Summary
Product identification
- Product Names: Memosens CPS31D, Ceratex CPS31
- Type: pH sensors for drinking water and swimming pool water.
- Technology: Digital with Memosens technology or analog.
Application
- Drinking water
- Swimming pool water
- pH compensation during the measurement of free chlorine.
Features and Benefits
- Gel-filled reference system with ceramic junction.
- One or three ceramic junctions.
- Optional salt storage for extended operating life.
- Very low ion depletion of electrolyte.
- Gel filling, no electrolyte refill required.
- Memosens plug-in head or ESA or GSA plug-in head.
Memosens Technology Advantages
- Maximum process safety with non-contact, inductive signal transmission.
- Data security thanks to digital data transmission.
- Easy to use: sensor data saved in the sensor.
- Predictive maintenance enabled by recording sensor load data.
Input
Measured variables
- pH value
- Temperature
Measuring range
- pH: 1 to 12
- Temperature: 0 to 80 °C (32 to 176 °F)
Conductivity
- Min. 100 µS/cm
- Min. 50 µS/cm for "AC" version (three junctions)
Operating limits
Process temperature range
- 0 to 80 °C (32 to 176 °F)
Process pressure range
- 1 to 4 bar (14.5 to 58 psi) (abs.)
Operating temperature range (environment)
- -15 °C to 50 °C (5 °F to 122 °F) (Note: Sensor must not be used below -15 °C (5 °F) due to risk of frost damage.)
Storage temperature
- 0 to 50 °C (32 to 122 °F)
Electrical
Electrical connection
- Memosens sensors use a Memosens plug-in head.
- Analog sensors can use ESA or GSA plug-in heads with measuring cables CPK9 or CPK1.
Mechanical construction
Design, dimensions
- Overall length: 120 mm (4.70 in)
- Shaft diameter: ø 12 mm (0.47 in)
Weight
- 0.1 kg (0.2 lb)
Materials
- Electrode shaft: Glass to suit process
- pH membrane glass: Type A
- Metal lead: Ag/AgCl
- Diaphragm: Ceramic
- Temperature sensor: NTC 30K (CPS31D), Pt 100 (CPS31)
- Plug-in head: Memosens (CPS31D), ESA, GSA (CPS31)
Process connections
- Pg 13.5
Reference system
- Ag/AgCl, gel, 3M KCl
- Optional: Salt ring, KCl saturated
Installation
- Do not install sensors upside-down.
- Angle of inclination from the horizontal must be at least 15°.
- Hand-tighten sensor with a torque of 3 Nm (2.21 lbf ft) when installing in Endress+Hauser assemblies.
Environment
Degree of protection
- GSA plug-in head: IP 67
- ESA plug-in head: IP 68 (1 m (3.3 ft) water column, 50 °C (120 °F), 168 h)
- Memosens plug-in head: IP 68 (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)
Electromagnetic compatibility (EMC)
- Interference emission and interference immunity as per EN 61326:2012