Datasheet summary
Memosens CLS16E
Engineering summary from PDF text extraction for Memosens CLS16E. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: Memosens CLS16E
- Product Type: Digital conductivity sensor with Memosens technology
- Document ID: TI01527C/07/EN/03.24-00
- Version: 71649265, dated 2024-03-31
Application
Measurements in pure and ultrapure water, including:
- Ion exchanger monitoring
- Reverse osmosis
- Distillation
- Electrodeionization
- WFI (water for injection) in the pharmaceutical industry
Sensors with temperature probes are used with conductivity measuring devices supporting automatic temperature compensation, such as:
- Liquiline CM442/CM444/CM448
- Liquiline CM42
- Liquiline CM14
Transmitters can also measure resistivity in MΩ · cm.
Features and Benefits
- High measuring accuracy due to individually measured cell constant.
- Manufacturer inspection certificate with individual cell constant.
- Hygienic process connections for pipe or flow vessel installation.
- Easy to clean with electropolished surfaces.
- Sterilizable up to 150 °C (302 °F).
- Material: Stainless steel 1.4435 (AISI 316L).
- Certifications: EHEDG Document 8.
- Optional: USP 87, USP 88 Class VI certificate, EN 10204 3.1 inspection certificate.
Memosens Technology Advantages:
- Maximum process safety.
- Data security via digital data transmission.
- Easy to use with sensor data saved in the sensor.
- Recording of sensor load data for predictive maintenance.
Input
- Measured Variables: Conductivity, Temperature
- Measuring Ranges:
- Conductivity: 40 nS/cm to 500 µS/cm (in relation to water at 25 °C (77 °F))
- Temperature: -5 to 150 °C (23 to 302 °F)
- Cell constant: k = 0.1 cm⁻¹
- Temperature compensation: Pt1000 (Class A according to IEC 60751)
Performance Characteristics
- Measurement uncertainty (Cell constant): 1.0 %
- Response time:
- Conductivity (t<0xE2><0x82><0x95>): ≤ 2 s
- Temperature (t<0xE2><0x82><0x99>): ≤ 9 s (1)
- Measurement error:
- Conductivity: ≤ 2 % of reading (in specified measuring range)
- Temperature: ≤ 0.5 K (in measuring range -5 to 120 °C); ≤ 1.0 K (in measuring range 120 to 150 °C)
- Repeatability:
- Conductivity: ≤ 0.2 % of reading (in specified measuring range)
- Temperature: ≤ 0.05 K
(1) DIN VDI/VDE 3522-2 (0.3 m/s laminar)
Environment
- Ambient temperature: -20 to 60 °C (–4 to 140 °F)
- Storage temperature: -25 to +80 °C (-10 to +180 °F)
- Degree of protection: IP 68 / NEMA type 6P (1 m water column, 25 °C, 24 h)
Process
- Process temperature:
- Normal operation: -5 to 120 °C (23 to 248 °F)
- Sterilization (max. 45 min): Max. 150 °C (302 °F) at 6 bar (87 psi) absolute
- Pressure:
- 13 bar (188 psi) absolute, at 20 °C (68 °F)
- 9 bar (130 psi) absolute, at 120 °C (248 °F)
- 0.1 bar (1.5 psi) absolute (vacuum), at 20 °C (68 °F)
Mechanical Construction
- Design:
- Memosens plug-in head
- Process connection: Electropolished stainless steel 1.4435 (AISI 316L) (Clamp, Varivent, BioControl)
- Coaxial measuring electrode: Electropolished stainless steel 1.4435 (AISI 316L)
- Dimensions: Refer to diagrams for specific versions (Clamp ISO2852, BioControl, Varivent).
- Weight: Approx. 0.13 to 0.75 kg (0.29 to 1.65 lbs) depending on version.
- Materials (in contact with medium):
- Sensor: Electropolished stainless steel 1.4435 (AISI 316L) or PEEK.
- Seal: Molded seal FFKM or Molded seal EPDM.
- Process connection: 1½", 2" (ISO 2852, suitable for TRI-CLAMP, DIN 32676), Tuchenhagen VARIVENT N DN 50 to 125, NEUMO BioControl D50.
- Surface roughness: Ra ≤ 0.38 μm, electropolished.
Certificates and Approvals
- Ex approvals: Various ATEX, CSA, EAC, NEPSI certifications listed by model suffix (e.g., CLS16E-BA, CLS16E-CI).
- Hygienic compatibility: EHEDG Type EL Class I, Regulation (EC) No. 1935/2004, FDA compliant, GB4806.1-2016.
- Pharmaceutical compatibility: cGMP requirements, USP 87, USP 88 Class VI, TSE-/BSE-free, ASME BPE compliant.
- CRN approval: Registered in all Canadian provinces according to CSA B51.
- Test reports: Manufacturer's certificate (cell constant), Surface roughness test, EN 10204 3.1 (optional).
- External standards: EAC certified.
Accessories
- Measuring cable: Memosens data cable CYK10, CYK11.
- Sensor regeneration: Factory service for seal replacement and recalibration.
- Calibration solutions: Conductivity calibration solutions CLY11 (e.g., CLY11-A 74 μS/cm).
- Calibration set: Conducal CLY421 for ultrapure water.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.