Summary
Memobase Plus CYZ71D is a multichannel multiparameter software for liquid analysis utilizing digital Memosens sensors. It manages the complete lifecycle of pH, ORP, conductivity, and oxygen sensors, suitable for all industries including pharmaceuticals.
Product Identification
- Product Name: Memobase Plus CYZ71D
- Document Title: Memobase Plus CYZ71D
- Technical Information Document Number: TI00502C/07/EN/05.19
- Document Revision Date: 2019-04-01
System Design
The complete measuring system includes:
- PC or Windows-based tablet with Memobase Plus software and database.
- MemoLink sensor terminal box (with Ex barrier).
- Memosens laboratory cable (CYK20) or process cable (CYK10).
- USB cable to connect MemoLink to PC.
- Memosens sensor.
Note: PC/tablet and Memosens sensors are not included and must be ordered separately.
System Requirements
- Operating System: Windows 7 Service Pack 1 (32/64-bit), Windows 10 (32/64-bit).
- Monitor: Minimum 1280x1024 resolution, touchscreen compatible.
- Processor: Minimum 1 GHz clock speed (not virtual machine).
- Hard Disk Space: Minimum 3 GB.
- RAM: 1 GB.
- USB: Type A interface, USB 2.0 minimum.
- Barcode Reader: Supported interfaces: USB-HID, USB-COM. Minimum resolution: 0.254 mm (10.0 mil).
- Other: CD/DVD drive or internet access, Adobe Reader, printer driver, Microsoft .NET Framework 4.7.
Software Functions
- Measure: Displays numerical and graphical measured values, sample description, and measurement settings. Features include sample data transmission via barcode scanning and measuring range monitoring (with "Advanced diagnostics" license).
- Calibrate: Offers guided calibration, reference solution management, live-graph monitoring, and adjustable stability criteria. Includes "as-found-as-left" reporting. The "Advanced diagnostics" license enables multipoint calibration for pH sensors. Supports FDA 21 CFR Part 11 compliance with "Pharmaceutical compliance" license.
- Sensors: Manages sensor settings, administration, status, and information. Includes time stamping for adjustments, sensor-to-measuring point assignment, calibration method specification, operating hours counter, and calibration validity checks ("Advanced diagnostics" license).
- Reports: Enables data retrieval via categories and filtering. Reports can be generated with company logos and exported to PDF, XML, or CSV formats. Supports audit trail compliance with user administration for pharmaceutical requirements. Available in multiple languages. Diagnostic messages are Namur NE 107 compliant.
Network Architecture
Memobase Plus uses a client-server architecture supporting multiple clients accessing a shared central database (Microsoft SQL Server included, Oracle interface available).
- Installation Options: Local (PC/tablet) or Central (multiple PCs/tablets).
- Operating Modes:
- Master Mode: Connects to a local or central database.
- Master-Slave Mode: A central database acts as the master, with one or more local databases as slaves for data replication.
Memosens Technology
Memosens technology enables safer and more reliable measurements through:
- Non-contact, digital signal transmission for galvanic isolation.
- Resistance to contact corrosion.
- Watertight design.
- Ability to calibrate sensors in a lab, increasing process availability.
- Predictive maintenance via recording sensor data (hours of operation, temperature, sterilizations).
MemoLink Terminal Box
- Input: Memosens port (M12 socket) for various Memosens sensors (pH/ORP, conductivity, dissolved oxygen), all containing a temperature sensor.
- Output: Mini USB 2.0 Type B, USB class HID.
- Output Voltage: 2.8 to 3.3 V.
- Output Current: 10 mA.
- Power Supply: 5 V DC via USB. USB hub requires a power unit.
- Cable Lengths:
- USB cable: 2.0 m (6.6 ft).
- Memosens lab cable CYK20: 1.5/3.0 m (4.9/9.8 ft).
- Memosens process cable CYK10: 3 to 100 m (9.8 to 328.1 ft).
Performance Characteristics
- Data Integrity: Digital data transmission ensures no measured data corruption.
Environmental Specifications
- Ambient Temperature:
- MemoLink: –10 to 50 °C (14 to 122 °F).
- CYK20 Cable: –10 to 50 °C (14 to 122 °F).
- CYK10 Cable: –25 to 135 °C (–13 to 277 °F).
- Storage Temperature:
- MemoLink: –25 to 85 °C (–13 to 185 °F).
- CYK20 Cable: –10 to 50 °C (14 to 122 °F).
- CYK10 Cable: –25 to 135 °C (–13 to 277 °F).
- Relative Humidity: Maximum 85 %, non-condensating.
- Degree of Protection:
- MemoLink: IP 65 (mated).
- CYK20 Cable: IP 68.
- CYK10 Cable: IP 68.
- Electromagnetic Compatibility: EN 61326-1:2006, Class B (Industrial).
Mechanical Construction
- Dimensions (MemoLink): 77.5 mm (3.05 in) x 31.5 mm (1.24 in) x 37 mm (1.46 in), overall length 97 mm (3.82 in). MemoLink units are stackable.
- Weight (MemoLink): 0.24 kg (0.53 lb) excluding cable.
- Materials: Housing: PBT. Housing feet: EPDM.
Certificates and Approvals
- CE Mark: Conforms to EU directives.
- Ex Approval (MemoLink): ATEX II (2) G [Ex ia Gb] II C. Operation is permitted only in non-hazardous areas. Non-Ex Memosens sensors can be connected without affecting the intrinsic safety of Ex-rated sensors.
Licensing Model
- One license per workstation, valid for unlimited time.
- Each license supports 1 to 4 MemoLink terminal boxes and sensors.
- Available licenses: Memobase Plus Basic, Advanced diagnostics, Pharmaceutics compliance, and a Demo version.
Accessories
- Kits: Kit CYZ71D MemoLink for Memosens (incl. USB cable), Kit CYZ71D USB cable.
- Measuring Cables: Memosens laboratory cable CYK20, Memosens data cable CYK10, Memosens extension cable CYK11.
- Standard Solutions: CPY20 pH buffer solutions, CLY11 Conductivity calibration solutions (various concentrations), COY8 Zero-point gel for oxygen sensors.
Sensors
The system is compatible with various Memosens sensors, including:
- pH Electrodes: Orbisint CPS11D, Memosens CPS31D, Ceraliquid CPS41D, Ceragel CPS71D, Memosens CPS171D, Orbipore CPS91D, Orbipac CPF81D, Ceramax CPS341D.
- ORP Sensors: Orbisint CPS12D, Ceraliquid CPS42D, Ceragel CPS72D, Orbipac CPF82D, Orbipore CPS92D.
- pH ISFET Sensors: Tophit CPS441D, Tophit CPS471D, Tophit CPS491D.
- pH and ORP Combined Sensors: Memosens CPS16D, Memosens CPS76D, Memosens CPS96D.
- Conductivity Sensors:
- Inductive: Indumax CLS50D, Indumax H CLS54D.
- Conductive: Condumax CLS15D, Condumax CLS16D, Condumax CLS21D, Memosens CLS82D.
- Oxygen Sensors: Oxymax COS22D, Oxymax COS51D, Memosens COS81D.