Datasheet summary
Memosens 31E
Engineering summary from PDF text extraction for Memosens 31E. Verify every value with the OEM datasheet.
Summary
Product identification
Product Name: Memosens CPS31E Product Type: pH sensor for standard applications in drinking water and swimming pool water Technology: Memosens 2.0
Application
- Drinking water
- Swimming pool water
- pH compensation during the measurement of free chlorine
The sensor is approved for use in hazardous areas (Zone 0, Zone 1, and Zone 2) with ATEX, IECEx, CSA C/US, NEPSI, Japan Ex, and INMETRO certifications.
Operating limits
- Maximum pressure: 4 bar (58 psi) absolute
- Maximum temperature: 80 °C (176 °F)
- Process temperature range: –15 to 80 °C (5 to 176 °F)
- Process pressure range: 0.8 to 4 bar (11.6 to 58 psi) absolute
Feed water
- Conductivity (Reference system AB and SB): Minimum 100 μS/cm
- Conductivity (Reference system AC and SC): Minimum 50 μS/cm
Product water
- pH measuring range: 1 to 12
- Temperature measuring range: –15 to 80 °C (5 to 176 °F)
Electrical
- Technology: Memosens digital technology with inductive, non-contact signal transmission.
- Integrated temperature sensor: NTC 30K for temperature compensation.
Mechanical / hydraulic
- Reference system: Ag/AgCl with gel (3M KCl) or salt storage (KCl-saturated).
- Junctions: 1 or 3 ceramic junctions.
- Electrolyte: Very low ion depletion.
- Dimensions:
- Installed length: 120 mm (4.72 in)
- Diameter: Ø 11.9 mm (0.47 in)
- Weight: 40 g (1.4 oz)
- Materials:
- Sensor shaft: Glass
- pH membrane glass: Type A
- Metal lead: Ag/AgCl
- O-ring: FKM
- Process coupling: PPS fiber-glass reinforced
- Plug-in head: Memosens plug-in head (pressure resistance 16 bar / 232 psi relative)
- Process connections: Pg 13.5
- Degree of protection: IP 68 (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)
- Installation orientation: Minimum 15° from the horizontal.
- Tightening torque: 3 Nm (2.21 lbf ft) (for Endress+Hauser assemblies).
Certificates and approvals
- CE marking
- Ex approvals: ATEX, IECEx, NEPSI, CSA C/US, Japan Ex, INMETRO (II 1G Ex ia IIC T4/T6 Ga or equivalent). Ex versions have an orange-red ring on the plug-in head.
- EAC conformity