Datasheet summary
Liquiline System CA80AL
Engineering summary from PDF text extraction for Liquiline System CA80AL. Verify every value with the OEM datasheet.
Summary
Product Identification
The Liquiline System CA80AL is a wet-chemical analyzer designed for the near-continuous determination of aluminum concentration in liquid media. It utilizes a colorimetric measuring principle.
Application
This analyzer is suitable for:
- Controlling phosphate precipitation in wastewater treatment plants.
- Monitoring precipitants in drinking water and wastewater treatment plants.
The "Cabinet, outdoor version" is designed for outdoor installation, mounting, and operation.
Key Features and Benefits
- Integrated controller with up to 2 measuring channels and digital Memosens technology.
- Expandable: Can be upgraded to a measuring station by connecting up to 4 Memosens sensors.
- Digital Fieldbuses: Supports protocols such as PROFINET, PROFIBUS DP, Modbus TCP, Modbus RS485, and Ethernet IP, along with an integrated web server.
- Easy Maintenance: Features simple, tool-free maintenance.
- Reliability: Memosens technology offers non-contact, digital signal transmission for improved safety and reliability, allowing for lab calibration.
- Data Logging: Includes integrated ring or stack memories for recording analog values, events, and analyzer data.
- Mathematical Functions: Capable of calculating up to 6 "virtual" process values using mathematical and Boolean operations.
Measuring Principle
The analyzer employs the Pyrocatechol violet method according to DIN ISO 10566. Catechol violet reacts with Al(III) in a buffered solution (pH 5.8-6.0) to form a blue dye. Absorption is measured at 568 nm, with absorption being proportional to aluminum concentration. A reference measurement is performed to compensate for interference influences.
Cross-Sensitivity
The following ions were tested and showed no observed cross-sensitivities up to the specified concentrations:
- 1000 mg/l (ppm): Cl⁻, SO₄²⁻
- 500 mg/l (ppm): Ca²⁺, K⁺, Na⁺
- 100 mg/l (ppm): NO₃⁻
- 10 mg/l (ppm): Zn²⁺
- 2 mg/l (ppm): F⁻
Measuring System Components
A complete measuring system includes the analyzer, reagents, cleaner, standard solutions, and optionally, sample conditioning units like:
- Liquiline System CAT810: Microfiltration (50 µm sieve filter).
- Liquiline System CAT820: Membrane filtration (0.1 µm or 0.4 µm ceramic membrane, or 0.04 µm plate filter).
- Liquiline System CAT860: Membrane filtration (0.1 µm or 0.4 µm ceramic membrane filter candle) with automatic backflush.
Input
- Measured Variables: Al [µg/l, ppb]
- Measuring Range: 15 to 1000 µg/l (ppb) Al. An optional dilution function allows for configurable measuring ranges up to 20,000 µg/l (ppb) with dilution factors of 5, 10, or 20.
- Channels: 1 or 2 measuring channels (analyzer main parameter).
- Optional Inputs: 1 to 4 digital sensor inputs (Memosens protocol), analog current inputs, binary inputs.
- Input Signal: 2 x 0/4 to 20 mA (optional), passive, potentially isolated.
- Memosens Sensor Cable Length: Max. 100 m (330 ft).
Output
- Output Signal:
- Standard: 2 x 0/4 to 20 mA, active, potentially isolated.
- Optional: 4 x or 6 x 0/4 to 20 mA, active, potentially isolated.
- Binary outputs.
- Signal on Alarm: Adjustable, as per NAMUR Recommendation NE 43 (failure current 0 to 23 mA or 2.4 to 23 mA).
- Load (Current Outputs): Max. 500 Ω.
- Relay Outputs:
- 1 x alarm relay (changeover contact).
- Optional: 2 or 4 additional changeover contacts.
- Maximum load: 0.5 A (alarm relay), 2.0 A (other relays).
Electrical Specifications
- Supply Voltage: 100 to 120 V AC / 200 to 240 V AC or 24 V DC (not available for "Outdoor" version). Frequency: 50 or 60 Hz.
- Power Consumption: Varies by version (e.g., 130 VA + 660 VA per hose heater for non-outdoor versions, 680 VA + 660 VA per hose heater for outdoor versions with cooling system).
- Fuse: 5 x 20 mm 10 A/250 V fine-wire fuse for hose trace heating system.
Mechanical Construction
- Dimensions: See figures in the datasheet for specific dimensions of closed, open, and base-mounted versions.
- Weight: Cabinet version: 39.5 kg (87.1 lbs); Open installation: 31.5 kg (69.45 lbs); Analyzer stand: 72.5 kg (159.8 lbs).
- Materials: Various plastics (ASA, PC, PP), stainless steel, glass, and elastomers are used for different components.
- Process Connection: Plug-in connector for rigid hoses (OD 4 mm) or hose barb for flexible hoses (ID 1.6 mm) for sample inlet. Hose barb for flexible hoses (ID 13 mm) for outlet.
Environment
- Ambient Temperature Range:
- Standard versions: +5 to +40 °C (41 to 104 °F).
- Outdoor version: -20 to +40 °C (-4 to 104 °F).
- Storage Temperature: –20 to 60 °C (–4 to 140 °F).
- Relative Humidity: 10 to 95 %, non-condensing.
- Degree of Protection: IP55 (cabinet, analyzer stand), TYPE 3R (cabinet, analyzer stand).
- Sample Temperature: 4 to 40 °C (39 to 104 °F).
- Sample Consistency: Low solids content (turbidity < 50 NTU), aqueous, homogenized.
- pH Value of Sample: < 6.
Communication Protocols
- Fieldbus Systems: PROFIBUS DP, Modbus TCP or RS485, PROFINET, EtherNet/IP.
- Web Server: For remote access to configuration, measured values, and diagnostics.
Display
- Type: Graphic display with 240 x 160 pixel resolution.
- Features: Backlight with switch-off function, red display background for alarms, transflective display technology.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.