Datasheet summary
Liquiline CM14
Engineering summary from PDF text extraction for Liquiline CM14. Verify every value with the OEM datasheet.
Summary
Product Identification
The Liquiline CM14 is a four-wire transmitter designed for panel or cabinet installation. It supports pH/ORP, conductivity, and oxygen measurements.
Applications:
- Wastewater treatment
- Water treatment and drinking water monitoring
- Ion exchangers
- Surface water monitoring
Key Features:
- Compact 48 x 96 mm panel-mounted housing
- Easy to use
- Memosens technology enabled
- Supports pre-calibrated sensors
- Active indication of cable break for safety
- Two contacts for limit switching
- Second current output available for temperature
Input
- Measured variables: pH value, ORP, conductivity, oxygen (dependent on connected sensor).
- Measuring ranges: Dependent on connected sensor.
- Types of input: Digital sensor inputs for sensors with Memosens protocol.
- Cable specification:
- Cable type: Memosens data cable CYK10 or fixed sensor cable.
- Max. cable length: 100 m (330 ft).
Output
- Output signal: 2 x 0/4 to 20 mA, active, galvanically isolated.
- Load: Max. 500 Ω.
- Linearization/transmission behavior: Linear.
- Alarm output: Open collector, normally closed, opens on fault.
- Max. current: 200 mA.
- Max. voltage: 30 V DC.
- Current outputs:
- Span: 0 to 23 mA.
- Signal characteristic: Linear.
- Relay outputs: 2 changeover contacts.
- Switching capacity: max. 3 A @ 24 V DC / 253 V AC.
- Minimum: 100 mW (5 V / 10 mA).
Electrical Specification
- Output voltage: Max. 24 V.
- Test voltage: 500 V.
- Cable specification (for outputs/relays): Max. 1.5 mm² (16 AWG) for signal, Max. 2.5 mm² (14 AWG) for relays.
- Power supply: Universal power supply 24 to 230 V AC/DC (±20 % / +10 %) 50/60 Hz.
- Power consumption: Max. 13.8 VA / 6.6 W.
Performance Characteristics
- Response time (current outputs): t90 = max. 500 ms for 0 to 20 mA.
- Reference temperature: 25 °C (77 °F).
- Measured error for sensor inputs: See documentation of the connected sensor.
- Current output resolution: > 13 bit.
- Repeatability: See documentation of the connected sensor.
Installation
- Mounting location: Panel mount, cutout 92 x 45 mm (3.62 x 1.77 in).
- Max. panel thickness: 26 mm (1 in).
- Orientation: Determined by display legibility; max. viewing angle range of ±45°.
Environment
- Ambient temperature: –10 to +60 °C (14 to 140 °F).
- Storage temperature: -40 to +85 °C (-40 to 185 °F).
- Humidity: 5 to 85 %, non-condensing.
- Operating height: < 2000 m (6561 ft) above MSL.
- Degree of protection:
- Front: IP65 / NEMA 4X.
- Casing: IP20.
- Electromagnetic compatibility: Interference emission and immunity as per EN 61326-1:2013, Class A for Industry.
Mechanical Construction
- Dimensions: 48 mm (width) x 96 mm (height) x 118.6 mm (depth).
- Weight: 0.3 kg (0.66 lbs).
- Materials:
- Housing, casing: Polycarbonate.
- Front foil: Polyester, UV-resistant.
- Terminals: Max. 2.5 mm² (22-14 AWG) for mains and relay connections (torque 0.4 Nm).
Human Interface
- Operating elements: LC display for measurements and configuration, status LEDs (power, alarm, relays), and operating keys.