Datasheet summary
Turbimax CUA262
Engineering summary from PDF text extraction for Turbimax CUA262. Verify every value with the OEM datasheet.
Summary
Product identification
Product: CUA262 Flow assembly Purpose: For installation of the CUS52D turbidity sensor in process water skids and closed pipe systems. Manufacturer: Endress+Hauser Document: TI01152C/07/EN/02.19
Application
The CUA262 is designed for direct installation of the CUS52D turbidity sensor into stainless steel pipes. It is suitable for:
- Turbidity measurement at all process stages in water treatment skids.
- Turbidity measurement in closed stainless steel pipe systems.
- Turbidity monitoring in filter modules.
Benefits include direct process installation, no water loss, self-venting with vertical installation, and suitability for drinking water applications due to stainless steel material.
Function and system design
A complete measuring system comprises:
- Flow assembly Flowfit CUA262
- Sensor Turbimax CUS52D
- Transmitter (e.g., Liquiline CM442)
- Measuring cable
Installation
General Instructions
- The flow assembly should be installed vertically.
- Connect the inflow at the bottom end for upward flow in the pipe.
- Avoid buckles and loops in hose systems.
- Pay attention to sensor installation instructions regarding flow direction.
Wall Effects
Backscattering from the pipe wall can cause false measurements for turbidity values < 200 FNU.
- If wall effects occur, change the orientation of the sensor.
- Avoid sudden changes in cross-section; gradual changes should be located far from the sensor.
- Do not install the sensor directly downstream from a bend; position it as far as possible from the bend.
- For reflective materials like stainless steel, pipe diameter should be at least 100 mm (4 in). On-site adjustment of installation position is recommended.
- Stainless steel pipes with diameter > DN 300 exhibit minimal wall effects.
Mounting the flow assembly
- Attach the flow assembly to the pipe as a weld-on element.
- Install the assembly vertically.
- Connect the inflow at the bottom end (upward flow).
Mounting the sensor
- Only insert turbidity sensors with a 2" clamp into the assembly.
- Install the sensor such that the optical windows are aligned against the direction of flow.
- Use the installation marking on the sensor for correct orientation.
Environment
- Ambient temperature range: 0 to 60 °C (32 to 140 °F)
- Storage temperature: 0 to 60 °C (32 to 140 °F), in original packaging
Process
- Process temperature range: 0 to 90 °C (32 to 194 °F)
- Process pressure range: 0 to 10 bar (0 to 145 psi)
Mechanical construction
- Dimensions: See diagram (Ø 64 / 2.52 in, Ø 100 / 3.94 in, Ø 104 / 4.09 in, 200 / 7.87 in, 164 / 6.46 in). DN 50.
- Weight: 1.11 kg (2.45 lb)
- Materials:
- Assembly housing: Stainless steel 1.4404 (AISI 316 L)
- Seals: EPDM
- Dummy cover: Stainless steel 1.4404 (AISI 316 L)
Certificates and approvals
- DRGL- 2014/68/EU / PED- 2014/68/EU: Manufactured according to good engineering practice, Article 4, Paragraph 3 of the Pressure Equipment Directive 2014/68/EU. Not required to bear the CE label.
Ordering information
Scope of delivery:
- 1 Flowfit CUA262 flow assembly
- 1 clamp seal and locking clamp
- 1 x Operating Instructions
Product page: www.endress.com/cua262
Accessories:
- Dummy cover for clamp connection; 1 pc (Order number: 71242180)
- Ultrasonic cleaning system CYR52