Memosens 61E

Engineering summary from PDF text extraction for Memosens 61E. Verify every value with the OEM datasheet.

Summary

Product Identification

Memosens CPS61E pH sensor for bioreactors in life sciences and for the food industry.

Application

Hygienic and sterile applications (sterilizable, autoclavable) in:

  • Bioreactor/fermenter
  • Biotechnology
  • Pharmaceutical industry
  • Foods

Approvals for use in explosive atmospheres (Zone 0, 1, 2): ATEX, IECEx, CSA C/US, NEPSI, JPN Ex, INMETRO.

Benefits

  • Biocompatibility (in vitro cytotoxicity and in vivo) tested for relevant parts in contact with the process medium.
  • Suitable for CIP/SIP cleaning and autoclavable up to 140 °C (284 °F).
  • Pressurized reference (TP reference system) for fermentation processes, with integrated pressure indicator.
  • Alternative TU reference system for upside-down installation with solidified gel.
  • Long service life with poison-resistant reference and improved ion trap.
  • Bridging electrolyte is silver ion-free and acrylamide-free.
  • Integrated NTC 30K temperature sensor for compensation.
  • Parts in contact with the process medium are not derived from animals, minimizing TSE/BSE risk per EMA.
  • Memosens technology provides:
    • Maximum process safety via non-contact inductive signal transmission.
    • Digital data transmission ensuring data security.
    • Ease of use with sensor data stored in the sensor.
    • Predictive maintenance capabilities by recording sensor load data.

Measuring Principle

  • pH Measurement: The electrode's membrane glass generates an electrochemical potential dependent on the medium's pH, due to H⁺ ion accumulation. An Ag/AgCl reference system is used. The measured voltage is converted to a pH value using the Nernst equation.
  • Temperature Sensor: Integrated NTC 30K for temperature compensation.

Measuring System Components

A complete measuring system includes:

  • pH sensor CPS61E
  • Memosens data cable (CYK10 or CYK20)
  • Transmitter (e.g., Liquiline CM44, Liquiline CM42)
  • Assembly (e.g., retractable Cleanfit CPA875, permanent installation Unifit CPA842)
  • Optional: Automatic cleaning and calibration system (e.g., Liquiline Control CDC90)

Communication and Data Processing

  • Digital sensors with Memosens technology must be connected to a Memosens transmitter.
  • The sensor stores data including: manufacturer details, serial number, order code, calibration data, calibration history, operating data, temperature and pH application ranges, commissioning date, operational limits, sterilization cycles, CIP counter, and sensor load.
  • This data can be displayed using Liquiline CM42, CM44x, CM44x/R, and Memobase Plus CYZ71D.

Dependability and Reliability

  • Integrated electronics store calibration and operational data.
  • Automatic data transfer to the transmitter upon connection.
  • Sensor calibration and adjustment can be performed independently of the measuring point.
  • Allows for easy calibration in a lab environment and quick replacement of pre-calibrated sensors.
  • Enables accurate maintenance interval definition and predictive maintenance.
  • Sensor history can be documented externally.

Integrity and Data Security

  • Measured values are digitized within the sensor and transmitted wirelessly, preventing interference.
  • Reliable detection and reporting of sensor failure or connection interruptions.

Safety

  • Inductive transmission via non-contact connection ensures process safety.
  • Eliminates issues caused by moisture, such as connection corrosion or distorted readings.
  • Transmitter is galvanically decoupled from the medium.
  • Electromagnetic compatibility (EMC) is ensured through screening.
  • Intrinsically safe electronics allow operation in hazardous areas with appropriate Ex approvals.

Input

  • Measured Variable: pH value, Temperature
  • Measuring Range:
    • pH: 0 to 14
    • Temperature: 0 to 100 °C (32 to 212 °F)
  • Conductivity: Minimum 100 μS/cm (at minimized flow, constant pressure, and temperature).

Performance Characteristics

  • Reference System:
    • TP reference system: Ag/AgCl reference lead with ion trap, 3 M KCl electrolyte (flowable gel), pressurized to 7 bar (102 psi) absolute. Features an integrated pressure indicator.
    • TU reference system: Ag/AgCl reference lead with ion trap, 3 M KCl electrolyte (solidified gel).

Mounting and Installation

  • Orientation:
    • TP reference system: Minimum 15° angle from the horizontal; not permitted to be installed upside-down.
    • TU reference system: Suitable for upside-down installation; can be installed at any angle.
  • Installation Torque: 3 Nm (2.21 lbf ft) when installing in Endress+Hauser assemblies.
  • Hygienic Requirements: For 3-A or EHEDG compliance, use certified assemblies, assemblies with protective guards, ensure self-draining installation, and avoid dead areas.

Environment

  • Ambient Temperature Range: Not specified, but avoid use below 0 °C (32 °F) due to risk of frost damage.
  • Storage Temperature: 0 to 50 °C (32 to 122 °F).
  • Degree of Protection: IP 68 (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl).
  • Electromagnetic Compatibility (EMC): Complies with EN 61326-1, EN 61326-2-3, and NAMUR NE21.

Process

  • Process Temperature Range:
    • Application M: 0 to 100 °C (32 to 212 °F); up to 130 °C (266 °F) for sterilization.
    • Application N: 0 to 100 °C (32 to 212 °F); up to 140 °C (284 °F) for sterilization.
  • Process Pressure Range: 0.8 to 7 bar (11.6 to 101.5 psi) absolute for Application M and N.
  • Pressure-Temperature Ratings: See the pressure-temperature diagram in the document for specific application limits.

Mechanical Construction

  • Design: Available in several installed lengths: 120 mm (4.72 in), 225 mm (8.86 in), 360 mm (14.17 in), 425 mm (16.73 in).
  • Weight: Varies by installed length from 40 g (1.4 oz) to 100 g (3.5 oz).
  • Materials:
    • Sensor Shaft: Glass
    • pH Membrane Glass: Type N
    • Metal Lead: Ag/AgCl
    • Junction: Ceramic (zirconium dioxide)
    • O-ring: FKM
    • Process Coupling: PPS fiber-glass reinforced
    • Nameplate: Ceramic metal oxide
    • Temperature Sensor: NTC 30K
    • Plug-in Head: Memosens plug-in head (pressure resistance 16 bar (232 psi) relative).
  • Process Connections: Pg 13.5

Certificates and Approvals

Current certificates and approvals are available via the Product Configurator on www.endress.com.

Scope of Delivery

  • Sensor (as ordered)
  • Operating Instructions
  • Safety instructions for hazardous areas (if applicable)

Accessories

  • Assemblies: Unifit CPA842, Cleanfit CPA875
  • Buffer Solutions: CPY20
  • Measuring Cables: Memosens data cable CYK10, Memosens laboratory cable CYK20

Disclaimer: This extracted summary may not include all figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used for the project.