unlocking button for position locking.
- Sensor Protection: Protection cap prevents sensor removal in the measuring position.
- Non-rotating Sensor Guide: Ensures precise sensor positioning.
- Limit Position Detection: Inductively detects service and measuring positions (can be retrofitted for pneumatic drives).
Cleaning and Sterilization
- Medium drains completely from chambers when mounted at an angle up to 15° to the horizontal.
- Special aseptic seals are available without cleanable openings (not for NA process connection).
- EHEDG-certified for cleanability and sterilizability.
- Double-chamber system offers contamination-free insertion/retraction and safe separation between process and service chambers.
Materials
- Seals: EPDM-FDA (USP Class VI), FKM-FDA (USP Class VI), FFKM-FDA (USP Class VI). All FDA-certified and meet USP Class VI.
- Wetted Parts (Immersion Tube, Process Connection, Service Chamber): Stainless steel 1.4435 (AISI 316L) with surface roughness Ra < 0.76 µm or optionally Ra < 0.38 µm (immersion tube only).
- Non-wetted parts: Stainless steel 1.4301 (AISI 304) or 1.4404 (AISI 316L), plastics PPS CF15, PBT, PP.
Operating Limits
- Ambient Temperature Range: -10 to +70 °C (+10 to +160 °F).
- Storage Temperature: -10 to +70 °C (+10 to +160 °F).
- Process Temperature Range: –10 to 140 °C (14 to 284 °F).
- Process Pressure Range:
- Pneumatic drive: 16 bar (232 psi) up to 140 °C (284 °F).
- Manual drive: 8 bar (116 psi) up to 140 °C (284 °F).
Mechanical / Hydraulic
- Weight:
- Pneumatic drive: 3.8 to 6 kg (8.4 to 13.2 lbs).
- Manual drive: 3 to 4.5 kg (6.6 to 9.9 lbs).
- Rinse Chamber Volumes:
- Single chamber, short stroke: 20.94 cm³ (1.28 in³) max., 10.51 cm³ (0.64 in³) min.
- Single chamber, long stroke: 42.97 cm³ (2.62 in³) max., 20.77 cm³ (1.27 in³) min.
- Double chamber (front): 18.53 cm³ (1.13 in³) max., 9.80 cm³ (0.6 in³) min.
- Double chamber (rear): 77.49 cm³ (4.72 in³) max., 47.04 cm³ (2.87 in³) min.
- Double chamber (total): 96.02 cm³ (5.87 in³) max., 56.84 cm³ (3.47 in³) min.
- Necessary Mounting Distance for Sensor Replacement (XA):
- 425 mm (16.73") for 225 mm sensors.
- 610 mm (24.02") for 360 mm sensors.
Electrical
- Limit Position Switches:
- Switching element function: NAMUR NC contact (inductive).
- Switching distance: 1.5 mm (0.06").
- Nominal voltage: 8 V.
- Switching frequency: 0 to 5000 Hz.
Certificates and Approvals
- CRN (Canadian Registration Number)
- Pharma CoC
- Biological reactivity (USP Class VI)
- EHEDG (TYP EL Class I, and EL aseptic class I for double-chamber)
- ASME BPE Standard 2012 compliance
- FDA listed materials
- EC VO 1935/2004 compliance
Accessories
A range of accessories is available, including weld-in adapters, dummy plugs, sensor dummies, seal kits, cables for limit switches, installation tools, and lubricants. Device-specific sensors for pH, ORP, ISFET, conductivity, oxygen, and NIR absorption are also listed.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.