Datasheet summary
CTS1
Engineering summary from PDF text extraction for CTS1. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: CTS1 Temperature sensor
- Purpose: Temperature measurement, e.g., for automatic temperature compensation during pH measurement.
- Document ID: TI00123C/07/EN/05.20
- Part Number: 71490541
- Date: 2020-07-27
Application
- Can be combined with:
- pH combined sensor without a temperature sensor
- pH glass electrode and separate reference electrode
- Suitable for installation in immersion assemblies and flow assemblies.
Input
Temperature measurement
- Measured variable: Temperature via temperature-dependent resistor (TDR)
- Measuring range (in conjunction with Liquiline M CM42): –35 to 250 °C (–31 to 482 °F)
- Measuring range (in conjunction with Liquisys M CPM223/253): –50 to 150 °C (–58 to 302 °F)
Operating limits
Process temperature range
- –15 to 135 °C (5 to 275 °F)
Process pressure range
- 0.8 to 14 bar (11.6 to 203 psi)
Mechanical construction
- Process connection: Pg 13.5
- Dimensions:
- Shaft diameter: 12 mm (0.47 in)
- Total length: 170 mm (6.69 in)
- Insertion length: 120 mm (4.72 in)
- Weight: 185 g (6.5 oz)
- Wetted materials:
- Shaft: Glass to suit process
- Sealing ring: Viton
- Thrust collar: EPDM
- Materials not in contact with the medium: The potting compound in the sensor shaft contains the SVHC substance terphenyl, hydrogenated (CAS number 61788-32-7) with more than 0.1 % (w/w).
Electrical
- Measuring resistor: Pt100
- Measuring cable connection options:
- Coaxial plug-in head
- TOP68 plug-in head
- Fixed cable
- Double Pt100 version: Possible
- Ex zone version: Available for Ex zone 0.
- Certificates:
- CE mark
- ATEX II 1G EEx ia IIC T3/T4/T6
- FM Class I Div. 2 (in conjunction with Mypro CPM431 and Liquiline M CM42 transmitter)
- TÜV certificate for TOP68 plug-in head
Accessories
- Measuring cable CPK3
- Measuring cable CPK9
- pH single electrode CPS64
- pH reference electrode OrbiSint CPS13
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.