Back to Endress + Hauser Liquid analysis
Endress+Hauser logo

Datasheet summary

Memosens CLL47E

Engineering summary from PDF text extraction for Memosens CLL47E. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product name: Memosens CLL47E
  • Product type: Contacting conductivity sensor for laboratory measurements and random sampling in the field
  • Technology: Digital with Memosens 2.0 technology

Application

Designed for measurements where very diverse conductivities need to be measured within a single system. Suitable for sticky and viscous samples due to easy cleaning with mechanically polished surfaces.

Measuring principle

A 4-electrode system applies an alternating current via the outer electrodes and measures the voltage at the inner electrodes. This allows for reliable electrolytic conductivity determination, suppressing electrochemical effects at the electrodes.

Input

  • Measured variables: Conductivity, Temperature
  • Measuring ranges:
    • Conductivity: 5 µS/cm to 200 mS/cm (in relation to water at 25 °C)
    • Temperature: 0 to 100 °C (32 to 212 °F)
  • Cell constant: k = 0.57 cm⁻¹
  • Temperature compensation: Pt1000 (Class A according to IEC 60751)

Performance characteristics

  • Cell constant uncertainty: 1.0 % (determined at factory)
  • Measured error:
    • Conductivity:
      • 5 µS/cm to 1 mS/cm: ≤ 2 % of reading
      • 1 mS/cm to 200 mS/cm: ≤ 4 % of reading
    • Temperature: ≤ 1.0 K (in measuring range 0 to 100 °C)
  • Repeatability:
    • Conductivity: ≤ 0.5 % of reading (in specified measuring range)
    • Temperature: ≤ 0.5 K

Environment

  • Ambient temperature: –20 to 60 °C (–4 to 140 °F)
  • Storage temperature: -25 to +80 °C (-13 to +176 °F)
  • Humidity: 5 to 95 %
  • Degree of protection: IP 68 / NEMA type 6P

Conditions for outdoor use

  • Connection via CYK10-A052 cable, use of spacer.
  • Maximum duration: 30 minutes.
  • Frequency: Maximum twice per week.
  • Maximum insertion depth: 5 m.
  • Maximum medium temperature: 50 °C.

Mechanical construction

  • Dimensions: Ø11.9 mm (0.47 in)
  • Weight: Max. 0.06 kg (0.13 lbs)
  • Materials (in contact with medium):
    • Sensor element: Platinum and ceramic (zirconium oxide)
    • Process connection: Stainless steel 1.4435 (AISI 316L)

Communication and data processing

Digital laboratory sensors with Memosens technology communicate with handheld devices like the CML18. Sensor data stored include manufacturer information, serial number, order code, calibration data, and application ranges.

Accessories

  • Memosens laboratory cable CYK20
  • Memosens data cable CYK10
  • Conductivity calibration solutions CLY11
  • Liquiline Mobile CML18 (transmitter)
  • Memobase Plus CYZ71D (PC software)

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.