Product identification
- Product Names: Memocheck CYP02D, Memocheck Sim CYP03D
- Product Type: Testing tools for analysis measuring points utilizing Memosens Technology.
Application
- Chemicals and process engineering
- Food, pharmaceutical industry and biotechnology
- Water and wastewater treatment
- Hazardous areas
Your benefits
- Increased plant safety through complete sensor and error simulation.
- Easy, quick, and reliable sensor simulation with user-definable measured values for CYP03D and fixed values for CYP02D.
- Error simulation capability.
- Verification of parameters including pH (glass and ISFET), ORP, conductivity (conductive and toroidal), dissolved oxygen (amperometric and optical), chlorine, turbidity, and nitrate.
- High flexibility during commissioning of the measuring point.
- Utmost reliability due to Memosens Technology: no contact corrosion, completely watertight, and no interference from external potential thanks to galvanic isolation.
Function and system design
Simulation setup in non-hazardous areas
A complete measuring system consists of:
- Memocheck CYP02D or Memocheck Sim CYP03D
- Transmitter with Memosens Technology (e.g., Liquiline M CM42 or Liquiline CM44x)
- Memosens data cable CYK10
Simulation setup in hazardous areas
A complete measuring system consists of:
- Memocheck Sim CYP03D
- Transmitter with Memosens Technology and Ex approval (e.g., Liquiline M CM42)
- Memosens data cable CYK10
Performance characteristics
Memocheck CYP02D simulation values
The CYP02D consists of two interconnected sensor plug-in heads, each supplying one specific parameter. Available simulation values depend on the order configuration and include:
- pH glass and pH ISFET:
- Fixed simulation value: pH 4 at 25 °C (77 °F), pH 7 at 60 °C (140 °F)
- pH glass and pH glass:
- Fixed simulation value: pH 4 at 25 °C (77 °F), pH 7 at 60 °C (140 °F)
- pH glass and ORP:
- Fixed simulation value: pH 4 at 25 °C (77 °F), ORP 750 mV at 90 °C (194 °F)
- pH glass and oxygen:
- Fixed simulation value: pH 4 at 25 °C (77 °F), Oxygen 60 nA¹⁾ at 60 °C (140 °F)
- ¹⁾ Sensor current
- pH glass and conductive measurement of conductivity:
- Fixed simulation value: pH 4 at 25 °C (77 °F), Conductivity 20 µS/cm at 60 °C (140 °F)
- pH glass and chlorine:
- Fixed simulation value: pH 4 at 25 °C (77 °F), Chlorine 60 nA¹⁾ at 25 °C (77 °F)
- ¹⁾ Sensor current
The data listed above are displayed on the transmitters. The Memocheck CYP02D is maintenance-free.
Memocheck Sim CYP03D simulation values
The CYP03D allows simulation of:
- Simulation values: Main values, raw values, temperature.
- Parameters:
- pH glass (pH glass)
- pH glass, SIL sensor (pH glass SIL)
- pH ISFET (pH Isfet)
- ORP (ORP)
- pH + ORP combined sensor (pH + ORP)
- Conductivity conductive (Cond c)
- Conductivity, conductive, 4-pin (Cond c 4-pol)
- Conductivity, inductive (Cond i)
- Oxygen, amperometric (Oxygen (amp.))
- Oxygen, optical, Memosens (Oxygen (opt.Memo.))
- Oxygen, optical, fixed cable (Oxy. (opt.fixed))
- Chlorine (Chlorine)
- Turbidity (Turbidity)
- Nitrate (Nitrate)
- SAC (SAC)
- Main simulation values can be selected as required within sensor specification values.
- Repeated ramp with any increment.
- Error simulation (e.g., glass breakage, alarm, warning).
- Calibration values.
All values can be freely configured to match your process. The data listed above are displayed on the transmitters. The Memocheck Sim CYP03D can also be used as a qualification tool with its quality or calibration certificate. Requalification and recalibration services are available by returning the device to Endress+Hauser. The recommended testing interval is 1 year.
Environment
| Parameter | Memocheck CYP02D | Memocheck Sim CYP03D |
|---|---|---|
| Ambient temperature | –15 to 70 °C (5 to 158 °F) | –20 to 50 °C (–4 to 120 °F) |
| Storage temperature | –15 to 70 °C (5 to 158 °F) | -20 to 55 °C (-4 to 130 °F) |
| Degree of protection | IP65 | IP55 |
Electromagnetic compatibility
Interference emission and interference immunity as per EN 61326-1:2013, Class A for Industry.
Mechanical construction
Dimensions
- Memocheck CYP02D: Ø 20 mm (0.79 in) diameter, 40.0 mm (1.57 in) length.
- Memocheck Sim CYP03D: 200.0 mm (7.87 in) length, 94.0 mm (3.7 in) width, 58.0 mm (2.28 in) depth.
Weight
- Memocheck Sim CYP03D (incl. batteries): 0.3 kg (0.7 lbs)
Materials
- Memocheck CYP02D: White plastic (PET), Blue plastic (PPS GF 40).
- Memocheck Sim CYP03D: Housing (ABS, UL 94 HB).
Batteries
- Memocheck CYP02D: No batteries required.
- Memocheck Sim CYP03D: Requires Energizer EN91 (AA, 1.5 V, LR6 as per IEC) x 3 batteries. Battery storage temperature: -20 to 35 °C (-4 to 95 °F).
Certificates and approvals
- CE mark: Product meets harmonized European standards and EU directives.
- Ex approvals:
- Memocheck Plus CYP02D: ATEX/IECEx II 2G Ex ia IIC T6 Gb; FM/CSA IS NI Cl.I Div.1&2, Group A-D
- Memocheck Sim CYP03D: ATEX II 2G Ex ia IIC T4 Gb; IECEx Ex ia IIC T4 Gb; CSA IS NI Cl. I, Div. 1&2, Group A-D
Scope of delivery
Memocheck Plus CYP02D
- 1 Memocheck CYP02D with 2 plug-in heads
- 1 Operating Instructions (de + en)
Memocheck Sim CYP03D
- Memocheck Sim CYP03D
- Operating Instructions
- 1 quality certificate (as ordered)
- Cable (optional)
- Case to store CYP03D and cable (optional)
- Certificate of calibration (optional)
Accessories
- Memosens data cable CYK10: Available for Non-Ex and Ex applications. Can be used with CYP02D and CYP03D. Required for simulating fixed cable sensors (turbidity, nitrate, toroidal conductivity, oxygen optical) with CYP03D.
- Storage case for Memocheck Sim CYP03D: Available (Order No. 71183327).
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.