Datasheet summary
FlowFit COA250
Engineering summary from PDF text extraction for FlowFit COA250. Verify every value with the OEM datasheet.
Summary
Product identification
Flowfit COA250 is a flow assembly designed for oxygen sensors with a diameter of 40 mm (1.57 inch). It is suitable for applications where direct sensor immersion is not feasible or desired.
Main areas of application:
- Wastewater treatment
- Drinking water
- Monitoring of surface waters
- Analysis stations
- Process applications
Benefits:
- Simple wall or unsupported pipe installation.
- Automatic self-venting due to bottom inlet/top outlet design.
- Minimized internal volume ensures quick oxygen sensor response.
- Construction is resistant to clogging.
- Oxygen sensor can be easily removed for calibration.
- Allows connection of a spray head for sensor membrane cleaning.
Function and system design
Measuring system components:
- Flowfit COA250
- Oxygen sensor (e.g., COS31, COS41, COS61, COS51D)
- Measuring cable (e.g., CYK10)
- Transmitter (e.g., Liquiline M CM42)
- Optional: Junction box for cable extension (e.g., RM junction box)
Operating limits
Environment:
- Ambient temperature range: 0 ... 50 °C (32 ... 120 °F)
- Storage temperature: 0 ... 50 °C (32 ... 120 °F)
Process:
- Process temperature range: 0 to 50 °C (32 to 120 °F), non-freezing
- Process pressure:
- max. 6 bar (87 psi) at 20 °C (68 °F)
- max. 4 bar (58 psi) at 50 °C (122 °F)
Mechanical / hydraulic
Design and dimensions:
- Weight: 0.5 to 0.8 kg (1.1 to 1.8 lbs), depending on process connection.
- Materials: EPDM (sealings), PVC.
- Process connection: G¾ (for usual fittings).
Suitable sensors:
- Amperometric and optical sensors with Ø40 mm (1.57 inch) and G1 thread.
Installation lengths:
- COA250-A (short sensor sleeve): 110 to 150 mm (4.3 to 5.9 inch)
- Example sensors: COS 3 / 3HD / 4 / 4HD / 41 / 51D
- COA250-B (long sensor sleeve): 140 to 190 mm (5.5 to 7.5 inch)
- Example sensors: COS 31 / 61
Accessories
Accessory kits:
- Long sensor sleeve (Order No. 51507172): For later installation of sensor COS31 / 61 into COA250-A assembly.
- Pipe clamp COY250 (Order No. 50065132): For panel or wall mounting.
Cleaning systems:
- Chemoclean CUR3: Spray head for flow assemblies CUA250 and COA250 (Order No. CUR3-1).
- Chemoclean: Injector CYR10 and programme sequencer CYR20.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.