Back to Endress + Hauser Liquid analysis
Endress+Hauser logo

Datasheet summary

Memosens CPS11E

Engineering summary from PDF text extraction for Memosens CPS11E. Verify every value with the OEM datasheet.

Summary

Product identification

Memosens CPS11E pH sensor for standard applications in process technology and environmental engineering. Digital with Memosens 2.0 technology.

Application

Designed for long-term monitoring and limit control in processes with stable conditions. Suitable for:

  • Chemical industry (strong acids/bases)
  • Plastic industry
  • Pulp and paper industry
  • Power plants (e.g., flue gas cleaning)
  • Oil and gas industry
  • Incinerator plants
  • Water and wastewater treatment
  • Boiler feedwater and cooling water
  • Well water and drinking water
  • All industrial and municipal treatment plants

Features ATEX, IECEx, CSA C/US, NEPSI, Japan Ex, and INMETRO approvals for use in hazardous areas Zone 0, Zone 1, and Zone 2.

Operating limits

  • Maximum pressure: Up to 17 bar (246.5 psi) absolute for applications A and B. Up to 7 bar (101.5 psi) absolute for application F.
  • Temperature ranges:
    • Application A: –15 to 80 °C (5 to 176 °F)
    • Application B: 0 to 135 °C (32 to 275 °F)
    • Application F: 0 to 70 °C (32 to 158 °F)
  • Ambient temperature range: Do not use below –15 °C (5 °F).
  • Storage temperature: 0 to 50 °C (32 to 122 °F).
  • Installation angle: Must be at least 15° from the horizontal.

Feed water

  • Conductivity:
    • Reference system AA, TA: minimum 50 μS/cm.
    • Reference system AS, TS: minimum 0.1 μS/cm.

Product water

  • pH measuring range:
    • Application A: 1 to 12 pH
    • Application B: 0 to 14 pH
    • Application F: 0 to 10 pH

Electrical

  • Technology: Memosens 2.0 digital technology with non-contact, inductive signal transmission.
  • Integrated sensor: NTC 30K temperature sensor for temperature compensation.
  • Data transmission: Digital, interference-free. Stores manufacturer data, serial number, order code, date of manufacture, calibration data, operating data, sensor load, and more.

Mechanical / hydraulic

  • Reference system options:
    • AA: PTFE reference junction, 3M KCl bridging electrolyte.
    • AS: PTFE reference junction, saturated KCl bridging electrolyte, salt storage.
    • TA: PTFE reference junction, Ag±-free, ion trap, 3M KCl bridging electrolyte.
    • TS: PTFE reference junction, Ag±-free, ion trap, saturated KCl bridging electrolyte, salt storage. (Note: TS may display incorrect values with rapid temperature changes; use with constant temperatures recommended).
  • Junction: Large PTFE ring junction.
  • Materials:
    • Sensor shaft: Glass.
    • pH membrane glass: Type A, B, F.
    • Metal lead: Ag/AgCl.
    • O-ring: FKM.
    • Process coupling: PPS fiber-glass reinforced.
    • Plug-in head: Memosens plug-in head, pressure resistance 16 bar (232 psi) relative.
  • Process connection: Pg 13.5.
  • Dimensions: See diagrams in the original document for specific lengths (120 / 225 / 360 / 425 mm) and weights.
  • Weight: Ranges from 40 g to 100 g depending on installed length.
  • Degree of protection: IP 68.

Disclaimer

This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.