Datasheet summary
Memocheck CYP02E
Engineering summary from PDF text extraction for Memocheck CYP02E. Verify every value with the OEM datasheet.
Summary
Product identification
Memocheck CYP02E is a service tool for fast on-site checking of measuring systems incorporating Memosens technology.
Application
Typical applications include:
- Installation check of transmitters with Memosens technology
- Fast configuration check of transmitter and process control system
- Troubleshooting by performing sensor simulation with fixed values
Benefits
- Increased plant safety thanks to sensor and error simulation
- Time savings thanks to quick and easy installation check during commissioning
- Quick and easy checking of transmitter and PLC configuration
- Robust sensor simulator for a variety of applications
Performance characteristics
The Memocheck CYP02E consists of two interconnected sensor plug-in heads. The following simulation values can be output depending on the order:
pH glass and pH ISFET simulation
- Temperature: 25 °C (77 °F)
- pH glass: 4
- pH ISFET: 7
- Temperature: 60 °C (140 °F)
pH glass and pH glass simulation
- Temperature: 25 °C (77 °F)
- pH glass: 4
- pH glass: 7
- Temperature: 60 °C (140 °F)
pH glass and ORP simulation
- Temperature: 25 °C (77 °F)
- pH glass: 4
- ORP potential: 750 mV
- Temperature: 90 °C (194 °F)
pH glass and oxygen simulation
- Temperature: 25 °C (77 °F)
- pH glass: 4
- Oxygen: 60 nA ¹)
- Temperature: 60 °C (140 °F) ¹⁾ Sensor current
pH glass and conductive measurement of conductivity simulation
- Temperature: 25 °C (77 °F)
- pH glass: 4
- Conductivity: 20 µS/cm
- Temperature: 60 °C (140 °F)
pH glass and chlorine simulation
- Temperature: 25 °C (77 °F)
- pH glass: 4
- Chlorine: 60 nA ¹)
- Temperature: 25 °C (77 °F) ¹⁾ Sensor current
The data listed above are displayed on the transmitters. The Memocheck CYP02E is maintenance-free.
Environment
- Ambient temperature range: –15 to 70 °C (5 to 158 °F)
- Storage temperature range: –15 to 70 °C (5 to 158 °F)
- Degree of protection: IP65
- Electromagnetic compatibility: Interference emission and interference immunity as per EN 61326-1:2013, Class A for Industry
Process conditions
NOTICE: Memocheck CYP02E is not suitable for use directly in the process. The device can be damaged or destroyed in the process.
- Only use the device to simulate measured values outside the process!
- The sensor simulators must not be operated under electrostatically critical process conditions. Avoid strong steam or dust currents that act directly on the connection system.
Mechanical construction
- Dimensions:
- Height: 89 mm (3.5 in)
- Width: 25.8 mm (1.02 in)
- Depth: 37.4 mm (1.47 in)
- Diameter of plug-in heads: Ø 20 mm (0.79 in)
- Weight: 40 g
- Materials:
- White plastic: PET
- Blue plastic: PPS GF 40
Certificates and approvals
- CE mark: The product meets the requirements of the harmonized European standards.
- Hazardous area approvals:
- ATEX: II 2G Ex ia IIC T6 Gb
- IECEx: Ex ia IIC T6 Gb
- NEPSI: NEPSI Ex ia IIC T6 Gb
- CSA C/US: CSA C/US IS Cl. I Div. 1 GP A-D T6; CSA C/US IS Cl. I Zone 0 AEx ia IIC T6
- Japan Ex: JPN Ex ia IIC T6 Gb
- INMETRO: INMETRO Ex ia IIC T6 Gb
Ex versions of digital sensor simulators with Memosens technology are identified by an orange-red ring on the plug-in head.
Scope of delivery
- 1 Memocheck CYP02E with 2 plug-in heads
- 1 x Operating Instructions
Accessories
- Memosens data cable CYK10
- Memosens laboratory cable CYK20
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.