Back to Endress + Hauser Liquid analysis
Endress+Hauser logo

Datasheet summary

Liquistation CSF34

Engineering summary from PDF text extraction for Liquistation CSF34. Verify every value with the OEM datasheet.

Summary

Product Identification

Product: Liquistation CSF34 Type: Automatic stationary sampler for liquid media; integrated controller with up to four measuring channels and digital Memosens technology.

Applications

  • Communal and industrial wastewater treatment plants
  • Laboratories and water management offices
  • Monitoring of liquid media in industrial processes

Key Features

  • All-purpose plastic ASA-PC housing material.
  • Two-door housing for reliable sample temperature regulation.
  • Cooling system with forced-air cooling and protective finish.
  • Swift menu guidance, navigator, and large display.
  • Dual bottle trays for easy sample transportation.
  • Practice-oriented programs ranging from simple time programs to event programs.
  • Functionality can be extended by installing modular electronic components.
  • Integrated data logger for recording measured values.
  • Service interface for data transmission.

Operational Modes

The CSF34 can operate using either a vacuum pump or a peristaltic pump.

Vacuum Pump Operation (4 steps)

  1. Blow clear: Suction line is cleared.
  2. Intake: Sample is drawn into the dosing beaker until conductivity probes are reached.
  3. Dose: Excess sample liquid flows back to the sampling point.
  4. Drain: Sample is drained into the sample bottle.

Peristaltic Pump Operation (3 steps)

  1. Rinse: Suction line is rinsed by forcing medium back to the sampling point.
  2. Intake: Medium is drawn in; pump is controlled by flow.
  3. Drain: Pump forces medium back to the sampling point.

Sampling Control Methods

  • Time-proportional sampling (CTCV): Constant sample volume at regular intervals.
  • Volume-proportional sampling (VTCV): Constant sample volume at variable intervals based on inflow volume.
  • Flow-proportional sampling (CTVV): Variable sample volume at regular intervals (peristaltic pump only).
  • Event-controlled sampling: Triggered by an event, can be time, volume, or flow-paced.

Input

  • 2 analog inputs (0/4 to 20 mA)
  • 2 binary inputs (passive, 12 to 30 V)
  • Optional: additional 2 binary inputs
  • Optional: 1 to 4 digital inputs for sensors with Memosens protocol

Output

  • 2 binary outputs (open collector, max. 30 V, 200 mA)
  • Optional: additional 2 binary outputs
  • Up to 2, 4, 6, or 8 x 0/4 to 20 mA active current outputs, galvanically isolated.
  • Optional HART communication (only via current output 1:1, limited with fieldbus communication).

Communication

  • HART
  • PROFIBUS DP (Profile 3.02)
  • Modbus TCP or RS485
  • PROFINET
  • EtherNet/IP
  • Web server
  • 1 service interface (optional front panel connection)

Power Supply

  • Supply Voltage: 100 to 120/200 to 240 V AC ±10 %, 50/60 Hz
  • Power Consumption:
    • Version with vacuum pump: 290 VA
    • Version with peristaltic pump: 290 VA
    • Version with sampling assembly: 290 VA
    • Version with 24V power supply: 240 W

Mechanical and Hydraulic

  • Dimensions: See official PDF for detailed diagrams.
  • Weight: Plastic version with refrigeration: 101 kg (223 lbs)
  • Materials:
    • Cabinet housing: Plastic ASA+PC V0
    • Sample compartment inner lining: Plastic PP
    • Window: Safety glass, coated
    • Insulation: Plastic EPS "Neopor®"

Environmental Conditions

  • Ambient temperature range: -30 to 50 °C (-20 to 120 °F)
  • Storage temperature: –20 to 60 °C (–4 to 140 °F)
  • Relative humidity: 10 to 95%, non-condensing
  • Degree of protection:
    • Front dosing compartment: IP 54
    • Rear dosing compartment: IP 33
    • Front panel with display: IP 65
    • Sample compartment: IP 54
    • Overall device: IP33
  • Process medium temperature range: 2 to 50 °C (36 to 122 °F)
  • Process pressure range:
    • Unpressurized, open channel
    • Max. 0.8 bar piping
    • Sampling assembly: Max. 6 bar