Datasheet summary

Memosens CCS53E

Engineering summary from PDF text extraction for Memosens CCS53E. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product Name: Memosens CCS53E
  • Description: Digital sensor with Memosens technology for determining total chlorine.
  • Document Title: Memosens CCS53E
  • Document Revision: TI01790C/07/EN/01.24-00

Application

The Memosens CCS53E sensor is used for total chlorine measurement in:

  • Wastewater treatment plants (disinfection monitoring and reuse).
  • Process water and utilities (chlorine detection in cooling towers).
  • Food and beverage industry (hygienic filling and washing water).
  • Drinking water (ensuring adequate disinfection without overdosing).
  • Swimming pools (efficient disinfectant dosing).

Operating limits

  • Ambient temperature range: –20 to 60 °C (–4 to 140 °F)
  • Storage temperature range:
    • Long-term storage (up to 2 years max.): 0 to 35 °C (32 to 95 °F) (non-freezing).
    • Short-term storage (up to 48 h max.): 35 to 55 °C (95 to 131 °F).
    • Without electrolyte: –20 to 60 °C (–4 to 140 °F).
  • Process temperature range: 0 to 55 °C (32 to 130 °F), non-freezing.
  • Process pressure: Up to 1 bar relative (14.5 psi relative) (2 bar absolute (29 psi absolute)).
  • pH range: 4 to 9

Feed water

  • Conductivity: Suitable for media with very low conductivity, such as demineralized water.
  • Flow limits (in Flowfit CYA27 assembly):
    • Minimum volume flow (5 l version): 5 l/h (1.3 gal/h)
    • Minimum volume flow (30 l version): 30 l/h (7.9 gal/h)
  • Flow velocity (minimum): 15 cm/s (0.5 ft/s)

Product water

  • Measuring range:
    • Version CCS53E-11AD: 0 to 5 mg/l (ppm) total chlorine.
    • Version CCS53E-11BF: 0 to 20 mg/l (ppm) total chlorine.

Electrical

  • Signal current:
    • Version CCS53E-11AD: 8 to 20 nA per 1 mg/l (ppm) HOCl.
    • Version CCS53E-11BF: 8 to 20 nA per 1 mg/l (ppm) HOCl.
  • Electrical connection: Via Memosens data cable CYK10.
  • Cable specification: Max. 100 m (330 ft), including cable extension.

Mechanical / hydraulic

  • Dimensions:
    • Sensor diameter: ⌀25 mm (0.98 in)
  • Weight: Approx. 95 g (3.35 oz) (sensor with membrane cap and electrolyte, without protection cap and adapter).
  • Materials:
    • Sensor shaft: POM
    • Membrane: PET/PVDF
    • Membrane cap: PVDF
  • Mounting orientation: Install at an angle of at least 15° to the horizontal; do not install upside down.
  • Immersion depth: At least 50 mm (1.97 in).

Performance characteristics

  • Response time: T90 < 180 s (after completing polarization).
  • Polarization time:
    • Initial commissioning: 45 min.
    • Recommissioning: 20 min.
  • Measured value resolution: 0.05 % of the measured value maximum.
  • Limit of detection (LOD): 0.008 mg/l (ppm) HOCl for both versions.
  • Limit of quantification (LOQ): 0.028 mg/l (ppm) HOCl for both versions.
  • Repeatability:
    • Version CCS53E-11AD: 0.004 mg/l (ppm).
    • Version CCS53E-11BF: 0.007 mg/l (ppm).
  • Nominal slope: 11 nA per 1 mg/l (ppm) HOCl for both versions.
  • Long-term drift: < 1 % per month (mean value).
  • Operating time of the electrolyte: 3 to 6 months (depending on water quality).