Datasheet summary
Memosens CPS12E
Engineering summary from PDF text extraction for Memosens CPS12E. Verify every value with the OEM datasheet.
Summary
Product identification
Memosens CPS12E ORP sensor for standard applications in process technology and environmental engineering with Memosens 2.0 technology.
Application
- Long-term monitoring and limit control in processes with stable process conditions
- Chemical industry
- Pulp and paper industry
- Power plants (e.g. flue gas cleaning)
- Incinerator plants
- Water treatment
- Drinking water
- Cooling water
- Well water
The sensor is available with ATEX, IECEx, CSA C/US, NEPSI, Japan and INMETRO approvals for use in hazardous areas Zone 0, Zone 1 and Zone 2.
Your benefits
- Robust sensor with long diffusion path for poisoning substances
- Low-maintenance due to large, dirt-repellent PTFE diaphragm
- Suitable for tough applications: process glass for highly alkaline media
- Can be used at pressures up to 17 bar (246.5 psi) (absolute)
- Integrated NTC 30K temperature sensor
- Different measuring elements for use in oxidizing and reducing media
Memosens technology advantages
- Maximum process safety thanks to non-contact, inductive signal transmission
- Data security thanks to digital data transmission
- Very easy to use as sensor data are saved in the sensor
- Predictive maintenance can be performed by recording sensor load data in the sensor
Function and system design
Measuring principle
- ORP measurement: Measures the potential of equilibria between oxidizing and reducing components using a platinum or gold electrode with an integrated Ag/AgCl reference system.
- Measuring system: Comprises the ORP sensor CPS12E, Memosens data cable (CYK10 or CYK20), and a compatible transmitter (e.g., Liquiline CM44, CM42). Various assemblies are available for installation.
ORP sensor options
- Gold electrode: For oxidizing media (e.g., cyanide oxidation, nitrite oxidation, ozone, hydrogen peroxide).
- Platinum electrode: For reducing media (e.g., chromate reduction) or chlorine dosing.
Communication and data processing
- Digital sensors with Memosens technology must be connected to transmitters with Memosens technology.
- Digital sensors store data including manufacturer information, serial number, order code, calibration data, operating data, and sensor history.
Input
- Measured variable: ORP, Temperature
- Measuring range: –1 500 to 1 500 mV
Performance characteristics
- Reference system: Ag/AgCl reference lead with Advanced Gel 3 M KCl
Installation
- Orientation: Install at an angle of at least 15° from the horizontal to prevent electrolyte separation from the diaphragm at elevated temperatures. Do not install upside-down.
- Installation instructions: Ensure assembly thread, O-rings, and sealing surface are clean and undamaged. Tighten sensor by hand with a torque of 3 Nm (2.21 lbf ft) when installing in Endress+Hauser assemblies.
Environment
- Ambient temperature range: –15 °C (5 °F) to 135 °C (275 °F) (Operating), 0 to 50 °C (32 to 122 °F) (Storage)
- Storage temperature: 0 to 50 °C (32 to 122 °F)
- Degree of protection: IP 68 (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)
- Electromagnetic compatibility (EMC): Interference emission and interference immunity as per EN 61326-1: 2013.
Process
- Process temperature range: –15 to 135 °C (5 to 275 °F)
- Process pressure range: 0.8 to 17 bar (11.6 to 246.5 psi) absolute. Avoid rapid heating of pressurized sensors under reduced or atmospheric pressure.
- Conductivity: Minimum 50 μS/cm for reference system AA (with minimized flow; stable pressure and temperature required).
Mechanical construction
- Design, dimensions: See diagram 5. Available in installed lengths of 120 mm (4.72 in), 225 mm (8.86 in), 360 mm (14.17 in), and 425 mm (16.73 in).
- Weight: Ranges from 40 g (1.4 oz) to 100 g (3.5 oz) depending on installed length.
- Materials:
- Sensor shaft: Glass
- ORP measuring element: Platinum or gold
- Metal lead: Ag/AgCl
- Aperture: Ring-shaped PTFE diaphragm
- O-ring: FKM
- Process coupling: PPS fibre-glass reinforced
- Nameplate: Ceramic metal oxide
- Temperature sensor: NTC 30K
- Plug-in head: Memosens plug-in head (pressure resistance 16 bar (232 psi) relative)
- Process connections: Pg 13.5
Certificates and approvals
- CE mark: Complies with EU directives.
- Ex approval: ATEX, IECEx, NEPSI, CSA C/US, Japan Ex, INMETRO approvals for use in hazardous areas. Ex versions are identified by an orange-red ring on the plug-in head.
- TÜV certificate: For Memosens plug-in head, pressure resistance 16 bar (232 psi) relative.
- EAC: Complies with TP TC 004/2011 and TP TC 020/2011.
Ordering information
- Product page for configuration: www.endress.com/cps12e
Scope of delivery
- Sensor in ordered version
- Operating Instructions
- Safety instructions for hazardous areas (for sensors with Ex approval)
Accessories
Various assemblies, buffer solutions, and measuring cables are available. Refer to the official product documentation for a complete list.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.