Engineering summary from PDF text extraction for Liquiline System CA80CR. Verify every value with the OEM datasheet.
Summary
Product identification
The Liquiline System CA80CR is a wet-chemical analyzer designed for the near-continuous determination of chromate (Cr (VI)) in liquid media.
Application
This analyzer is suitable for:
- Optimizing the treatment efficiency of industrial wastewater treatment plants
- Monitoring drinking water
- Monitoring membrane filtration systems
- Monitoring limit values for industrial wastewater discharge
A "Cabinet, outdoor version" product variant is available for outdoor setup, mounting, and operation.
Key Features
- Integrated controller with up to 2 measuring channels
- Digital Memosens technology
- Easy upgrade to a measuring station by connecting up to 4 Memosens sensors
- Two-channel device option
- Digital fieldbuses supported: PROFINET, PROFIBUS DP, Modbus TCP, Modbus RS485, and Ethernet IP
- Web server functionality
- Simple, tool-free maintenance
Measuring Principle
The analyzer uses a colorimetric measuring principle. A specific color reagent is added to a prepared sample in a mixing/reaction chamber. The resulting color change is measured by a multispectral photometer based on light absorption. A reference measurement is performed to compensate for interference influences.
Chromium and Chromate
The document notes that chromium exists in trivalent (Cr³⁺) and hexavalent (chromate, CrO₄²⁻ and dichromate, Cr₂O₇²⁻) forms. Cr(VI) is a known carcinogen, while Cr(III) has low toxicity. Limit values for chromate concentration are stipulated by national standards (e.g., German drinking water ordinance for total chromium at 50 µg/l, and limits for industrial wastewater like 2 mg/l for tanneries and 0.1 mg/l for electroplating plants).
Photometric Determination of Chromate
The analyzer utilizes the diphenylcarbazide method. Chromate (Cr(VI)) reacts with diphenylcarbazide in an acidic solution to form a red-violet chelate complex. This method specifically records Cr(VI) ions.
Cross-sensitivity
The document lists ions and substances checked for cross-sensitivity with specified concentrations. No cross-sensitivities were observed up to these limits. Some ions (Fe³⁺, Fe²⁺, Cu²⁺, Sn²⁺) were noted to decrease the measured value.
- Cl⁻: 15,000 mg/l (ppm)
- Na⁺: 10,000 mg/l (ppm)
- K⁺, Ca²⁺: 500 mg/l (ppm)
- Ag⁺: 100 mg/l (ppm)
- Cd²⁺: 70 mg/l (ppm)
- SO₄²⁻, CO₃²⁻, NO₃⁻, Zn²⁺, Ni²⁺, CO₂²⁺: 50 mg/l (ppm)
- Hg²⁺: 5 mg/l (ppm)
- Turbidity: 50 NTU
- pH: 4 to 12
- Hardness: 70 °dH
Measuring System
A complete measuring system includes the Liquiline System CA80CR analyzer, reagents, cleaner, standard solution, and optionally, sample conditioning systems like Liquiline System CAT8x0 (microfiltration or membrane filtration).
Input
- Measured Variables: Chromate (Cr(VI)), CrO₄ [mg/l, µg/l, ppm, ppb]
- Measuring Range:
- CA80CR-**AF: 0.03 to 2.5 mg/l Cr(VI)
- CA80CR-**AG: 0.2 to 5.0 mg/l Cr(VI)
- Types of Input:
- 1 or 2 measuring channels (analyzer main parameter)
- 1 to 4 digital sensor inputs for Memosens protocol sensors (optional)
- Analog current inputs (optional)
- Binary inputs (optional)
- Input Signal:
- 2 x 0/4 to 20 mA (optional), passive, potentially isolated
- Current Input, Passive:
- Span: > 0 to 20 mA
- Signal Characteristic: Linear
- Internal Resistance: Non-linear
- Test Voltage: 500 V
- Cable Specification (for optional sensors with Memosens technology):
- Cable Type: Memosens data cable CYK10 or sensor fixed cable
- Cable Length: Max. 100 m (330 ft)
Output
- Output Signal:
- 2 x 0/4 to 20 mA, active, potentially isolated (standard)
- 4 x 0/4 to 20 mA, active, potentially isolated (with 2 additional analog outputs)
- 6 x 0/4 to 20 mA, active, potentially isolated (with 4 additional analog outputs)
- Binary outputs
- Signal on Alarm: Adjustable per NAMUR Recommendation NE 43. Failure current adjustable from 0 to 23 mA (in 0-20 mA range) or 2.4 to 23 mA (in 4-20 mA range). Factory setting for failure current is 21.5 mA.
- Load: Max. 500 Ω
- Current Outputs, Active:
- Span: 0 to 23 mA
- Signal Characteristic: Linear
- Output Voltage: Max. 24 V
- Test Voltage: 500 V
- Relay Outputs:
- 1 x single-pin changeover contact (alarm relay)
- 2 or 4 x single-pin changeover contacts (optional)
- Maximum load for alarm relay: 0.5 A
- Maximum load for other relays: 2.0 A
Protocol-specific Data
The device supports several communication protocols:
- PROFIBUS DP: Profile 3.02, supports up to 16 AI and 8 DI blocks, 4 AO and 8 DO blocks.
- Modbus RS485: RTU/ASCII protocol, supports function codes 03, 04, 06, 08, 16, 23.
- Modbus TCP: TCP port 502, supports function codes 03, 04, 06, 08, 16, 23. Configurable via DHCP or menu.
- Web server: Accessible via standard WiFi/WLAN/LAN/GSM or 3G routers with a user-defined IP address. TCP port 80.
- EtherNet/IP: ODVA certified. Supports CIP connections and explicit messages.
- PROFINET: PNIO Version 2.34, 100 MBit/s. Conformance Class B, Netload Class II.
Power Supply
- Supply Voltage:
- 100 to 120 V AC / 200 to 240 V AC or 24 V DC (24 V DC not available for "Outdoor" version)
- 50 or 60 Hz
- Power Consumption: Varies by version (e.g., 130 VA for non-outdoor versions without heating, up to 680 VA for outdoor versions with cooling). 24-V version: max. 105 W.
Mechanical Construction
- Dimensions: Drawings provided for closed version, open version, and with base.
- Weight: Varies by order configuration (e.g., Cabinet version: 39.5 kg, Analyzer stand: 72.5 kg).
- Materials:
- Parts not in contact with medium: ASA + PC, PP, Safety glass, EPP, Powder-coated sheet steel.
- Parts in contact with medium: PP, TPE, FKM, C-Flex (TPE), NORPRENE (PP).
Process Connection
- Sample Inlet: Plug-in connector for rigid hoses (OD 4 mm) or hose barb for flexible hoses (ID 1.6 mm).
- Dilution Water: Hose barb for flexible hoses (ID 3.2 mm).
- Outlet: Hose barb for flexible hoses (ID 13 mm).
Operability
- Operation Concept: Intuitive operation with navigator and soft keys, plain-text display.
- Display: Graphic display (240 x 160 pixel) with backlight and alarm indication.
Environment
- Ambient Temperature Range:
- Standard housing versions: +5 to +40 °C (41 to 104 °F)
- Outdoor version: -20 to +40 °C (-4 to 104 °F)
- Storage Temperature: –20 to 60 °C (–4 to 140 °F)
- Relative Humidity: 10 to 95 %, non-condensing
- Degree of Protection: IP55 (cabinet, analyzer stand), TYPE 3R (cabinet, analyzer stand)
- Electromagnetic Compatibility: EN 61326-1, class A for industrial areas.
- Electrical Safety: EN/IEC 61010-1:2010, Class I equipment.
- Pollution Degree: Pollution level 2.
Process
- Sample Temperature: 4 to 40 °C (39 to 104 °F)
- Consistency of the Sample: Low solids content (turbidity < 50 NTU), aqueous, homogenized.
- Sample Supply: Unpressurized.
Performance Characteristics
- Maximum Measured Error: 2 % of end of measuring range. With dilution and dilution factor 5, an additional 3.0 % of end of measuring range is added.
- Repeatability:
- Up to 0.5 mg/l (ppm) Cr(VI): ± 10 µg/l (ppb) Cr(VI)
- ≥ 0.5 mg/l (ppm) Cr(VI): ± 2 % of the display value
- With dilution and dilution factor 5: Additionally 1.5 % of end of measuring range.
- Measuring Interval: Continuous (approx. 5 min), adjustable from 10 min to 24 h.
- Sample Requirement:
- Without dilution module: 22 ml (0.74 fl oz)/measurement.
- With dilution module: Sample: 11 to 15 ml (0.37 to 0.51 fl oz)/measurement; Dilution water: 17 to 21 ml (0.57 to 0.71)/measurement.
- Reagent Requirement: Approx. 65 μl per reagent and measurement.
- Calibration Interval: 1 h to 90 days, depending on application and ambient conditions.
- Maintenance Interval: Every 3 to 6 months, depending on application.
Disclaimer: Extraction may miss figures, footnotes, or revisions; contractual data must match the OEM PDF revision used on the project.