ISEmax CAS40D

Engineering summary from PDF text extraction for ISEmax CAS40D. Verify every value with the OEM datasheet.

Summary

Product Identification

Product Name: ISEmax CAS40D Product Type: Ion-selective sensor for continuous measurement of ammonium, nitrate, and other ions. Application: Online measurement of nutrient parameters in municipal wastewater treatment plants.

Function and System Design

Measuring Principle

The sensor utilizes an ion-selective electrode (ISE) with a membrane containing an ionophore. This membrane facilitates the selective migration of specific ions (e.g., ammonium or nitrate) to the electrode, creating a potential proportional to the logarithm of the ion concentration. This potential is measured against a reference electrode and converted to concentration using the Nernst equation. Color and turbidity do not affect the measurement.

Interference

Potential measurement errors can occur due to interference ions.

  • Potassium (K⁺) can cause higher measured values when measuring ammonium due to chemical similarity.
  • Chloride (Cl⁻) can cause higher measured values when measuring nitrate. To mitigate these errors, optional electrodes can be used to measure and compensate for potassium or chloride concentrations.

Measuring System Components

A complete measuring system typically includes:

  • Sensor CAS40D
  • Ion-selective electrode(s) for ammonium, nitrate, potassium, or chloride
  • pH glass electrode (Orbisint CPS11-1AS2GSA)
  • Temperature sensor (CTS1)
  • Transmitter (Liquiline CM44x)
  • Optional: Assembly holder, weather protection cover, compressed air generator.

Input Parameters

Measured Values

Depending on the sensor version, the following can be measured:

  • Ammonium (NH₄-N, NH₄⁺) in mg/l
  • Nitrate (NO₃-N, NO₃⁻) in mg/l
  • Potassium (K⁺) in mg/l
  • Chloride (Cl⁻) in mg/l
  • pH value
  • Temperature

Measuring Ranges

  • Ammonium: 0.1 to 1000 mg/l (NH₄-N)
  • Nitrate: 0.1 to 1000 mg/l (NO₃-N)
  • Potassium: 1 to 1000 mg/l
  • Chloride: 1 to 1000 mg/l

Electrical Connection

  • Options: M12 connector or fixed cable with ferrules for connection to Liquiline CM44x transmitter.
  • Maximum Cable Length: 100 m (328 ft).

Performance Characteristics

  • Response Time (t90) of Ion-Selective Sensors: < 2 min (for a change between 0.5 and 1 mmol/l at 25 °C).
  • Measured Error: ± 5 % of the measured value ± 0.2 mg/l.
  • Repeatability: ± 3 % of the display value.
  • Compensation:
    • Ammonium: Temperature (2-40 °C), pH (5-8.3 without compensation, 5-10 with compensation), Potassium (1-1000 mg/l).
    • Nitrate: Temperature (2-40 °C), pH (2-12), Chloride (10-1000 mg/l).
    • Potassium: Temperature (2-40 °C), pH (2-12).
    • Chloride: Temperature (2-40 °C), pH (1-10).

Maintenance and Durability

  • Max. Operating Life (Membrane and Electrolyte): Approximately 0.5 years in use, 2 years in storage.
  • Automatic Cleaning:
    • Medium: Compressed Air
    • Pressure: 3 to 3.5 bar (45 to 50 psi)
    • Volume per Cycle: 3 to 4 l (0.8 to 1 US gal)
    • Duration per Cycle: 4 to 15 s
    • Intervals (at T > 10 °C):
      • Sludge activation inlet: 15 s cleaning, 30 min pause.
      • Sludge activation: 15 s cleaning, 1 hr pause.

Environmental and Process Limits

  • Ambient Temperature: –20 to 50 °C (–4 to 122 °F).
  • Storage Temperature: 2 to 40 °C (36 to 104 °F).
  • Degree of Protection: IP68.
  • Electromagnetic Compatibility: As per EN 61 326, Namur NE21.
  • Process Temperature: 2 to 40 °C (36 to 104 °F).
  • Process Pressure: Max. 400 mbar overpressure.
  • pH Value of Medium:
    • Ammonium: pH 5 to 8.3 (without compensation), pH 5 to 10 (with compensation).
    • Nitrate, Potassium: pH 2 to 12.
    • Chloride: pH 1 to 10.

Mechanical Construction

  • Dimensions: See diagram on page 6 for detailed dimensions in mm (inch).
  • Weight: Approx. 3.5 kg (7.7 lbs).

Materials

  • Sensor Body: POM
  • Protective Cage: POM
  • Electrode Holder: Silicone
  • Radial Seal: EPDM
  • O-rings (ISE holder): VITON
  • O-rings (air nozzle): PP
  • Sensor Pipe: Stainless steel
  • Sensor Head: Glass
  • Ion-Selective Electrodes (Membrane Cap): POM
  • Ion-Selective Electrodes (Shaft): POM
  • Ion-Selective Electrodes (Membrane): PVC, plasticizer

Compressed Air Cleaning Unit

  • Operating Interval: Max. 3 minutes cleaning, with a break at least six times the cleaning duration.
  • Power Supply Options: 230 V or 115 V.
  • Conveying Rate (atmospheric pressure): 50 l/min (13.2 gal/min).
  • Power Consumption: 240 W.
  • Current Consumption: 1.3 A.