Datasheet summary
CYA680
Engineering summary from PDF text extraction for CYA680. Verify every value with the OEM datasheet.
Summary
Product identification
CYA680 Flow Assembly for the Life Sciences and Chemical Industry.
Application
- For 12 mm sensors with Pg 13.5
- Installations with small pipe diameters
- Hygienic and sterile applications
- Chromatography
- Filtrations
- Reverse osmosis
Benefits
- Fast measured values in small sample volumes for maximum product yield.
- High product safety: Flow assembly is suitable for CIP and SIP.
- PVDF version is resistant to aggressive and corrosive media.
- Complies with Life Sciences industry requirements:
- Certified biocompatibility (USP <87>, USP <88>).
- FDA-listed seals.
- Hygienic, electropolished surfaces (Ra ≤ 0.38 µm).
Function and system design
A complete measuring system comprises:
- Transmitter (e.g., Liquiline CM44P)
- One or two 12 mm sensors (e.g., CLS82D and/or CPS71D)
- Flow assembly CYA680
- Measuring cable (e.g., CYK10)
Mounting and Operating Limits
The CYA680 Flow Assembly is available in various nominal diameters and materials. It can be installed in horizontal and vertical pipes.
Process Connection
| Material | Nominal Diameter | Nominal Pressure | Temperature |
|---|---|---|---|
| 316L Tri-Clamp | 0.25 to 2" | 16 bar (230 psi) | 0 to 130 °C (32 to 266 °F) |
| 1.4435 Tri-Clamp | 0.25 to 2" | 16 bar (230 psi) | 0 to 130 °C (32 to 266 °F) |
| PVDF Tri-Clamp (Kynar) | 0.25", 0.5", 0.75" | 4 bar (58 psi) | 0 to 130 °C (32 to 266 °F) |
Observe the maximum permitted process temperature and pressure of the sensor.
Mechanical construction
Dimensions
| Flange | A (Internal diameter) | B (Flange diameter) | C (Sensor holder pH) | D (Sensor holder conductivity) |
|---|---|---|---|---|
| ¼" Tri-Clamp | 4.57 mm (0.18") | 25 mm (0.984") | 138.4 mm (5.45") | 143.4 mm (5.65') |
| ½" Tri-Clamp | 9.53 mm (0.375") | 25 mm (0.984") | 138.4 mm (5.45") | 143.4 mm (5.65') |
| ¾" Tri-Clamp | 15.24 mm (0.60") | 25 mm (0.984") | 138.4 mm (5.45") | 143.4 mm (5.65') |
| 1" Tri-Clamp | 22.1 mm (0.87") | 50.39 mm (1.984") | 144 mm (5.67") | 149 mm (5.87') |
| 1 ½" Tri-Clamp | 34.44 mm (1.356") | 50.39 mm (1.984") | 144 mm (5.67") | 149 mm (5.87') |
| 2" Tri-Clamp | 45 mm (1.856") | 63.91 mm (2.516") | 150 mm (5.92") | 155 mm (6.10') |
Materials
- Flow assembly: Stainless steel 1.4404/1.4435 or PVDF.
- O-rings: EPDM FDA, KALREZ FDA, VITON FDA.
- PVDF is not suitable for all hazardous areas.
Weight (Stainless steel version)
| Flange | 1 Sensor location | 2 Sensor locations |
|---|---|---|
| ¼" Tri-Clamp | Approx. 1.30 kg (2.86 lbs) | Approx. 1.65 kg (3.64 lbs) |
| 2" Tri-Clamp | Approx. 2.20 kg (4.85 lbs) | Approx. 2.55 kg (5.63 lbs) |
Ordering information
Configure the device using the Product Configurator on www.endress.com/cya680.
Scope of delivery
- Assembly in the version ordered.
- Operating Instructions.
Accessories
- pH sensors: Memosens CPS61E, Ceragel CPS71, Memosens CPS71E.
- ORP sensors: Memosens CPS62E.
- pH-ISFET sensors: Memosens CPS47E, Memosens CPS77E.
- Conductivity sensors: Memosens CLS82E.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.