Product identification
- Product Name: Memosens CCS51E
- Description: Digital sensor with Memosens technology for determining free chlorine.
- Document Revision: TI01768C/07/EN/01.23-00, dated 2023-10-13
Application
The Memosens CCS51E sensor is suitable for measuring free chlorine in various water qualities, including:
- Drinking water
- Utilities
- Process water for food and beverage packaging
- Swimming pools
Measuring principle
The sensor uses an amperometric measuring principle to determine free chlorine via hypochlorous acid (HOCl). HOCl diffuses through a membrane and is reduced at a gold working electrode, while silver is oxidized at a silver counter electrode, generating a current proportional to the free chlorine concentration. An additional pH measurement is recommended for compensation due to HOCl's dependence on pH. The transmitter converts the current signal to a concentration value (mg/l or ppm).
Cross-sensitivity
The sensor exhibits cross-sensitivities to:
- Chlorine dioxide
- Ozone
- Free bromine
There are no cross-sensitivities to H₂O₂ or peracetic acid.
Measuring system
A complete measuring system includes:
- Memosens CCS51E disinfection sensor (membrane-covered, ⌀25 mm)
- Appropriate mounting adapter (e.g., Flowfit CYA27 flow assembly or Flexdip CYA112 immersion assembly)
- Memosens measuring cable (CYK10 or CYK20)
- Transmitter (e.g., Liquiline CM44x or CM44xR with specific firmware)
- Optional accessories like extension cables or pH sensors.
Input
Measured variables
- Free chlorine (HOCl) in mg/l, μg/l, ppm, ppb
- Temperature in °C, °F
Measuring range
- CCS51E-**11AD** : 0 to 5 mg/l (ppm) HOCl
- CCS51E-**11BF** : 0 to 20 mg/l (ppm) HOCl
- CCS51E-**11CJ** : 0 to 200 mg/l (ppm) HOCl
Signal current
- CCS51E-**11AD** : 33 to 63 nA per 1 mg/l (ppm) HOCl
- CCS51E-**11BF** : 9 to 18 nA per 1 mg/l (ppm) HOCl
- CCS51E-**11CJ** : 9 to 18 nA per 1 mg/l (ppm) HOCl
Performance characteristics
Reference operating conditions
- Temperature: 20 °C (68 °F)
- pH value: pH 5.5 ±0.2
- Flow: 40 to 60 cm/s
- Medium: HOCl-free base medium (Mains water)
Response time
- T<0xE2><0x82><0x90> < 25 s (after polarization completion)
Polarization time
- Initial commissioning: 45 min
- Recommissioning: 20 min
Measured value resolution of sensor
- CCS51E-**11AD** : 0.03 µg/l (ppb) HOCl
- CCS51E-**11BF** : 0.13 µg/l (ppb) HOCl
- CCS51E-**11CJ** : 1.10 µg/l (ppb) HOCl
Measurement error (LOD / LOQ, based on ISO 15839)
- CCS51E-**11AD** : LOD 0.002 mg/l, LOQ 0.005 mg/l
- CCS51E-**11BF** : LOD 0.002 mg/l, LOQ 0.007 mg/l
- CCS51E-**11CJ** : LOD 0.008 mg/l, LOQ 0.027 mg/l
Repeatability
- CCS51E-**11AD** : 0.0031 mg/l (ppm)
- CCS51E-**11BF** : 0.0035 mg/l (ppm)
- CCS51E-**11CJ** : 0.062 mg/l (ppm)
Nominal slope
- CCS51E-**11AD** : 48 nA per 1 mg/l (ppm) HOCl
- CCS51E-**11BF** : 14 nA per 1 mg/l (ppm) HOCl
- CCS51E-**11CJ** : 14 nA per 1 mg/l (ppm) HOCl
Long-term drift
- < 1 % per month (mean value)
Operating time of the electrolyte
- At 10% of measuring range and 20 °C: 2 years
- At 50% of measuring range and 20 °C: 1 year
- At maximum concentration and 55 °C: 60 days
Installation
Orientation
The sensor must be installed at an angle of at least 15° to the horizontal. Installation upside-down is not permitted.
Immersion depth
- At least 50 mm (1.97 in), corresponding to the mark on the sensor.
Installation instructions
- Flowfit CYA27: Ensure minimum flow of 15 cm/s (0.49 ft/s) and minimum volume flow (5 l/h or 30 l/h depending on version). Maximum permissible backpressure is 1 bar relative. Avoid negative pressure. Filter heavily contaminated water.
- Other flow assemblies: Ensure a minimum flow velocity of 15 cm/s at the membrane, upward flow direction, and direct exposure of the membrane to the flow.
- CYA112 immersion assembly: Ensure sufficient flow towards the sensor.
Environment
Ambient temperature
- –20 to 60 °C (–4 to 140 °F)
Storage temperature
- Long-term (up to 2 years): 0 to 35 °C (32 to 95 °F) (non-freezing)
- Short-term (up to 48 h):
- With electrolyte: 0 to 35 °C (32 to 95 °F)
- Without electrolyte: –20 to 60 °C (–4 to 140 °F)
Degree of protection
- IP68 (submersible up to 1.8 m for 7 days at 20 °C)
Process
Process temperature
- 0 to 55 °C (32 to 130 °F), non-freezing
Pressure
- Up to 1 bar relative (14.5 psi relative) / 2 bar absolute (29 psi absolute). Measurement can occur with a free outlet.
pH range
- Range of effectiveness: pH 4 to 9
- Calibration: pH 4 to 8
- Measurement: pH 4 to 9
Conductivity
- Can be used in low conductivity media, but attention must be paid to reduced pH buffer capacity.
Flow
- Minimum flow velocities and volume flows:
- Flowfit CYA27 (5 l version): 5 l/h (1.3 gal/h)
- Flowfit CYA27 (30 l version): 30 l/h (7.9 gal/h)
- Flexdip CYA112: 15 cm/s (0.5 ft/s)
- Maximum flow velocities:
- Flowfit CYA27 (5 l version): 30 l/h
- Flowfit CYA27 (30 l version): 60 l/h
- Flexdip CYA112: 80 cm/s
Mechanical construction
Dimensions
- See diagram in the original document (overall length approximately 161 mm).
Weight
- Approximately 95 g (3.35 oz) (sensor with membrane cap and electrolyte, without protection cap and adapter).
Materials
- Sensor shaft: POM
- Membrane: PVDF
- Membrane cap: PVDF
- Protective cap: PC Makrolon (vessel, cover), Kraiburg TPE TM5MED (seal)
- Sealing ring: FKM
- Sensor shaft coupling: PPS
Cable specification
- Maximum cable length: 100 m (330 ft), including extension cables.
Accessories
Maintenance kit
- CCV05: Contains membrane caps and electrolyte.
Device-specific accessories
- Memosens data cables (CYK10, CYK11, CYK20)
- Flowfit CYA27 flow assembly
- Flexdip CYA112 immersion assembly
- Adapter kits for CYA27 and CYA112
- COY8 Zero-point gel