Back to Endress + Hauser Liquid analysis
Endress+Hauser logo

Datasheet summary

CPS31D

Engineering summary from PDF text extraction for CPS31D. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product Names: Memosens CPS31D, Ceratex CPS31
  • Type: pH sensors for drinking water and swimming pool water.
  • Technology: Digital with Memosens technology or analog.

Application

  • Drinking water
  • Swimming pool water
  • pH compensation during the measurement of free chlorine.

Features and Benefits

  • Gel-filled reference system with ceramic junction.
  • One or three ceramic junctions.
  • Optional salt storage for extended operating life.
  • Very low ion depletion of electrolyte.
  • Gel filling, no electrolyte refill required.
  • Memosens plug-in head or ESA or GSA plug-in head.

Memosens Technology Advantages

  • Maximum process safety with non-contact, inductive signal transmission.
  • Data security thanks to digital data transmission.
  • Easy to use: sensor data saved in the sensor.
  • Predictive maintenance enabled by recording sensor load data.

Input

Measured variables

  • pH value
  • Temperature

Measuring range

  • pH: 1 to 12
  • Temperature: 0 to 80 °C (32 to 176 °F)

Conductivity

  • Min. 100 µS/cm
  • Min. 50 µS/cm for "AC" version (three junctions)

Operating limits

Process temperature range

  • 0 to 80 °C (32 to 176 °F)

Process pressure range

  • 1 to 4 bar (14.5 to 58 psi) (abs.)

Operating temperature range (environment)

  • -15 °C to 50 °C (5 °F to 122 °F) (Note: Sensor must not be used below -15 °C (5 °F) due to risk of frost damage.)

Storage temperature

  • 0 to 50 °C (32 to 122 °F)

Electrical

Electrical connection

  • Memosens sensors use a Memosens plug-in head.
  • Analog sensors can use ESA or GSA plug-in heads with measuring cables CPK9 or CPK1.

Mechanical construction

Design, dimensions

  • Overall length: 120 mm (4.70 in)
  • Shaft diameter: ø 12 mm (0.47 in)

Weight

  • 0.1 kg (0.2 lb)

Materials

  • Electrode shaft: Glass to suit process
  • pH membrane glass: Type A
  • Metal lead: Ag/AgCl
  • Diaphragm: Ceramic
  • Temperature sensor: NTC 30K (CPS31D), Pt 100 (CPS31)
  • Plug-in head: Memosens (CPS31D), ESA, GSA (CPS31)

Process connections

  • Pg 13.5

Reference system

  • Ag/AgCl, gel, 3M KCl
  • Optional: Salt ring, KCl saturated

Installation

  • Do not install sensors upside-down.
  • Angle of inclination from the horizontal must be at least 15°.
  • Hand-tighten sensor with a torque of 3 Nm (2.21 lbf ft) when installing in Endress+Hauser assemblies.

Environment

Degree of protection

  • GSA plug-in head: IP 67
  • ESA plug-in head: IP 68 (1 m (3.3 ft) water column, 50 °C (120 °F), 168 h)
  • Memosens plug-in head: IP 68 (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)

Electromagnetic compatibility (EMC)

  • Interference emission and interference immunity as per EN 61326:2012

Disclaimer This extracted information may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.