Datasheet summary
Liquiline System CA80FE
Engineering summary from PDF text extraction for Liquiline System CA80FE. Verify every value with the OEM datasheet.
Summary
Product identification
The Liquiline System CA80FE is a wet-chemical analyzer designed for the near-continuous determination of iron concentration in liquid media.
Applications include:
- Monitoring residual dissolved iron after precipitation, flocculation, and filtration.
- Monitoring iron-rich process waters.
- Quality control of raw water for the food industry.
A "Cabinet, outdoor version" is available for outdoor installation.
Key Features
- Integrated controller with up to 2 measuring channels.
- Supports digital Memosens technology, allowing easy upgrade to a measuring station by connecting up to 4 Memosens sensors.
- Offers digital fieldbuses including PROFINET, PROFIBUS DP, Modbus TCP, Modbus RS485, and Ethernet IP.
- Includes an integrated web server.
- Designed for simple, tool-free maintenance.
Measuring Principle
The analyzer uses a colorimetric measuring principle. A specific color reagent is added to a prepared sample in a mixing/reaction chamber, causing a color change. A multispectral photometer measures light absorption at defined wavelengths, which is proportional to the iron concentration. A reference measurement is performed to compensate for interference.
The photometric determination of iron utilizes the Ferrozine method, reacting with iron (II) in a buffered solution. A reduction agent is included to convert Fe(III) to Fe(II) for total dissolved iron measurement. Absorption is measured at 568 nm.
Performance Characteristics
- Measuring ranges:
- CA80FE-**M2: 0.05 to 2.5 mg/l (ppm) Fe
- CA80FE-**M3: 0.1 to 5 mg/l (ppm) Fe
- Maximum measured error:
- CA80FE-AAM2: 0.04 mg/l (ppm) Fe within its range.
- CA80FE-AAM3: 0.1 mg/l (ppm) Fe within its range.
- With dilution factor 5: additionally 2.0 % of end of measuring range.
- Repeatability:
- < 0.25 mg/l (ppm): ± 0.02 mg/l (ppm) Fe
- 0.25 to 0.50 mg/l (ppm): ± 0.03 mg/l (ppm) Fe
-
0.50 mg/l (ppm): ± 0.06 mg/l (ppm) Fe
- With dilution factor 5: Additionally 1.0 % of end of measuring range.
- Measuring interval: Continuous (approx. 5 min), adjustable from 10 min to 24 h.
- Sample requirement:
- Without dilution module: 22 ml/measurement.
- With dilution module: 11 to 15 ml sample and 17 to 21 ml dilution water per measurement (depending on dilution factor).
- Reagent requirement: Approx. 65 μl per measurement (approx. 280 ml per month).
Input
- Measured variables: Dissolved iron Fe [μg/l, mg/l, ppb, ppm].
- Types of input:
- 1 or 2 measuring channels (analyzer main parameter).
- Optional: 1 to 4 digital sensor inputs for Memosens sensors.
- Optional: Analog current inputs.
- Optional: Binary inputs.
- Input signal (optional analog current inputs): 2 x 0/4 to 20 mA, passive, potentially isolated.
- Cable specification (for Memosens sensors): Memosens data cable CYK10 or sensor fixed cable, max. 100 m length.
Output
- Output signal:
- Standard: 2 x 0/4 to 20 mA, active, potentially isolated.
- Optional: Up to 6 x 0/4 to 20 mA, active, potentially isolated.
- Binary outputs.
- Signal on alarm: Adjustable, as per NAMUR Recommendation NE 43. Failure current factory setting is 21.5 mA.
- Load: Max. 500 Ω for current outputs.
Electrical
- Supply voltage: 100 to 120 V AC / 200 to 240 V AC or 24 V DC (24 V DC not available for "Outdoor" version). Frequency 50 or 60 Hz.
- Power consumption: Varies by version, e.g., 130 VA + 660 VA per hose heater (max. 1450 VA) for non-"Outdoor" versions without heaters, or max. 105 W for 24-V version. "Outdoor" version max. 2000 VA with cooling system.
- Fuse: 5 x 20 mm 10 A/250 V for hose trace heating system.
Mechanical/Hydraulic
- Dimensions:
- Closed version: 530 (W) x 463 (H) x 793 (D) mm.
- Open version: 530 (W) x 793 (H) x 417 (D) mm.
- With base: 1723 (H) x 930 (W) x 654 (D) mm.
- Weight:
- Cabinet version: 39.5 kg.
- Open installation: 31.5 kg.
- Analyzer stand: 72.5 kg.
- Materials (parts in contact with medium): PP, TPE, FKM, Borosilicate glass, FFKM, EPDM.
- Process connection:
- Sample inlet: Plug-in connector (OD 4 mm) or hose barb (ID 1.6 mm).
- Dilution water: Hose barb (ID 3.2 mm).
- Outlet: Hose barb (ID 13 mm).
- Hose entries: 4 x bores for M16, M20, G/NPT 3/8" or 1/2".
Environment
- Ambient temperature range:
- Standard housing: +5 to +40 °C.
- Outdoor version: -20 to +40 °C.
- Storage temperature: –20 to 60 °C.
- Relative humidity: 10 to 95 %, non-condensing.
- Degree of protection: IP55 / TYPE 3R (for cabinet and analyzer stand).
- Electromagnetic compatibility: EN 61326-1, class A for industrial areas.
- Electrical safety: EN/IEC 61010-1, Class I equipment.
- Pollution degree: 2.
- Sample temperature: 4 to 40 °C.
- Consistency of the sample: Low solids content (turbidity < 50 NTU), aqueous, homogenized.
- Sample supply: Unpressurized.
Communication
- Protocols: PROFIBUS DP, Modbus TCP/RS485, PROFINET, EtherNet/IP.
- Web server: Access via standard network connections for configuration, measured values, diagnostics, and logbooks.
Disclaimer
This summary is based on extracted text and may not include all figures, footnotes, or revisions. Contractual data must be verified against the specific OEM PDF revision used for the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.